Requiredshould be as indicated in the bid document. Documentary evidence in the form of certified Audited Balance Sheets of relevant periods or a certificate from the Chartered Accountant/Cost Accountant indicating the turnover details for the relevant period shall be uploaded with the bid. In case the date of constitution/incorporation of the OEM is less than 3 year old
Requiredthe average turnover in respect of the completed financial years after the date of constitution shall be taken into account for this criteria. AUTOMATED HEMATOLOGY ANALYSER ( 3 PART ) ( 1 pieces ) Technical Specifications 1/9 * As per GeM Category Specification Specification Specification Name Values Bid Requirement (Allowed Values) Standards Certifications for the product - * Certification number and date - * Confirmity to quality management standards ISO 9001 & ISO 13485 * Confirmity to electrical safety standards IEC 61010-1
RequiredIEC 61010-2-081
RequiredIEC 61010-2-101 or BIS equivalent * Performance Parameters UMDNS Name and Code "Analyzers
RequiredLaboratory
RequiredHematology
RequiredCell Counting
RequiredAutomated [17-741]
RequiredHematological cell analyser Analyzers
RequiredLaboratory
RequiredHematology
RequiredCell Counting
RequiredSemiautomated [17-742] Analyzers
RequiredPoint-of-Care
RequiredWhole Blood
RequiredHematology [18-513] Computer Aided Detection Systems
RequiredPathology Slide
RequiredBlood Cell Classification [24-248]" * Clinical Purpose Hematology Analyzers are used to count and characterize blood cells for disease detection and monitoring. Basic analyzers return a complete blood count(CBC) with a three-part differential white blood cell(WBC) count. * Type of Configuration Bench top * Type of system offered closed system closed system Type of Automation Fully Automatic * Automatic Start Up
Requiredshut down and sample analysis Yes * Analysis principle Based on principle of counting and sizing * 2/9 Multi channel analysis for better resolution Yes * Type of cell counting 3 part WBC differential * Testing mode selection available CBC Mode * Analysis available WBC
RequiredLymph#
RequiredLymph%
RequiredGranulocytes #
RequiredGranulocytes%Mixed #
RequiredMixed %
RequiredRBC
RequiredHGB
RequiredHCT
RequiredMCV
RequiredPCV
RequiredMCH
RequiredMC HC
RequiredRDW-SD
RequiredRDW-CV
RequiredPL T
RequiredMPV
RequiredPDW-SD
RequiredPDW- CV
RequiredPCT
RequiredPLCR
RequiredP-LCC * Should have three histograms of WBC
RequiredRBC
RequiredPLT Yes * MID% analysis No No MtD# analysis No No Gran% analysis Yes Yes Gran# analysis Yes Yes Analysis method for WBC electrical Impedance electrical Impedance Method for platelet measurement electrical Impedance electrical Impedance RBC Measurement method electrical Impedance electrical Impedance Hb measurement photonetric/direct celluar measurement photonetric/direct celluar measurement Types of modes of running sample Open vial
Requiredcapillary
RequiredClosed
RequiredPre dilute Open vial
Requiredcapillary
RequiredClosed
RequiredPre dilute Maximum sample aspiration volume needed in any of modes should be less than 50 µl should be less than 50 µl Minimum sample volume required should be 10 µl or less should be 10 µl or less Throughput capacity of analyser in (samples/hour) 60-70 60-70 Linearity of Platelet " 0 to 300 * 10³ cells micro litres" * RBC Linearity 0 to 8 x 10^6 per micro litre or more * Hemoglobin linearity 0 to 25 gm per litre 0 to 25 gm per litre WBC linearity " 0 to 3000 * 10³ cells * 3/9 micro litres" Analyses time for cytopenic samples Yes * Directly measures MCV Yes * Time taken by the analyser to produce the test results(Analysis time) in seconds 40-60 * Availability of Auto dilution Yes Yes Types of reagents 1 hemolyzing reagent
Required1 diluent
Required1 cleansing solution * Quality assurance system with calibration and controls Yes * Atleast 3 quality control programs Yes * Type of Calibration both (automatic and manual) both (automatic and manual) Direct aspiration for capillary blood from finger prick Yes Yes Floating discriminator for platelets and RBC counting for reliable RBC and PLT data Yes * Automatic probe wipe Yes * Separate diluting nozzles for RBC and WBC Yes Yes Double bathing mechanism Yes Yes Automatic electric clog removal Yes * Data Management and Display Type of data management Inbuilt system Inbuilt system Display LCD LCD Inbuilt monitor size in inches More than 5 More than 5 PC Monitor size in inches (When PC provided externally) NA (if Inbuilt system) NA (if Inbuilt system) PC hard disk NA NA RAM capacity of PC System NA (if no PC provided) NA (if no PC provided) 4/9 Processor i5 or more * HIS/LIS Interface RS232 RS232 Type of external storage SD Card SD Card No of USB Port 2 2 Data management systems Provide histograms in display and print Provide histograms in display and print Facility for user defined flagging Yes * Type of user Interface or data entry touchscreen touchscreen "Database capability of storing sets of results and graphics" >= 2000 to 5000 >= 2000 to 5000 Facility for workload recording Yes * "Auto stop function in event of unacceptable control data" Yes * Ability to transmit results to host computer Yes * Have auto cleaning function in the analyser’s software Yes * Printer facility Yes Yes Type of printer unit External External Printer type Thermal printer Thermal printer Power Requirements Type of power supply in VAC
RequiredHz 230-240 VAC
Required50-60 Hz * Power Backup facility Yes * Type of UPS Online Online Rating of UPS in KVA 1 KVA 1 KVA Back up time in minutes 30 min 30 min Accessories
Requiredspare parts and consumables "Offered equipment unit to be supplied with sufficient consumables (with at least 2/3rd of total shelf life) required for
Requiredsufficient to carry out haematological 500 500 5/9 testing of samples" Two set of all tubings Yes * Reagent expiry time should be minimum of 1 year Yes * Alerts for operator for level of reagents and to empty waste when indicated Yes * Availability of micro capillary adapter Yes * Operating temperature and humidity Capable of operating continuously in ambient temperature of 15 to 35 deg C and relative humidity of 15 to 85% in ideal circumstances * Miscellaneous Parameters Warranty in years 3 3 Number of installations in Central/State/PSU Govt Hospitals ( Hint: Seller should supply a performace certificate of the device to the buyer incase demaneded after placement of order) 0 0
RequiredLess than 3
RequiredMore than 3 OEM/Reseller ( if supplied by reseller shall ensure uninterputed availabilty of all spares for 10 years) Yes * Availability of toll free facility for technical support maintened by OEM or authorized agencies Yes Yes During the warranty period it shall be ensured that all the break down calls are attended with in 48 hrs and the complete details of service agents/contact details should be furnished to buyer and consignee at Yes * 6/9 time of supplies User/Technical/Mainten ance manuals to be supplied in English in hard and soft copy Yes * Details of equipments and procedures required for local calibration and routine maintenance to be supplied and advanced maintenance task documentation also to be furnished Yes * List of important spares and accessories
Requiredwith their part numbers to be supplied to the buyer at the time of supplying the equipment Yes * Installation and demonstration of equipment and training to be provided after completing supplies before acceptance Yes * The Principal Manufacturer must have direct Presence/approved service center In India Yes * Copies of reports and certifications to be furnished to buyer on demand at time of supplies Yes * Calibration certificates as per NABH requirement Yes * * Specifications highlighted in bold are the Golden Parameters. * Bidders may note that In respect of non-golden Parameters
Requiredthe specifications 'Values' chosen by Buyer will generally be preferred over 'Bid requirement ( allowed Values) by the Buyer. Additional Specification Documents