Garrison Engineer (East) Allahabad
Stanley Road, Teliyarganj, Allahabad
PROVISION OF PIPELINE DELIVERY SYSTEM ALONGWITH SUCTION UNIT FOR 40 BEDS FOR STATIC OXYGEN GENERATOR PLANT (200L/MIN) AT VARANASI CANTT.
2021_MES_446121_1
8470/E8
Open Tender
Civil Works
Item Rate
60 days
Varanasi
As Per NIT
8 documents required · 8 mandatory
₹500
Yes
GE (East) Allahabad
₹50,000
Yes
25 Jun 2021
27 May 2021
18 Jun 2021
27 May 2021
17 Jun 2021
4 Jun 2021
27 May 2021 - 3 Jun 2021
Amount
Supply , installation, commissioning and testing of Semi-Automatic control Panel with maximum input pressure 220 bar, output pressure 4.2 bar, flow capacity 200 litre per minute, to reduce the variable cylinder pressure down to constant line distribution pressure with regulator, safety valve, visual indication of pressure gauge, operating temperatures +10 to +45 degree Centigrade, Digital LCD display, Alarm, sound and as per technical specifications, approved medical standards, complete all as specified and directed and shall be consisting of following features :- (a) Semi-Automatic control system consists of two first stage regulators and consisting high pressure reducer panel, Alarm and control panel, left / right Ramp pressure sensors, pipeline pressure sensor, high pressure input valve, input filter, 315 bar input gauge, high pressure reducer body, pressure set-up screw, high pressure reducer drain, high pressure gauge; second stage regulator, regulator set up, automatic change-over, automatic output pressure gauge in emergency outlet, back up system input, plant output regulators, plant output pressure gauges, plant output connected piping, control / power wiring and all necessary arrangement complete and shall comply Indian Medical Standards and shall comply standard : HTM 02-01/DIN/EN/ISO-7396-1 and shall be CE certified
SL. NO. 1.00 Contd… (b) Both the first stage regulators in the oxygen control panel shall have non-halogenated polymer in the high pressure side to ensure that there will be no ignition due to adiabatic compression (c) 40 micron filter shall be provided at the inlet of each high pressure regulators of the oxygen control panel (d) The first stage regulators shall be connected to a common second stage regulator .The first two regulators meant for first stage shall be capable of switchover system incorporated from “RUNNING” to “RESERVE” bank due to differential pressure (e) The control panel shall provide for two individual content contact pressure gauges to indicate the cylinder pressure in the two wings of the manifold and common pressure gauge to indicate the delivery / line pressure (f) The control panel shall have built-in audio-visual signal lamp indications for bank Changeover (g) The control panel shall be covered with aesthetically suitable cover for safe operation indicating the respective services.
Supply, installation, testing and commissioining of (6+6) Oxygen Manifold System for oxygen gas connections : Main confirming to ISO standards with Middle Frames suitable to withstand a pressure of 40 Kg per sq.m, along with high-pressure copper annealed tail pipes with end brass adapter suitable for Oxygen Cylinders and manifold. Top frame comprising of high pressure copper pipes of size 15 mm inner dia with high pressure brass fittings made of high tensile brass and connections through non- return valves; high pressure copper tail pipes, made of high pressure copper pipe of size 8 mm x 15 swg. The design of middle and bottom frames should be provided to fit both round and flat bottom cylinders safely. The manifold shall be tested (hydraulically) at 3500 PSI.Each cylinder Manifold shall be provided with a terminal header and NPT connection for the semi automatic control panel. Necessary test certificates accompany along with the supply. Only Non Halogenated Polymer materials are to be used in the Non Return Valves supplied along with the manifold., complete all as specified and directed. MAKE:- MAX FLOW / ASB SYSTEM / ASCENT MEDITRONIC / PRIME MEDICAL / LIFELINE MEDISOL
Supply, installation, testing of cylinders set (1+1) emergency supply for oxygen service with high flow regulator isolating the main control panel to ensure uninterrupted oxygen supply, complete all as specified and directed. NOTE:- The quoted rate shall deemed to be inclusive of the provision of cylinder brackets and fixing chains which shall be zinc plated for cylinder placing arrangement. MAKE:- MAX FLOW / ASB SYSTEM / ASCENT MEDITRONIC / PRIME MEDICAL / LIFELINE MEDISOL
Outlet for oxygen service with probe gas specific for the services indicated to accept only compatible quick connect geomatic index type steel adaptors all as per HTM / NFPA standard and as required, complete all as specified and directed. MAKE:- MAX FLOW / ASB SYSTEM / ASCENT MEDITRONIC / PRIME MEDICAL
Supply, installation, testing and commissioning of Oxygen flow meter with mask control range of 0–15 litre per minute (calibration within ± 10%) with 250 ml humidifier bottle made of unbreakable polycarbonate material and autoclavable at 121 degree Centigrade temperature. The flow meter shall be strict precision, durability standard and body shall be made of brass chrome plated materials. The flow tube and shroud components shall be made of clear, impact resistant polycarbonate. Flow Tube shall have large and expanded 0–5 litre per minute range for improved readability at low flows with inlet filter of stainless steel wire mesh to prevent entry of foreign particles, complete all as specified and directed. MAKE:- OXICARE ENGINEERS / GLOBAL MEDLIFE SOLUTION / LIFELINE MEDISOL / AKTIV / LMPL
Supply, installation, testing and commissioning of Medical Vacuum Plant Reciprocating Vacuum 200 litre per minute duplex system, Piston displacement 58 CFM, Two cylinder singl e stage, Air cooled type suitably mounted on MS channel base frame electric driven AC 3 Phase 440 Volts TEFC electric motor 866 rpm with vertical Air receiver tank of capacity 500litre made of mild steel body thickness of 6 mm, disc thickness 8mm, working pressure 10 kg per sq.cm ant Belt Guard and DOL starter and other connected accessories, complete all as specified and directed. NOTE:- Vaccum pump shall be fitted with MS channel frame complete having vibration isolators with 3 Nos.silencers at vacuum pump discharge end. The pump shall be air cooled.3 sets of isolation valve, non-return valve and filter shall be provided with pumps; the system shall be expandable to receive one additional pump in future. The medical vacuum plant shall be CE Marked / ISO compliant and shall be comprising of air cooled, oil lubricated rotary vane / reciprocating vacuum pumps to provide a flow rate. MAKE:- ANESTAIWATA / IR / INDOAIR
Outlets for vaccum service with probe specific for the services indicated to accept only compatible quick connect geomatric index type steel adaptors as HTM / NFPA standard and as required, complete all as specified and directed. MAKE:- MAX FLOW / ASB SYSTEM / ASCENT MEDITRONIC / PRIME MEDICAL
Vaccum jar with rubber tubing complete with regulator capable of controlling vaccum between 0 to 750 mm Hg and polycarbonate autoclavable reusable type vaccum jar of not less than 600 ml, vaccum regulator shall have dial gauge, on / off switch valve control knob, complete all as specified and directed. MAKE:- MAX FLOW / ASB SYSTEM / ASCENT MEDITRONIC / PRIME MEDICAL
Low pressure tube with 93/42/EEC medical devices shall be EN 739 duly colour coated for individual services i.e. white for Oxygen and yellow for Vaccum antistatic rubber tube, complete all as specified and directed. MAKE:- MAX FLOW / ASB SYSTEM / ASCENT MEDITRONIC
Isolation valves 22mm (ball valves) with stainless steel bus and PTFE or Teflon seals; they shall be fitted into the pipeline by means capiliary head soldered joint containing silver; it shall be Open / Closed with help of handle with rotation of 90 degree, complete all as specified and directed.
Isolation valves 28mm (ball valves) with stainless steel bus and PTFE or Teflon seals; they shall be fitted into the pipeline by means capiliary head soldered joint containing silver; it shall be Open / Closed with help of handle with rotation of 90 degree, complete all as specified and directed.
Supply, installation and testing of area alarm panel having capacity of monitoring two gas services of size 300mm x 200mmx 65mm with audio-visual alarm (Master and Local area alarm) provided with gas colour coded gauge or regulator on pipeline side on check wall in the pipeline inlet and test fuction provided in the lix pressure alarm with audio-visual indicators, test button, emergency indicator and beep alarm complete with 220 Volt 50Hz AC input and low voltage transformer less than 24, complete all as specified and directed. MAKE:- MAX FLOW / ASB SYSTEM / ASCENT MEDITRONIC / PRIME MEDICAL
Supply, Instllation, testing and commissiong of Copper Piping 22 mm OD x 0.9 mm thick for medical gases with complete fittings / fixtures as per technical specification and BS: EN 1057/13348, complete all as specified and directed and as mentioned below:- (a) Installation of piping shall be carried out as per approved medical standards with utmost cleanliness (b) Only pipes, fittings and valves which have been degreased as per approved medical standards shall be used (c) Pipe fixing clamps for upto 22mm diameter (d) Pipes shall be non-ferrous suitable for the diameter of the pipe and wherever required rigid metallic hanging or cemented supports to be used. (e) The main lines to the building to be taken overhead through heavy grade metallic poles or GI tubes with inspection points in metallic boxes wherever required . Adequate supports shall be provided while laying pipelines to ensure that the pipes do not sag. Suitable sleeves shall be provided wherever pipes cross through walls / slabs. All pipe clamps shall be non-reactive to copper (f) All pipe joints should be made using inert gas fluxless brazing method. (g) All joints should be of copper to copper and should be brazed by silver brazing filler material without flux while being brazed joints shall be continuously purged with oil free dry nitrogen to prevent the formation of copper oxide on the inside surface of the joint. (h) All pipes should be installed without springing or forcing. (j) All pipes should be protected against mechanical injury in a manner satisfactory to authorities having jurisdiction.
SL. NO. 13.00 Contd… (k) Test: After erection, all the new pipes cleaned or purged with the help of dry nitrogen gas. Complete system should be tested with dry nitrogen at 2 times of working pressure for 24 hours. (l) Painting: All existing and proposed exposed pipes/should be painted with two coats of Synthetic enamel paint & color codification as per IS : 2379 of 1963. and approved medical standards. (m) All concealed pipes to have gas identification bands / labels at appropriate distance. (n) Similarly all pipes which need embedding in the wall should be tested/painted/labeled and properly insulated. (p) Certification: To be certified that pipes are suitable for the particular service and complete cross connection (anti-confusion) test should be carried out. (q) Distribution piping system: MATERIAL (PIPE): Solid drawn, seamless, deoxidized, non arsenical, half hard, tempered and degreased materials conforming to BS: EN 13348. Copper pipe must have reputed third party inspection certificate (E.g.Lloyd‟s or Kite marked or TUV or SGS). Fittings should be made of copper conforming to BS 864 and suitable for a working Pressure of upto 17 bar and especially made forbrazed socket type connections. All copper pipes should be inspected and certified for medical use before dispatch and the pipe should be delivered plugged or capped at both ends. Note: (a) The tenderer shall design & submit complete layout drawings with proper colour code, fittings, fixtures complete for approval of GE or rep of accepting Officer before placing order and the quoted includes for the same.(b) The quoted rate inclusive of all mechanical protections with metallic pipes to pipe wherever required . Make: Mexflow/Janya/ Metal Gem / Cubex .
All as per Sl. No. 13.00 here-in-before but Copper Piping 15 mm OD x 0.9mm thick, complete all as specified and directed.
All as per Sl. No. 13.00 here-in-before but Copper Piping 12 mm OD x 0.7mm thick, complete all as specified and directed.
Bed Head Panel 4 ft long made of high pressure extruded aluminium with powder coating painting and provision of gas / vaccum outlets and 4 Nos 15/5 Amps 240 Volt AC electrical socket outlet with individual switch control, complete all as specified and directed. MAKE:- MAX FLOW / ASB SYSTEM / ASCENT MEDITRONIC
NOTE:- (APPLICABLE FOR SL. NO. 1.00 TO 16.00 HERE-IN-BEFORE) The quoted rate shall deemed to be inclusive of the following :- (a) Contractor shall provide comprehensive on-site warranty (Including all spares, accessories and labour) for a period of 2 years from the date of final acceptance of the complete system after successful and complete installation and commissioning with regular updation of newer technology as and when evolved. (b) Contractor shall install, depute trained manpower to maintain manifold and medical gas pipeline system in case of any breakdowns within 24hours. The technical maintenance personnel of the contractor shall be responsible for supervision and maintenance shall be available to reach the site(s) within 1 hour notice. (c) If the performance of any individual equipment or system is not satisfactory, the same shall be replaced by the contractor free of cost. (d) If it is found that to meet the performance criteria, any extra equipment is required the same shall be provided free of cost by the contractor. (e) All faults appearing and their rectification shall be periodically advised to the hospital, the period being not more than a month. (f) Any lacuna or lacunae noticed in the functioning of the installation as a result of any design feature shall be rectified by the contractor free of cost.
NOTE:- (CONTD...) (g) Contractor shall fully associate the engineers and technicians for MH Varanasi during installation, testing, commissioning, operation and maintenance period. The testing of the entire system shall be carried out in the presence of IEM from MES and Officer rep (Min 02 Nos.) from Hospital authorities. (h) Contractor shall train the staff of MH Varanasi the operation of complete medical gas and oxygen plant system. (j) Contractor shall submit MoU with authorised Medical Service engineer / registered Bio-medical engineer services experienced in the relevant field for not less than 5 years immediately before placing work order. (k) All electrical wiring / cables, cable tray, earthing, Control panels, connected electrical & pipeline distribution system with all connected accessories of complete medical gas system and oxygen plant system shall also deemed to be included in the quoted rates.
Excavating in trenches, not exceeding 1.5 m wide and not exceeding 1.5metre in depth; for foundation, etc. or for shafts, wells, cesspits, manholes, pier holes, tubular poles or laying of pipes, or for removal / laying of cables etc., not exceeding 10 sq m on plan and not exceeding 1.5 m in depth and getting out in soft / loose soil including dressing / trimming sides and levelling / grading and ramming of bottoms of excavation, complete all as specified and directed.
Returning, filling in, in soils, including spreading, levelling, watering and well ramming in layers not exceeding 25cm., complete all as specified and directed.
Removing surplus excavated soils to a distance not exceeding 50metre and depositing where directed at a level not exceeding 1.50m above the starting point, complete all as specified and directed.
Spreading of natural, clean and dry river sand in trenches for cable protection, complete all as specified and directed. NOTE:- Punned thickness shall only be measured and paid.
Supply and lay burnt clay fly ash bricks class 'B', old size, designation 75, size 230mm x 115mm x 75mm, laid in trenches for cable protection with length of brick equal to width of trench, complete all as specified and directed.
Steel tubing, light grade, ISI marked, galvanised including short lengths, laid in trenches / fixed to walls / ceilings or laid in floors / laid in ducts / laid under road crossing for cables, complete with all fittings (such as bends, elbows, Tee, short pieces, connectors, back nuts, diminishing pieces, caps, plugs and unions etc.,), 80mm bore of pipe, complete all as specified and directed. NOTE:- Necessary excavation and earthwork for taking down old and unserviceable GI pipe shall be measured and paid separately under respective items of this Schedule.
LED street light luminaires 90 Watt with pressure aluminium die- cast powder coating housing and frame with heat resistant, high efficiency toughened glass with EPDM gasket for IP-66 grade protection, complete assembly with LED, driver and accessories pre-wired up to terminal block including connection on pole or building including taking down of old / unserviceble street light fitting, complete all as specified and directed. MAKE:- Cat No. LSTO- 90- CDL of Crompton Greaves / Cat No. LHETAXU7IN6J090 of Havells / Cat No. LTESS40X2.5HPLED/LS of C&S
XLPE insulated, screened, PVC bedded, galvansied steel strip or wire armoured heavy duty electric cable with stranded aluminium conductor, 1100 volts grade, ISI marked as per IS-7098, cross-sectional area 25 sq.mm 4 core, laid in trenches / on poles / walls including provision of suitable hardware accessories, connecting up, testing on completion and commissioning including taking down existing unserviceable cable of any type and depositing the retrieved materials in MES storeyard Varanasi, complete all as specified and directed. NOTE:- Necessary excavation and earthwork for taking down old and unserviceable cable of any type, sand cushioning to cables and bricks as cable protection shall be measured and paid separately under respective items of this Schedule.
XLPE insulated, screened, PVC bedded, galvansied steel strip or wire armoured heavy duty electric cable with stranded aluminium conductor, 1100 volts grade, ISI marked as per IS-7098, cross-sectional area 70 sq.mm 3.5 core, laid in trenches / on poles / walls including provision of suitable hardware accessories, connecting up, testing on completion and commissioning including taking down existing unserviceable cable of any type and depositing the retrieved materials in MES storeyard Varanasi, complete all as specified and directed. NOTE:- Necessary excavation and earthwork for taking down old and unserviceable cable of any type, sand cushioning to cables and bricks as cable protection shall be measured and paid separately under respective items of this Schedule.
XLPE insulated, screened, PVC bedded, galvansied steel strip or wire armoured heavy duty electric cable with stranded aluminium conductor, 1100 volts grade, ISI marked as per IS-7098, cross-sectional area 25 sq.mm 4 core, laid in trenches / on poles / walls including provision of suitable hardware accessories, connecting up, testing on completion and commissioning including taking down existing unserviceable cable of any type and depositing the retrieved materials in MES storeyard Varanasi, complete all as specified and directed. NOTE:- Necessary excavation and earthwork for taking down old and unserviceable cable of any type, sand cushioning to cables and bricks as cable protection shall be measured and paid separately under respective items of this Schedule.
XLPE insulated, screened, PVC bedded, galvansied steel strip or wire armoured heavy duty electric cable with stranded aluminium conductor, 1100 volts grade, ISI marked as per IS-7098, cross-sectional area 10 sq.mm 2 core, laid in trenches / on poles / walls including provision of suitable hardware accessories, connecting up, testing on completion and commissioning including taking down existing unserviceable cable of any type and depositing the retrieved materials in MES storeyard Varanasi, complete all as specified and directed. NOTE:- Necessary excavation and earthwork for taking down old and unserviceable cable of any type, sand cushioning to cables and bricks as cable protection shall be measured and paid separately under respective items of this Schedule.
Earthing with GI earth plate electrode 600mm x 600mm x 6.3mm thick buried directly in ground vertically to a depth of not less than 2.25m deep below ground level with the top edge of earth plate at a depth of not less than 1.5m below ground level connected to GI earth strip of 25mm x 4mm size by means of bolts, nuts, washers of galvanised iron or steel, GI earth strip shall be protected by 40mm dia GI pipe 'light grade', watering pipe 20mm dia GI pipe 'medium grade', GI wire meshed funnel, alternate layer of charcoal and salt on all side to earth plate upto a thickness of 150mm, filling of black cotton soil in earth pit upto GL, and connected to earth plate including necessary excavation & earthwork, PCC (1:2:4) type B-1 earth chamber, 40mm thick precast RCC cover (1:2:4) type B1 mix with 6mm dia nominal reinforcement @ 100mm c/c both ways etc complete all as shown in electrical plate No.3 connected to earthing test point, including testing on completion, complete all as speciified and as directed. NOTE:- Rate quoted shall deemed to include cost of excavation in any type of strata, RFR and removal upto 50metre, PCC pit, funnel and RCC cover plate. (b) Earth continuity conductor and protection pipe beyond 7.5metre shall be measured and paid under respective item of this Schedule.
Earth continuity conductor or main earthing lead fixed to wall on batten or recess or chases or buried in ground or drawn in conduit/pipe or fixed to poles or any other indicated situation for loop earthing etc. as required of size 25mm x 4mm galvanised iron strip, complete all as specified and directed.
Steel tubing, light grade, ISI marked, galvanised including short lengths, laid in trenches / fixed to walls / ceilings or laid in floors / laid in ducts / laid under road crossing for cables, complete with all fittings (such as bends, elbows, Tee, short pieces, connectors, back nuts, diminishing pieces, caps, plugs and unions etc.,), 40mm bore of pipe, complete all as specified and directed.
Supplying, installation, testing and commissioning of factory made feeder pillar box, 415 Volt, with synchroscope 4-wire, AC 50 Hz, of size 1000mm x 750mm x 600mm, made out of 3.15mm thick CRCA sheet, weather and vermin proof powder coating finish both interior and exterior side, for outdoor mouting with having degree of protection as per IP-43 with locking arrangements, two earth terminal suitable cable gland for incoming and outgoing and removable neutral link, connecting strip / solid coductor brought out to cable chamber brass / copper nut, bolts, washers etc., including grouting of feeder pillar box 600mm above ground level including provision of suible size angle iron stand made up of 50mm x 50mm x 6mm welded on all the four corner embedded in PCC 1:2:4 type B1 (using 20mm graded stone aggregate) upto 300mm above ground including provision of necessary earthwork and excavation, complete with internal wiring, complete all as specified and directed with and including provision of the following :-
SL. NO. 31.00 Contd... (a) Busbar 200A cap (R,Y,B) for incoming and out going cable (b) Busbar 100A cap neutral for outging cable (c) MCCB, FP, 400 Amp, 415 Volt, 50kA breaking capacity, 50 Hz with thermal magnetic release - 01 (d) MCCB, FP, 250 Amp, 415 Volt, 35kA breaking capacity, 50 Hz with thermal magnetic release - 01 (e) MCCB, FP, 100 Amp, 415 Volt, 35kA breaking capacity, 50 Hz with thermal magnetic release - 01 (f) RCCB two pole 63 Amp rating with sensitivity of 30mA suitable for 240 Volt - 04 Nos (g) Voltmeter, Ammeter, frequency meter, load sharing relay and other accessories with internal wiring to withstand upto 160kVA
M/s Rajeev Kumar
M/S UJJAWAL ENTERPRISES
PANDEY TRADERS
stage.html
html • 0.05 MB
tech_bid_open.pdf
tech_eval.pdf
fin_bid_open.pdf
boq_comp_chart.xlsx
xlsx
fin_eval.pdf
aoc.pdf
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.03 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .