RequiredExperience with Government Bid to RA enabled No Fully Automatic Bio Chemistry Analyzer ( 1 pieces ) Technical Specifications * As per GeM Category Specification Specification Specification Name Values Bid Requirement (Allowed Values) Standards Conformity to Standards CE CE Certification No and Date - * Performance Parameters Type of system Fully Open Random Access system * 1/8 Purpose Should be capable of all routine
RequiredSTAT and special biochemical tests including specific proteins
Requiredtherapeutic drugs(TDM)
RequiredDrugs of abuse
RequiredImmunoturbidemtric assays and user definable applications in blood
Requiredserum
Requiredbody fluids
Requiredurine and equipment to be configurable as per requirements of consignee * Model type of Biochemistry Analyzer Bench top Type Bench top Type Type of samples serum
Requiredplasma
Requiredurine
RequiredCSF and any body fluid samples * Type of Assay End point
RequiredRate
RequiredKinetic turbidometric and bi chromatic assay * Throughput tests hour Photometric Tests with ISE module 200 or more 200 or more Throughput tests/hour Photometric Tests without ISE Module 150 or more 150 or more Calibration facility Linear (one
Requiredtwo and m ultipoint)
RequiredExponential
RequiredL ogit-log
RequiredNon linear Linear (one
Requiredtwo and multipoint)
RequiredExponential
RequiredLogit-log
RequiredNon linear Equipment must have self diagnostic tests with error message and online display Yes * Equipment must be programmable for all test menus and state of the art work station should be upgradable by addition of ISE module Yes * Type of sample loading Sample tray Sample tray Equipment should have both internal and Yes * 2/8 external probe cleaning/washing capacity Repeat facility repeat facility to be available for calibrator and control * Reagent refill message and monitoring facility Yes * Minimum number of test parameters on board at a time 36 * Minimum sample capacity at a time hint: Seller may select the value based on the number of samples that can be tested at a time and if it is any value betweem 35-49 select option of 35-49 and for other ranges also similar choices may be made 35 - 49 35 - 49 Minimum number of STAT samples at a time 28 * Out of range samples Should have pre post and auto dilution of samples and return capability for out of range samples * Probe dispensers Probe dispensors must have level detectors and must typically use between 2-25 µl of sample * Spectoral range in nm 340-700 * Wave length selection method Interference filter(With minimum 8 interference filter) Interference filter(With minimum 8 interference filter) Light source Halogen lamp low cost with long life * Life of lamp in hrs 10000 * Power consumption of lamp in VA 1000 * Type of sample cups Reusable Reusable Number of sample cups provided with machine 1000 1000 3/8 QC program QC Programme with L-J graphs
RequiredPrint out of reports * The reaction cuvette must be reusable and durable Yes * System should have on board cooling system Refrigerated * DI water plant for the washing of the cuvettes/to avoid carry over and contamination Yes Yes Compatible on line UPS Yes Yes Back up time 1 hour 1 hour Data management software Equipment to be supplied with compatible programmable windows based comprehensive data processing and management system graphical user interface software
RequiredLM Capability complete back up of data base for calibration control and patient sample results
Requiredcapability for Bi- directional interfacing with LIS * Patient result storage capacity 10000 10000 Onboard cuvette washing & drying facility Yes Yes Photometeric Checking of cuvettes before next reaction - Create Another field Yes * System should have facility for reading results on monitor and print out facility Yes * Reagents to be supplied with machine Creatinine
Requiredurea
Requireduric acid
RequiredTriglycerides
Requiredsugar
RequiredCholesterol
RequiredTP
RequiredCalcium albumin
RequiredSGOT
RequiredSGPT
Requiredalkaline Creatinine
Requiredurea
Requireduric acid
RequiredTriglycerides
Requiredsugar
RequiredCholesterol
RequiredTP
RequiredCalcium albumin
RequiredSGOT
RequiredSGPT
Requiredalkaline 4/8 phosphatase
RequiredBilirubin (Total & Direct) calibrator 5 ml - 4 and quality control 5ml -1 Nos each for normal and abnormal phosphatase
RequiredBilirubin (Total & Direct) calibrator 5 ml - 4 and quality control 5ml -1 Nos each for normal and abnormal Reports Prinout of reports and full patient demographics to be available * System to be supplied with necessary pre requisites and start up kits normal and abnormal QC and calibrators Yes * Availability of RS 232 port/USB for data transfer Yes Yes Power supply 200-240V AC 50Hz Single phase * Reagents position More than 30 * Bar code reader for reagents No No Bar code reader for samples No No Refrigerator supplied with Analyzer No No Capacity of Refrigerator in L Not provided Not provided Star rating Not provided Not provided Type of Micro pipette Supplied 10 to 100 microlitres variable micropipettes and 100 to 1000 microlitres variable micro pipette one each 10 to 100 microlitres variable micropipettes and 100 to 1000 microlitres variable micro pipette one each Variable Micro pipette Supplied Yes Yes Seller must provide original documentary proof of the date and place of manufacturing of supplied equipment Yes * Service manual and operating manual to be provided Yes * Warranty in years (including machine
Required3 3 5/8 DI Water Plant if provided
RequiredUPS
Requiredcomputer and printer) Additional Parameters Sytem shall be supplied with a compatible desk top PC No No The PC shall be of reputed brand and name of brand to be specified - * Speed of processor in GHZ NA (No PC is supplied) NA (No PC is supplied) RAM in MB NA (No PC is supplied) NA (No PC is supplied) Hard disk drive capacity in GB NA (No PC is supplied) NA (No PC is supplied) Monitor Not applicable Not applicable Monitor size in inches Not applicable Not applicable Compatible operating system provided in PC Not applicable Not applicable Printer provided No No Type of printer Not applicable Not applicable The unit shall be capable of operating continuously in ambient temperature and relative humidity "Temp of 10 - 40deg C and humidity of 15-90%" * Startup kit
RequiredCalibrators
Requiredconsumables required in training the staff/Technician/Doctor s should be provided at the time of installation &Training Yes * The reaction cuvettes should be replaced at free of cost when ever required during the warranty period Yes * Reports ISO Certification of Manufacturer ISO 13485 * ISO certification number and date - * Availability of test report from central govt/NABL/ILAC No No 6/8 accredited lab covering all parameters Test report number if test report not available put NA - * Test report date if test report not available put NA - * Name of test lab if test report not available put NA - * Address of test lab if test report not available put NA - * Copies of all certifications and reports to be provided to buyer on demand at time of supplies Yes Yes * Specifications highlighted in bold are the Golden Parameters. * Bidders may note that In respect of non-golden Parameters
Requiredthe specifications 'Values' chosen by Buyer will generally be preferred over 'Bid requirement ( allowed Values) by the Buyer.