RequiredExperience with Government
RequiredAnnual Turnover
RequiredCertificate (Requested in ATC)
RequiredOEM Authorization Certificate
RequiredOEM Annual Turnover Bid to RA enabled Yes Microwave Based Disinfection System (SAMEER Technology) ( 4 pieces ) Technical Specifications * As per GeM Category Specification Specification Specification Name Values Bid Requirement (Allowed Values) Performance Parameters Technology Microwave technology based disinfection of hazardous biomedical waste designed developed and Microwave technology based disinfection of hazardous biomedical waste designed developed and 1/7 patented by SAMEER Mumbai patented by SAMEER Mumbai Nominal capacity of Disinfection system in litres per cycle 60 60 Microwave out put power in kw 3 3 Microwave frequency in MHz 2450 * Power consumption in watts hour (Approximate) 4600 4600 Duration of one disinfection cycle in minutes 45-50 * Microwave generator provided Magnetrons strategically positioned and tuned to specific frequency to ensure total coverage and disinfection * Material of Construction of external body MS powder coated external body with SS microwave applicator aith microwave leak proof front cover * Inner microwave chamber construction Specially designed rectangular chamber made of SS 304 Grade stainless steel * Monitoring On line temperature display along with printing of process cycle using printer to record process parameters .On line monitoring of process status and magnetron current * Report Generation inbuilt printer to record process start time
Requiredend time and temperature log of the process along with weight of the hospital waste processed * Turn table Precisely balanced slow speed turn table with concealed drive mechanism to ensure 360 degree uiform microwave heating and * 2/7 with weight sensor Waste container Microwave proof moulded container to hold the medical waste * Exhaust System Specially desigend closed loop water ejector system for the microwave chamber to avoid discharge of any fumes/gases to the atmosphere * Automatic moistening System internally installed automatic water spray mechanism to facilitate moistening of the waste for uniformity of microwave absorption and heating * Control System High end programmable logic control system to ensure total automatic disinfection cycle with minimum human intervention and single button operation * Safety interlocks Insufficient hospital waste
Requiredclosure of applicator door
Requiredhigh voltage safety door
Requiredmagnetron cooling blower magnetron temperature and current etc * Dimensions in mm x mm x mm 1700 X 1050 X 630 1700 X 1050 X 630 Weight in kg 300 200
Required250
Required300 Remote monitoring of machine status No Yes
RequiredNo Print out of weight processed by system Yes Yes Parameters for which disinfection efficiency to be tested Disinfection efficacy 10⁵ to 10⁷ log
Requireddecontamination of bacterial suspension
RequiredD econtamination of mycobacterium tuberc ulosis(MTB)
RequiredDecontami nation of fungal suspen sion
RequiredDecontamination of spores of bacillus * 3/7 atrphaeus
Requiredinactivation of cell lines
RequiredDestruction of vaccines eginfluenza live attunuated.
RequiredDestruction of viruses eg HIN1 and HIVI
RequiredDestruction of blood vials
RequiredDestruction of m
Requiredaggots
RequiredDecontami nation of operation thestre monitoring swabs Technology acquisition From Society of applied mocrowave electronics engineering and research Mumbai * Operating voltage and frequency 230 ± 10 VAC
Required50 Hz * Warranty in years 5 5 Certifications Certification Report Efficincy certification as per CPCB Biomedical waste rule * Certification Lab Half kine institute
RequiredParel Mumbai * Certification No and Date - * Manufacturing LICENCE(TOT Agreement) issued by SAMEER and its validity - * Details of certifications with manufacturer(TOT holder) such as ISO - * Copies of all certifications and reports to be provided to buyer on demand at time of supplies Yes * * Bold specifications are the golden parameters. * Bidders may note that In respect of non-golden Parameters
Requiredthe specifications 'Values' chosen by Buyer will generally be preferred over 'Bid requirement ( allowed Values) by the Buyer.