Tender Value
₹82 L
EMD Value
₹1.6 L
Closing Date
17 Jan 2022, 6:00 pmClosed
Executive Engineer M and H, Dn. I Jaipur
Executive Engineer M and H, Dn. I Jaipur
SITC of Oxygen Generation plant at CHC Deedwana, Distt. Dausa and CHC Jalsu, Distt. Jaipur
2022_MEDIC_254275_1
NIT 106/2021-22 EE M and H Dn. I Jaipur/
Open Tender
Electrical Works
Percentage
60 days
Jaipur
Please refer Tender documents.
3 documents required · 3 mandatory
₹1,000
RTGS/NEFT/Any other money Transfer System
₹1.6 L
27 Jan 2022
12 Jan 2022
18 Jan 2022
12 Jan 2022
17 Jan 2022
12 Jan 2022
Amount
SITC of Pressure Swing Absorption Oxygen Generator Plant capacity 150 to 160 LPM or 353 SCFH equal to 35 (Approx. )Jumbo Cylinders per day with stand alone two compressor and two dryers working in tendem As per ISO 13485 and as per following Specifications:
Purity of Oxygen 93%± 3% @4.1 bar pressure
Rotary Screw type air compressor of suaitable capacity ( 2 Nos) as per afore mention capacity 400 V /3 Phase /50 HZ ,Outlet pressure 5BAR,Maximum Output Pressure 10 BAR, Power Supply 110-240 V, 1 phase , 50 Hz , UPS Module for control only with medical upgrade kit , Coal Tower , carbon filter , Bacteria/ Sterile .
The unit should comprise four stage air filtration , consists of moisture seperator , 0.1 micron pre filter before the Refrigerated air dryer , another two filters 0.1 and 1 micron after the dryer then a carbon tower which makes sure 0 oil and vapour in the down stream air. The carbon should last for more than two years . and it should be provided 1 micron filter after carbon tower. Beneath this filter a moisture alarm should be incorporated to switch off the PSA system in case of dryer auto drain failure .
This unit should also comprise of 1nos air receiver 750 Litres , oxygen /Buffer tanks 750 Ltrs as per standards, an in-line flow meter , certifying the capacity shall be installed at the outlet of the oxygen receiver
In case PSA Oxygen Generator failure , an auto switch over system to regular supply for oxygen is incorporated so that automatically back up from cylinders shall be taken over. To show this effect an alarm system should be installed with this system.
All the system piping and fittings , all the interconnections , and piping up to manifold room shall be of stainless steel .
This system should be comprising of OFF -Spec Valve to prevent the supply of impure oxygen to go into the system if the purity goes to below the 90% .
The Voltage Stabilizer of suitable capacity ( Minimum 60 KVA Rating )and standerd make as per manufacturer guideline should be provided.
The Unit can be monitor on line through by service team /authorised person 24/7 on no extra cost
The size of the display screen (Touch screen digital Display) 7” to 10 “
The Makes of PSA Plant should be ISO:13485/CE/US FDA certified.
The bidder shall have to be submit test results from NABL notified third party inspection after installation and commissioning of PSA Plant.
SITC of MGPS pipe of medically grade as per BSEN-13348 with third party certification from Lloyds/TUV degreased (ie free from oil, grease and particulate matters) seamless copper pipe confirming to ASTM B88 and all copper joining done by brazing with AWSA5.8, classification BCuP-5 or BS 1845 following procedure in accordance with ASTM B828 after purging with nitrogen. All pipe cutting shall be carried out with a radial pipe cutter and copper pipe fitting's shall comply with B4 alloy designation. Adequate support shall be provided while laying pipeline testing shall be done at 17 Bar and after testing pipe shall be painted with two coat of enamel paint as per colour codification as per ISO 7396-Part-I 2016. Third party test report shall be submitted. All as per pre approved by Engineer in charge. For additional technical parameters of products/work, refer Annexure "A" attached with P.W.D. Electrical BSR 2021
SITC of non lubricated ball type Nickel Coated Isolation Valve with adapter & PTFE Seat having quarter turn/Butterfly handle opening of following sizes suitable for flow capacity of required LPM as per the size of valve at 4.7 bar with maximum allowable pressure of 30 bar and suitable for working temperature ranging - 20°C to 130°C. Appropriate Test certificate/report of product/accessories from OEM shall be submitted. All as per pre approved by Engineer in charge. For additional technical parameters of products/work, refer Annexure "A" attached with PWD Electrical BSR 2021
FOR 22 MM OD Pipe
FOR 15 MM OD Pipe
Neometrix Engineering (P) Limited
Dhawan Box Sheet Containers Pvt Ltd
JK ENGINEERING WORKS
Med Freshe Pvt. Ltd.
UNISSI INDIA PVT LTD
Dhawan Box Sheet Containers Pvt Ltd
Neometrix Engineering (P) Limited
JK ENGINEERING WORKS
UNISSI INDIA PVT LTD
Med Freshe Pvt. Ltd.
stage.html
html • 0.04 MB
tech_bid_open.pdf
tech_eval.pdf
fin_bid_open.pdf
boq_comp_chart.xlsx
xlsx
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.04 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .