| 1 | Course particle dust dispersion system for particles | Technical Specification
Feed rate should be less than 10 milometers per hr,
Potentiometer with variable speed,
Maximum feed rate be 300 millimetres per hr,
Mass flows of system for compact density be 1 gram per cubic centimetre,
Volume flow rate should be 0.5 to 5 cubic meter per hr,
Course particle material in bulks and non cohesive powders,
Dosing time is continuous max 10 hrs,
Particles type for tracer is LDA, PIV, etc.,
Total particle number concentration is Ca. 10000000 particles per cubic centimetre,
Required Mass flow of particles is from 0.04 to 185 grams per hr for compact density of 1 gram per cubic centimetre,
Required Particle size range is from 0.1 to 100 micrometres,
Required Carrier or dispersion gas is Air, Nitrogen gas or Oxygen gas,
Required Inlet pressure is 4 to 8 bar,
Required Counter pressure is 0 to 0.2 barg,
Required Feed rate is 1 to 300 millimetres per hour,
Required Air connection is Quick coupling,
Required Inner diameter of sink is 7 mm, 10 mm, 14 mm, 20 mm, 28 mm,
Required Height of filling is 70 mm, Required Dispersion cover is Type A, Type B, Type C, and Type D,
Required Outlet connection for aerosol are should be next four Type A, Type B, Type C, and Type D given
Dispersion cover type A should be Inner diameter equal to 5 mm, Outer diameter equal to 8mm,
Dispersion cover type B should be Inner diameter equal to 3.6 mm, Outer diameter equal to 6 mm,
Dispersion cover type should be Inner diameter equal to 2.5 mm, Outer diameter equal to 6 mm,
Required Power supply is from 115 to 230 Volts, 50 or 60 Hz,
Required Filling quantity is 2.7 g at reservoir diameter equal to 7 mm, 5.5 g at reservoir diameter equal to 10 mm, 10.8 g at reservoir diameter equal to 14 mm, 22 g at reservoir diameter equal to 20 mm, 43 g at reservoir diameter equal to 28 mm,
Required Warranty is 1 years,
Required Dimensions are 465 by 320 by 200 millimetres for H by W by D,
The approximate weight of dust dispersion system can be up to 19 kg | 1 | set | MANORANJAN SAHU Aerosol Generator For Solid 400076,ESED ,IIT Bombay, Powai, Mumbai-400076 K Cl Or Na Cl Particles Up | 30 |
| 2 | Aerosol generator for solid KCl or NaCl particles up to 10 �m | Technical Specifications for aerosol generator for solid KCl or NaCl particles up to 10 �m areRequired Number of course particles of KCl or NaCl particles is up to 10 �m with KCl,
Required Reproducibility is with regard to particle size and particle concentration,
Required Air volume flows is 20 litre per minutes up to 600 litre per minutes,
Required Components and dispersion nozzle material is plastic (POM),
Required with Simple handling, robust, durable and, low maintenance,
Required Particle discharge is bipolar path must be necessary,
Required Test standards is EN 779 to ISO 16890 for air filters,
Required Stable and reproducible atomization of NaCl and other salt solutions should be integrated,
Required Particle count is at least 500 counts per size interval (as per ISO 16890),
Required Connection to MFP system is Direct, Flat filter media considering ISO 16890 is Fraction separation efficiency,
Required for Laboratory equipment for the generation of salt aerosols,
Required Warranty is 1 year for aerosol generator.
Required Number of course particles of KCl or NaCl particles is up to 10 �m with KCl,
Required Reproducibility is with regard to particle size and particle concentration,
Required Air volume flows is 20 litre per minutes up to 600 litre per minutes,
Required Components and dispersion nozzle material is plastic (POM),
Required with Simple handling, robust, durable and, low maintenance,
Required Particle discharge is bipolar path must be necessary,
Required Test standards is EN 779 to ISO 16890 for air filters,
Required Stable and reproducible atomization of NaCl and other salt solutions should be integrated,
Required Particle count is at least 500 counts per size interval (as per ISO 16890),
Required Connection to MFP system is Direct, Flat filter media considering ISO 16890 is Fraction separation efficiency,
Required for Laboratory equipment for the generation of salt aerosols,
Required Warranty is 1 year for aerosol generator. | 1 | set | MANORANJAN SAHU 400076,ESED ,IIT Bombay, Powai, Mumbai-400076 | 30 |