RequiredOEM Authorization Certificate Bid to RA enabled Yes Fourier Transform Infra Red (FTIR) Spectrometer ( 1 pieces ) Technical Specifications * As per GeM Category Specification Specification Specification Name Values Bid Requirement (Allowed Values) OPTICAL PERFORMANCE Type of Optical System Single beam * Wavenumber- Maximum (Per cm) 7800 7500
Required7800
Required8000 Wavenumber - Minimum (Per cm) 350 350
Required375 Spectral (Wavenumber) Resolution (Per cm) 0.25 0.07
Required0.25
Required0.5
Required0.8 Spectral (Wave Number) Accuracy 0.05 0.05 1/7 (Per cm) (+/-) Interferometer Michelson interferometer * Wavelength setting and scanning Automatic * Beam Splitter KBr KBr Automatic beam splitter changeover for MIR to NIR or MIR to FIR without manual intervention Yes Yes Mirror Coating Aluminium Aluminium
RequiredGold Peak to Peak Signal to Noise Ratio for 1 minute scan 50000:1 or more 50000:1 or more Measurement modes Transmittance
RequiredAbsorbance & Reflactance * DETECTOR Detector Type Deuterated Triglycine Sulphate (DTGS) Deuterated Triglycine Sulphate (DTGS) Detector having temperature control No Yes
RequiredNo LIGHT SOURCE Light Source High intensity Ceramic light source High intensity Ceramic light source SAMPLE TYPE ANALYSIS FACILITY Analyze Liquid samples Yes Yes Analyze Solid samples Yes Yes Analyze Powder samples Yes Yes Analyze thin films on metals samples Yes Yes Analyze Gas samples Yes Yes SOFTWARE software - inclusive in the scope of supply Latest version of windows based IR Software (compatible with Windows 7/8/10/latest) with License to control all fuctions of the FTIR system like facility for self-diagnostic
Requiredsoftware controlled interferometer auto tuning
Requiredauto alignment
Requiredauto component recognition
Requiredand auto optimization data collection
Requireddata * 2/7 processing
Requiredlive data disply
Requiredspectral quality check
Requiredspectral search from commercial library
RequiredSpectral match
RequiredCurve fitting
Requirednormalization
RequiredPeak Area
RequiredPeak Height calculations
Requiredcreation of user generated library
Requiredautomatic atmospheric correction
Requiredspectral interpretation etc. Free updates Yes * Customizable Report Template Yes Yes Application Note for the instrument in the intended application field (May provide URL & other reference details) - * FTIR LIBRARIES FTIR Georgian Crime Library - inclusive in the scope of supply Yes Yes FTIR Canadian Forensic Library - inclusive in the scope of supply Yes Yes FTIR Sadler Drug Library - inclusive in the scope of supply Yes Yes FTIR Polymers & Polymer additives Library
Requirednclusive in the scope of supply Yes Yes FTIR Surfactants Library
Requiredinclusive in the scope of supply Yes Yes FTIR Minerals & inorganic CompoundsLibrary
Requiredinclusive in the scope of supply Yes Yes FTIR Organic Compounds Library
Requiredinclusive in the scope of supply Yes Yes FTIR Drugs Library
Requiredinclusive in the scope of supply Yes Yes 3/7 FTIR Adhesives & Sealants Library
Requiredinclusive in the scope of supply Yes Yes FTIR Cosmetics & Personal products Library
Requiredinclusive in the scope of supply Yes Yes ACCESSORIES Suitable Hydraulic Press 15 Ton for pellet/sample preparation - inclusive in the scope of supply Yes Yes Suitable Mortar and pestle - inclusive in the scope of supply Yes Yes Suitable Dies - inclusive in the scope of supply Yes Yes Suitable KBr set for Pellets - inclusive in the scope of supply Yes Yes KBr Powder - inclusive in the scope of supply Yes Yes Suitable Pellet holder - inclusive in the scope of supply Yes Yes Types of Suitable ATR - inclusive in the scope of supply Yes Yes ATR crystal - inclusive in the scope of supply Diamond Diamond Suitable Universal liquid cell holder - inclusive in the scope of supply Yes Yes Suitable Dry Box - inclusive in the scope of supply Yes Yes Suitable Replacement dessicant (rechargeable) - inclusive in the scope of supply Yes Yes Suitable Liquid Cell Window - inclusive in Yes Yes 4/7 the scope of supply Liquid Cell Window material KBr KBr Shape of the Cell Window Rectangular Rectangular Suitable Gas cell - inclusive in the scope of supply Yes Yes Validation Kit - inclusive in the scope of supply Yes * GENERIC Sample Chamber Dimensions (mm x mm x mm) - * FTIR and Sample compartment having provision for purging with gas Yes Yes FTIR provided with in-built air drier for humidity and vapour protection Yes Yes Display "Through PC (PC is not part of supply)" * Connectivity USB USB Power Supply 230 V ± 10
Required50 Hz * Overall Size (Dimensions) (mm x mm x mm) - * Weight 39 * Any other features/accesories covered in scope of supply - * List of Items and Quantity of each item included in the offer - * OPERATING CONDITIONS Minimum Operating Temperature 20 20.0 Maximum Operating Temperature 40 40.0 Operating Humidity (RH) (%) at 40 degree C 65 65.0 WARRANTY AVAILABILITY OF SPARES/TRAINING Warranty on equipment 1 1.0 Warranty on Source 10 10.0 5/7 Warranty on interferometer 10 10.0 Warranty on Laser 10 10.0 Availability of Spares and Consumables for 7 years from the date of installation (Not inclusive in the scope of supply) Yes * Onsite training 5 5.0 CERTIFICATION Certifications ISO
RequiredGLP
RequiredGMP
RequiredCE * Availability of UL/CE Certification as per EN 61010-1:2010 Safety requirements for electrical equipment for measurement
Requiredcontrol
Requiredand laboratory use General requirements Yes Yes Conformance to EMC/EMI as per EN 61326-1:2013 Electrical equipment for measurement
Requiredcontrol and laboratory use Yes Yes Availability of Test Reports from Central Govt/NABL approved/ILAC accredited Lab to prove conformity to the specification Yes Yes If Yes
RequiredTest Report to be furnished to the Buyer on demand Yes * * Bold specifications are the golden parameters. * Bidders may note that In respect of non-golden Parameters
Requiredthe specifications 'Values' chosen by Buyer will generally be preferred over 'Bid requirement ( allowed Values) by the Buyer.