Loading…
Loading…
| # | Company | Amount | Rank | Status |
|---|---|---|---|---|
| 1 | L-1₹6.8 LAccepted-AOC | L-1 | Accepted-AOC ok | |
| 2 | Not Admitted-Fee/PreQual/Technical | - | Not Admitted-Fee/PreQual/Technical Not Complied with technical specification |
Tender Value
₹8.7 L
EMD Value
₹17,400
Closing Date
16 Nov 2021, 1:00 pmClosed
EE PHED No. 2
Jhajjar
Recalled tender for Procurement of equipments in PHE District Water Testing Laboratory Jhajjar, Supply of UV-VIS dual beam spectrophotometer bench top and BOD Incubator at District PHE laboratory Jhajjar Town and all other works contingent thereto.
2021_HRY_192102_2
SE JJR No. 8073 dt. 08/10/2021 Proc. Lab equipment
Open Tender
Civil Works
Supply
30 days
Jhajjar
Please refer Tender documents.
2 documents required · 2 mandatory
₹1,000
₹17,400
Yes
6 Jan 2022
10 Nov 2021
16 Nov 2021
10 Nov 2021
16 Nov 2021
10 Nov 2021
Amount
BOD Incubator Make:- NST, Ambassdor, optics Technology Inner Size:18"×18"×24" Rating Watts 750 No. of Shelves:3 Capacity: 120 Ltrs Construction: Double walled construction outer body made of Mild steel powder coated & inner Body made of Stainless- Steel Insulation: Glass wool insulation between two walls. Temp range: room temperature to 600C Accuracy:+/- 0.50C Temperature controlled by: Controlled by PID Micro process controller with dual display Selves: With G.I. wire Mesh selves Door: 2 door inner acrylic door to view sample & outer metal door with air circulation motor with blower for uniform temperature throughout chamber Heater: Special Nichrome hearter designed for long life & saving of power Consumption
UV-VIS dual beam Spectrophotometer Make:- Thermo Fischer Scientific, Hach, Lovibond, WTW ( Xylem), ELICO The Spectrophotometer will be dual beam (viz. UV- Visible) operation and with the reference cell position for kinetics application. It will mandatorily have an 8-cell carousel. The spectrophotometer must have room- light resistance, allowing lid-open operation for improved speed and convenience The Spectrophotometer will have option of automated cell changers and cell holders, sipper, and fiber optic probes for measurements outside of the sample compartment. It must accommodate high throughput measurement and temperature control. The Spectrophotometer will have following features: i. Spectrophotometer will be Double Beam(UV- Visible)/ Reference Cell ii. It will have high resolution touch-screen user interface. The spectrophotometer must be capable of operating through computer also iii. There will be 8- position cell changer with switchable turret. iv. It will have onboard methods, reporting and data export v. It is capable of open-lid operation vi. There will be room light resistance vii. It will have sipper (optional). viii. It must have optional facility for 4- position cell changer ( long path cells) and Peltier- thermostat single cell holder(200C- 600C). Specification: Detector Type Dual Silicon Photodiodes Display 7- inch colour touch-screen, high definition, 800×1280 pixels Photometric Accuracy instrument ±0.002A at 0.5A ±0.004A at 1.0A ±0.008A at 2.0A Wavelength Accuracy ±0.5nm Baseline Flatness ±0.002A Compartments 8-position cell changer (standard) Drift (At 500nm after 1-hour warm up) <0.0005A/Hr Keypad Touch-screen Lamp Xenon Flash Lamp (>5 years typical, 3 years guaranteed) Noise (RMS at 500 nm 60 consecutive measurements) ≤0.00020A at 0A at 260 and 500nm ≤0.00030A at 1A at 260 and 500nm ≤0.00040A at 2A at 260 and 500nm Optical Design Double Beam Photometric Display-3A to +5A Photometric range -2A to +3.5A Photometric Repeatability ±0.001A at 1A measured at 1.0A at 546nm Spectral Bandwidth 2 nm Stray Light <1.0%T 198nm (KCl), <0.05%T at 220nm (NaI), <0.03% at 340nm(NaNO2) Wavelength Data Interval 0.1nm, 0.2nm, 0.5nm, 1nm, 2nm, 5nm Wavelength Range 190nm-1100nm Wavelength Repeatability <±0.2 nm Wavelength Scan Speed Slow, medium, and fast- up to 1600nm/min Computer operation: The spectrophotometer can be operated from compatible computer and software Computer and operational software to run spectrophotometer will be a part of the spectrophotometer Connections Singles USB-A supports flash memory devices for method and data storage Warranty: 3 years Calibration Standard for iron, nitrate, sulphate, fluoride Reagent for iron, nitrate, sulphate, fluoride
stage.html
html • 0.04 MB
tech_bid_open.pdf
tech_eval.pdf
fin_bid_open.pdf
boq_comp_chart.xlsx
xlsx
fin_eval.pdf
aoc.pdf
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.03 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .