Sturdy & light weight machine
Compatibility with mains 220-240 V (normal), 50/60 Hz power supply
ECG recording with 12 leads a. Standard Leads (the limb leads or bipolar Limb leads: I, II & III)
Built-in ECG auto-measurement including: HR, P-R interval, P-Duration, QRS duration, Q-T interval, Q-TcF (Friedericia), P Axis, QRS Axis, T Axis, R(V5), S(V1), R(V5)+S(V1) -QTcF interval reading/ measurement should also be available with Limb leads alone .
-High-resolution thermal printing array system -Built in printer should work with standard universal thermal printer paper
he machine should be supplied with power cord , patient cable, user manual and warranty card, Operation Manual with user demonstration video CD, interpretation manual & 10 thermal recording paper rolls , 5 bottles of jelly,
Should be provided with terminal for good earth connection to preclude electric disturbances while recording- -Must have a safety certificate or valid detailed electrical and functional safety test report from a recognised competent authority -Copy of the certificate / test report shall be produced along with the technical bid.
The firm should install the instrument at the designated location and provide one-day training/ demonstration of operation of ECG machine.
- performance warranty of at-least one year from date of installation + additional two years comprehensive warranty, -In case of breakdown of the machine, the supplier shall make the machine functional by repair (including replacement of parts) free of cost at the user site, within 3 (three) days of the receipt of complaint, or replace the machine (if necessary).
-The suppliers should have adequate after sales service facilities covering all districts of the country. -They should have infrastructure and trained manpower to attend to any complaints within 3 days of receipt of complaints
2260Lx1060Wx500H mm
Single mold sturdy stable base stand with built in light & regulator, inclined Binocular body 30 °, 360º rotatable head with focus adjustment controls. A durable textured acid resistant finish. All optical parts including objectives, eye pieces and prisms should have anti- reflective coating which also gives anti- fungal property. All metallic parts should be corrosion-proof, acid proof and stainproof.
Optical system should be infinity corrected. Built-in LED light source with whitelight with intensity control and LED life of more than 10, 000 Hrs
The system should have a build-in variable light source (Illuminator). This light source should have a 20 W,6 V LED lamp. The system should be provided with a step down transformer and an on-offswitch and intensity control. The lamp should be provided with a lamp socket which has the facility foreasy replacement of the bulb.
Highest quality 10 X/20mm wide angle anti fungus field eyepiece. One with pointer. Diopter adjustment must be present on both eye pieces. (the image of the object as seen through the binocular eyepiece should be well defined centrally in at least 2/3 field of view) Achromatic, wide field, 10 x with inbuilt pointer. The eyepiece should be aplanatic and have a minimum field number of 18 Diopter adjustment must be present on one / both eye pieces or on the eye piecetube.
Backward tilted revolving nose piecesuitable to accommodate four objectives with click stop. It should be provided with rubber ribbed grip for easy rotation mounted on a precision ball bearing mechanismfor smooth and accurate alignment.
Stage uniformly horizontal, mechanical stage having dimensionsof length 140 mm (+/- 20mm) with fine Vernier graduations (minimum reading accuracy of 0.1 mm). It should be designed with convenient sub-stage vertical coaxial adjustment for slide manipulation. The stage should have ball-bearing arrangement to allow smooth travel intransverse directions i.e. 80 mm (+/- 5mm) and front to back direction, 50mm (+/- 5mm).
Coaxial coarse and fine focusing knob, capable of smooth, fine focusing movement sensitivity; minimum: 300 micron; focusing stop for slide safety.
Abbe-type condenser with numericalaperture (N.A.) 1.25 focusable with rack and pinion arrangement incorporating a spherical lens and aniris-diaphragm
Voltage 220 V AC, 50Hz. shouldhave one on-off power switch
Capable of operating continuously inambient temperature of 10 to 50 ° C and relative humidity of 15 to 90% in ideal circumstances. Capable of being stored continuously in ambient temperature of 0 to 50 °C and relativehumidity of 15 to 90%
Should be provided with accessories like working manual, immersion oil, lens tissue paper, lens cleaning solution, one antistatic cleaning brush etc
Equipment shall operate on 230 V, single phase, 50 Hz electric supply. The necessary protective relaying / circuitry shall be there with the machines. The mains supply voltage variation may be max.±10% and frequency variation maximum ±3 %
Microscope should be ISI marked as per IS:4381:2017 with valid BIS license / CE certified under IVD / Valid CDSCO manufacturing license. In case of CE (Other/General IVD, following documents are required. a. a. Declaration of conformity by manufacturer or EU representative of Manufacturer for the quoted model b. b. Documentary evidence regarding firm registered with EEA (European Economic Area) competent authority is required.OR European Representative registered with EEA (European Economic Area) competent authority appointed by firm is required. OR Other documents like certificates from notified body along with declaration of conformity. In case of imported equipment, CDSCO import license is required along with valid USFDA/CE certification documents as described above.
Three-year comprehensive warranty to be followed by 7 years CMC. Technical support, required spares and consumables should be assured for two years after initial 3 + 7years period is over.
The equipment should be provided with one hard copy in original of the detailed service manual and operation manual. Further, a soft copy is also required.
The equipment must be tropicalized as below: Operating room temperature: upto 40° C Storage room temperature: upto 60º C Relative Humidity: upto 90% non-condensing
Among the other things, the responsiveness of the bid will be based on successful demonstration of the offered model of the equipment
The bidder has to submit users list with address & contact telephone number/s.
Prospective tenderers should have a full-fledged and wellestablished service centre in Mumbai with engineers qualified in servicing of binocular microscope.
Description
Laser Jet Tank
22 PPM or higher
22 PPM Black
(PCL5 and 6, PS3 emulation) or PDL or equivalent
250 Sheets Input Tray
At least 100 Sheets
Built-in Automatic
USB 2.0 with cable; 1 Ethernet 10/100T network
Windows 10(32/64bit), Linux