Executive Engineer Mech
SKIMS Soura
Repairing, providing and fitting of Automatic Medical gas Manifold panels alongwith allied items in Main Hospital of SKIMS
2022_SKIMS_180439_1
e-NIT NO. 70 OF 2022-23 Dated 22.07.2022
Open Tender
Machineries/ Mechanical Engg Items
Item Rate
20 days
SKIMS Soura
Please refer Tender documents.
8 documents required · 8 mandatory
₹500
Yes
As Described in SBD
₹12,000
Yes
13 Aug 2022
22 Jul 2022
1 Aug 2022
22 Jul 2022
30 Jul 2022
23 Jul 2022
Amount
Repairing oxygen manifold of ward block of Main Hospital by way of providing and fitting of New automatic Medical gas manifold panel (Fully Automatic) for twin bank type (15+15) configuration with auto regulation, 02 No by-pass emergency kits/points regulator, lead (high pressure pipes),gauges, interconnecting copper piping, tail pipes with NRV for cylinders, Isolation control valve on the supply line including MS support and allied items. The manifold control panel shall be capable of regulating minimum 1500 LPM Oxygen flow and shall be designed with standing the oxygen cylinder pressure in the range of 0-200kgf/cm2 with safety features related to hazard and continuity of flow of medical grade oxygen. The panel shall operate at a constant pressure through two banks of bottled gas cylinders with following status and indication system. (i)Bank in use :- Green Indication (ii)Bank ready for use :- yellow indication (iii).Empty Bank :- Red indication The Control panel shall have built-in LED display to indicate normal, high and low line-pressure. The Control panel shall incorporate safety provision in case of failure of electronic as well as pneumatic system. Control should be able to be used manually through by pass system. The mounting angle iron of 50x50x5mm shall be fixed on the wall with clamps/dash fasteners to hold the 15+15 cylinder system in place. The job includes dismantling, removing of old panel and cleaning of surface. The job includes cleaning of exiting old panel and stacking same at same place as per instruction of site Engineer. All wiring, thimbling, piping, valves etc required for successfully installation testing and commissioning of the battery bank are to be provided by the contractor. The equipment shall conform to the relevant ISO/BIS standard and contractor or firm shall have to submit relevant test certificates.
Repairing Medical gas manifold ( Nitrous oxide) of main hospital by way of providing and fitting of New automatic Medical gas manifold panel (Fully Automatic) suitable for line pressure range 0 to 10kgf/ cm2, twin bank type (10+10) configuration with auto regulation, 02 No by-pass emergency kits/points regulator, lead (high pressure pipes),gauges, interconnecting copper piping, tail pipes with NRV for cylinders, Isolation control valve on the supply line including MS support and allied items. The manifold control panel shall be capable of regulating minimum required Nitrous oxide, flow and shall be designed to bear the cylinder pressure of 0-200kgf/cm2 with safety features related to hazard and continuity of flow of medical grade Nitruous oxide . The panel shall operate at a constant pressure through two banks of bottled gas cylinders with following status monitoring and indication system. (i)Bank in use :- Green Indication (ii)Bank ready for use :- yellow indication (iii).Empty Bank :- Red indication The Control panel shall have built-in LED display to indicate normal, high and low line-pressure. The Control panel shall incorporate safety provision in case of failure of electronic as well as pneumatic system. Control should be able to be used manually through by pass system. The mounting angle iron of 50x50x5mm shall be fixed on the wall with clamps/dash fasteners to hold the 15+15 cylinder system in place. The job includes dismantling, removing of old panel and cleaning of surface. The job includes cleaning of exiting old panel and stacking same at same place as per instruction of site Engineer. All wiring, thimbling, piping, valves etc required for successfully installation testing and commissioning of the battery bank are to be provided by the contractor. The equipment shall conform to the relevant ISO/BIS standard and contractor or firm shall have to submit relevant test certificates.
Ajay Electro Medical
amba steel & metal udyog
stage.html
html • 0.03 MB
tech_bid_open.pdf
tech_eval.pdf
fin_bid_open.pdf
boq_comp_chart.xlsx
xlsx
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.03 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .