Supply, installation, testing and commissioning of Lab Equipment High Performance Liquid Chromatography - HPLC
2026_DSTE_171139_1
PBTI/PROC/26-27/08/07_01
Open Tender
Laboratory and scientific equipment
150 days
PBTI, Mohali
Please refer Tender documents.
2 documents required · 2 mandatory
₹2,360
Yes
₹1.1 L
PBTI, Mohali
2 Jul 2026
2 Jul 2026
23 Jul 2026
2 Jul 2026
23 Jul 2026
9 Jul 2026
8 Jul 2026
PUNJAB BIOTECHNOLOGY INCUBATOR
COMMERCIAL BID FORMAT - HPLC with PDA, RID & FLD
Name of the Manufacturer
Make of the Equipment
Country of Origin
Name of the Bidder
Part A - Main Equipment Qty Rate in Rs. GST% Amount in Rs.
For Honey application as per National/International requirements 1 set
High pressure Liquid Chromatography (HPLC) equipped with Quaternary gradient solvent delivery system with capability of operating detectors. The system should be quoted with all accessories required to make it fully operational and any other item required for stated applications be quoted as optional.
Technical Specification:
Quaternary Gradient with Online Degasser.
Pressure operating range 7000 psi or better
Flow Rate Range: Programmable 0.01 to 5 ml/ min in 0.01 ml/min increments
Flow Precision ±0.1% RSD or below
Flow Rate Accuracy ±1%
Eluent Degassing Online Degasser for all channels
Gradient Mixer Quaternary mixing & gradient capability using proportionate valve
Solvent Setting Range 4 solvents setting range:0-100% with 0.1% step
Diagnostic Features Error detection and display, Leak detection & safe leak handling
Wavelength range 190-700 nm
Linearity range Linearity range <5%
Baseline noise 3.0 X 10-5 AU at 254 nm or better
Drift <1.0x 10-3 AU/h at 254 nm
Suitable peak purity software, Auto threshold for peak purity
Refractive Index Range 1.00 to 1.75 RIU
Noise Level ± 2.5X10-9 RIU
Drift Drift 2x10-7 RIU/hr
Cell Volume Approximately 10 μl or better
Temperature Control Temp controlled Flow cell unit
Light Source Xe lamp with standard wattage
Wavelength Range 200-600nm (Excitation)
Wavelength Accuracy ±3nm
Wavelength Reproducibility/ Repeatability ±0.25nm
Sensitivity S/N Raman peak of water ≥ 1000 or S/N >550 ASTM
Flow Cell volume approximately 8 μl or more
Wavelength Scanning Scanning of Excitation and Emission wavelengths
Pressure 1 Mpa or better
5 COLUMN COMPARTMENT
Temp. range Ambient to 80° C or better
For column length 300 mm
No. of Columns accommodated Minimum 2
Temperature Stability ±0.3 ºC of set temperature
Injection Volume Range 0.1-100μl (Standard)
Sample Capacity >80 x 1.5/2 ml vials or more
Injection Volume Accuracy ±1%
Injection Precision <0.5% RSD or better
Carry over 0.005% from previous injection
Tray Temperature Operating Range 4 - 40°C programmable in 1°C increments
7 SOFTWARE AND HARDWARE
Software: Software with multitasking and capable of performing the following functions: Control the system, acquire, store, process and reproduce the data. It must be able to control all the devices from same software Factory certified and/or Branded Computer System with i7/ Xenon or better processor, 32 GB RAM or better and 2TB HDD or better, 22-23” or larger Full HD Monitor with licensed latest windows compatible with equipment Software, MS office Latest standard version and Antivirus for 3 year. System should have minimum 2 network ports. Printer – Laser B/W - duplex - Legal, A4 with Network Card and Bluetooth/Wifi facility. 1 set
8 DOCUMENTS AND TRAININGS
On site IQ, OQ of instrument along with document
Training Satisfactory technical and application training to the personnel at site immediately after installation.
Total Amount of Part A
PART B - Other Requisites for HPLC
HPLC COLUMNS C18=300 x 4 (id) mm Carbohydrate or equivalent - 2 no./Set 1 Set
C8 = 150 X 4.6 X 5μm - 2 no./Set 1 Set
Cyano = 150 X 4.6 X 5μm 1 no. 1 No.
Amino = 250 X 4.6 X 5μm - 2 no./Set 1 Set
Phenyl = 250 X 4.6 X 5μm 1 no. 1 No.
Silica = 250 X 4.6 X 5μm 1 no. 1 Set
· Fittings, Frits, ferules and Tubing’s 1 Set
· Tubing cutter - 02 no./Set 1 Set
· Solvent bottles 1000ml capacity - 04 No./Set 1 Set
· Solvent filters (Glass & SS both, 04 no. each) 1 Set
Total Amount of Part B
PART C - Extended Warranty Charges (After expiry of Standard Warranty of One (01) Year
1st Year Extended Warranty Charges
2nd Year Extended Warranty Charges
3rd Year Extended Warranty Charges
4th Year Extended Warranty Charges
5th Year Extended Warranty Charges
6th Year Extended Warranty Charges
Total Amount of Part C
Total FOR Amount upto PBTI Site - Part A + Part B + Part C
Total Amount in Words:
ü It is mandatory to quote item wise rates for Part B & Part C, lump sum rates are not acceptable for these.
ü Bidder must quote amount in the given cells, if amount against any cell is included in main component, bidder should put "Zero" value.
ü Taxes must be mentioned, if it is not mentioned, the same shall be treated as included in given rates.
Signatory Name:
HPLC - List of any other items like consumables required for trouble free operation of equipment for 06 years after the expiry of guarantee/warranty period which are not covered in CMC
Sr. No. Name Qty. Rate
Remarks, if any:-
Name of Company
Name of Authorzed Signatory
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.03 MB
Tendernotice_1.pdf
PDF • 0.31 MB
e-tender-HPLC.pdf
Tender Documents • 0.92 MB
PriceBidFormatHPLC.xls
Additional Documents • 0.06 MB
Download all tender documents and submit your bid