RequiredExperience with Government
RequiredAnnual Turnover
RequiredCertificate (Requested in ATC)
RequiredOEM Authorization Certificate
RequiredOEM Annual Turnover Bid to RA enabled No C ARM FLUOROSCOPE X RAY MACHINE ( 2 pieces ) Technical Specifications * As per GeM Category Specification Specification Specification Name Values Bid Requirement (Allowed Values) Standards Conformity to Certifications CDSCO/AERB European CE
RequiredCDSCO/AERB Certification Number and Date - * Type test approval from AERB required Yes * Type test certificate date and number - * 1/6 Conformity to Safety Standards IEC 60601-2-54 or BIS equivalent * Performance Parameters Purpose C arms fluoroscopes are used in a variety of general surgical
Requiredcardiac
Requiredand neurological applications
Requiredincluding aneurysm repair
Requiredpacemaker implantation
Requiredhip replacement
Requiredfracture reduction
Requiredforeign-body location
Requiredneedle biopsy
Requiredcatheter placement
Requiredpain management
Requiredper- cutaneous lithotripsy
Requiredand brachytherapy * UMDNS Name and Code 1. Fluoroscopic Units
RequiredPortable - 18-434 2. Radio graphic/Fluoroscopic Units
RequiredMobile - 11-758 3. Radio graphic/Fluoroscopic Units
RequiredExtremity - 18-435 * The System should be microprocessor controlled
Requiredcompact and mobile offering excellent Image details Yes * Type of Mechanical Motion of C Arm Motorized Motorized Horizontal movement Of C Arm in mm 190 * Vertical Movement Of C Arm in mm 410 * Angulation Movement of C Arm in degrees ± 180 ± 180 Radius (Immersion Depth)of C arm in mm 650 * Source to Image Distance in mm 1000-1050 1000-1050 Arc Orbital Movement of C Arm At least +90° - 25° At least +90° - 25° Total width of C Arm in mm( Free space of C arm) 650 * 2/6 Panning Movement of C Arm ( Swivel movement) ±12.5º ±12.5º Safety Feature available for C Arm Movement a. Locks for all the movements. b. Control Stand foot lock. c. Steering wheel for easy steering & movement * Range of Frequency of X ray Generator in khz 41-50 41-50 Generator Power output in Kw 5 5 Fluoroscopy Range in kvp 40-120 40-120 Fluoroscopy mA range
Requiredmaximum in mA 0.2 to 7 0.2 to 7 Dual focus Rotating Anode X-Ray Tube of focal spot 0.3mm² (small) & 0.6mm² (large) 0.3mm² (small) & 0.6mm² (large) Anode heat storage capacity with tolerance ±10% 100 KHU 100 KHU Modes Available Continuous fluoro
RequiredPulsed Fluoro
RequiredRadiographic Mode * Pulses per sec 1 to 5 1 to 5 Mode selection for Kv and mA Manual Manual Type of Image Intensifier 9 " triple Field 9 " triple Field Number of Monitors provided on intensifier system 2 (One for LIH and one for memory display ) 2 (One for LIH and one for memory display ) Type of monitor to be provided with intensifier system LCD LCD Size of Monitors in inch 17-20 17-20 Type of Collimator controlled Fixed Collimation Slot Fixed Collimation Slot Type of Camera CCD Camera CCD Camera Camera Resolution in pixels 0.5K x 0.5K 0.5K x 0.5K Acquisition of Camera 14 bits 14 bits 3/6 Contrast ratio of intensifier unit 10:1 to 30:1 * Image storage capacity in images 20000 to 50000 20000 to 50000 Image reversal Yes * Zoom function available 2x 2x Monitor with trolley Yes * Sterile covers for C- Arm/X-Ray tube/Image Intensifier Yes * DICOM Compatibility No No The system should be ready to connect with HIS/PACS Yes Yes Data transfer Both USB and CD?DVD Both USB and CD?DVD Lead Aprons - Full body and Skirt type (0.5pb) 4 number each * No of Thyroid shield and Gonald Shield 4 number each * Radiation/Fuoro Exposure delivery Switch Hand Switch and Foot Switch * Display of mA
RequiredKV and other safety features on C - arm Console Yes * Radiography through Cassette with holder Yes Yes Size of Cassette holder provided (for radiographic mode) in inch x inch 10 x 12 10 x 12 Patient demographic detail entry facility Yes * Upgradable option to digital substration of angigraphy and peripheral angiography Yes Yes "Automatic image parameter selection with capability of switching on to manual selection" Yes * Operating temperature & humidity 0 to 25 deg C and relative humidity of 15-90% * 4/6 Type of Power Supply Single phase
Required220- 240 Vac
Required50Hz-16 Amps Socket * UPS Availability External External UPS backup time in minutes 15 15 Servo Stabilizer supplied No No Rating of Servo Stabilizer Not provided Not provided User/Technical/Mainten ance manuals to be supplied in English in hard and soft copy Yes * Installation
RequiredDemonstration and training to be provided at consignee end Yes * The Principal Manufacturer Must Have Direct Presence/approved service center In India Yes * Warranty in Years 5 * QA
RequiredQC and other tests as per NABH shall be carried out by OEM or authorized agency of OEM Yes * Reports Confirmity to Manufacturing Standards ISO 13485 ISO 13485 Certification number and date - * Copies of all calibration and test reports and certifications to be submitted at the time of Supply of Machine Yes * * Bold specifications are the golden parameters. * Bidders may note that In respect of non-golden Parameters
Requiredthe specifications 'Values' chosen by Buyer will generally be preferred over 'Bid requirement ( allowed Values) by the Buyer.