Loading…
Loading…
| # | Company | Amount | Rank | Status |
|---|---|---|---|---|
| 1 | 1₹8.3 LAccepted-AOC OFFICE NO 3 CSC 5 1ST FLOOR GOLE MARKET ROHINI NORTH WEST DELHI DELHI 110085 | NORTH WEST DELHI | DELHI | 110085 | ₹8.3 L | 1 | Accepted-AOC lowest |
| 2 | 2₹13.5 L+₹5.2 L (62.7%)Rejected-AOC | ₹13.5 L+₹5.2 L (62.7%) | 2 | Rejected-AOC higher |
| 3 | 3₹15 L+₹6.7 L (80.7%)Rejected-AOC | ₹15 L+₹6.7 L (80.7%) | 3 | Rejected-AOC higher |
| Sl No | Description | Qty | Unit | RAMPHAL PHOGAT CONT L2 | Gennext Lab Technologies Pvt Ltd L1 | SUNIL KUMAR CONTT L3 |
|---|---|---|---|---|---|---|
| 1.00NS Item | ||||||
| 1.10 | `The Spectrophotometer will be dual beam (viz. UV-Visible) operation and with the reference cell position for kinetics application. It will mandatorily have an 8-cell carousel. The spectrophotometer must have room-light resistance, allowing lid-open operation for improved speed and convenience.
The Spectrophotometer will have option of automated cell changers and cell holders, sipper, and fiber optic probes for measurements outside of the sample compartment. It must accommodate high throughput measurements and temperature control.
The Spectrophotometer will have following features:
i. Spectrophotometer will be Double Beam (UV-Visible)/Reference Cell
ii. It is will have High resolution touchscreen user interface. The spectrophotometer must be capable of operating through computer also.
iii. There will be 8-position cell changer with switchable turret
iv. It will have onboard methods, reporting and data export
v. It is capable of open-lid operation.
vi. There will be room light resistance.
vii. It will have sipper(optional)
viii. It must have optional facility for 4-position cell changer (long path cells) and Peltier-thermostat single cell holder (20 °C - 60 °C)
Specifications:
Detector Type Dual Silicon Photodiodes
Display 7-inch colour touchscreen, high definition, 800 × 1280 pixels
Photometric Accuracy Instrument ±0.002A at 0.5A
±0.004A at 1.0A
±0.008A at 2.0A
Wavelength Accuracy ±0.5 nm
Baseline Flatness ±0.002A
Compartments 8-position cell changer (standard)
Drift (At 500 nm after 1-hour warmup) (0.0005A/Hr
Keypad Touchscreen
Lamp Xenon Flash Lamp ()5 years typical, 3 years guaranteed)
Noise (RMS at 500 nm 60 consecutive measurements)
=0.00020A at 0A at 260 and 500 nm
=0.00030A at 1A at 260 and 500 nm
=0.00040A at 2A at 260 and 500 nm
Optical Design Double Beam
Photometric Display -3A to plus 5A
Photometric Range –2A to plus 3.5A
Photometric Repeatability ±0.001A at 1A measured at 1.0A at 546 nm
Spectral Bandwidth 2 nm
Stray Light ( 1.0%T 198 nm (KCl), (0.05%T at 220 nm (NaI), (0.03%T at 340 nm (NaNO2)
Wavelength Data Interval 0.1 nm, 0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm
Wavelength Range 190 nm – 1100 nm
Wavelength Repeatability ( ±0.2 nm
Wavelength Scan Speed Slow, medium, and fast - up to 1600 nm/min
Computer operation: The spectrophotometer can be operated from compatible computer and software. Computer and operational software to run Spectrophotometer will be a part of the spectrophotometer.
Connections Single USB-A supports flash memory devices for method and data storage
Warranty 3 years | 1 | Each | 13,50,000 ₹13,50,000 | 8,30,000 ₹8,30,000 Lowest | 15,00,000 ₹15,00,000 |
Tender Value
₹7.5 L
EMD Value
₹15,000
Closing Date
11 Apr 2026, 12:00 pmClosed
SOHAN LAL
Charkhi Dadri PHED
supply installation commissioning and training of Pre programmed water quality analysis spectrophotometer Make HACH
2026_HRY_513000_1
2026BC36114D 74C1 410E 955C 73D89D201E96555PUH
Open Tender
Civil Works
Works
365 days
Charkhi Dadri
2 documents required · 2 mandatory
₹1,000
₹15,000
Yes
8 Jul 2026
6 Apr 2026
11 Apr 2026
6 Apr 2026
11 Apr 2026
6 Apr 2026
| Sl No | Description | Qty | Unit | Est. Rate | Est. Amount |
|---|---|---|---|---|---|
| 1NS Item | |||||
| 1.1 | `The Spectrophotometer will be dual beam (viz. UV-Visible) operation and with the reference cell position for kinetics application. It will mandatorily have an 8-cell carousel. The spectrophotometer must have room-light resistance, allowing lid-open operation for improved speed and convenience.
The Spectrophotometer will have option of automated cell changers and cell holders, sipper, and fiber optic probes for measurements outside of the sample compartment. It must accommodate high throughput measurements and temperature control.
The Spectrophotometer will have following features:
i. Spectrophotometer will be Double Beam (UV-Visible)/Reference Cell
ii. It is will have High resolution touchscreen user interface. The spectrophotometer must be capable of operating through computer also.
iii. There will be 8-position cell changer with switchable turret
iv. It will have onboard methods, reporting and data export
v. It is capable of open-lid operation.
vi. There will be room light resistance.
vii. It will have sipper(optional)
viii. It must have optional facility for 4-position cell changer (long path cells) and Peltier-thermostat single cell holder (20 °C - 60 °C)
Specifications:
Detector Type Dual Silicon Photodiodes
Display 7-inch colour touchscreen, high definition, 800 × 1280 pixels
Photometric Accuracy Instrument ±0.002A at 0.5A
±0.004A at 1.0A
±0.008A at 2.0A
Wavelength Accuracy ±0.5 nm
Baseline Flatness ±0.002A
Compartments 8-position cell changer (standard)
Drift (At 500 nm after 1-hour warmup) (0.0005A/Hr
Keypad Touchscreen
Lamp Xenon Flash Lamp ()5 years typical, 3 years guaranteed)
Noise (RMS at 500 nm 60 consecutive measurements)
=0.00020A at 0A at 260 and 500 nm
=0.00030A at 1A at 260 and 500 nm
=0.00040A at 2A at 260 and 500 nm
Optical Design Double Beam
Photometric Display -3A to plus 5A
Photometric Range –2A to plus 3.5A
Photometric Repeatability ±0.001A at 1A measured at 1.0A at 546 nm
Spectral Bandwidth 2 nm
Stray Light ( 1.0%T 198 nm (KCl), (0.05%T at 220 nm (NaI), (0.03%T at 340 nm (NaNO2)
Wavelength Data Interval 0.1 nm, 0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm
Wavelength Range 190 nm – 1100 nm
Wavelength Repeatability ( ±0.2 nm
Wavelength Scan Speed Slow, medium, and fast - up to 1600 nm/min
Computer operation: The spectrophotometer can be operated from compatible computer and software. Computer and operational software to run Spectrophotometer will be a part of the spectrophotometer.
Connections Single USB-A supports flash memory devices for method and data storage
Warranty 3 years | 1 | Each | - | - |
stage.html
html • 0.05 MB
tech_bid_open.pdf
tech_eval.pdf
fin_bid_open.pdf
boq_comp_chart.xlsx
xlsx
fin_eval.pdf
aoc.pdf
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.03 MB
Tendernotice_1.pdf
PDF • 0.69 MB
BOQ_591817.xls
BOQ • 0.23 MB
AD0112112.pdf
Additional Documents • 0.69 MB
OT0112112.pdf
Other Document • 0.69 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .