Loading…
Loading…
| # | Company | Amount | Rank | Status |
|---|---|---|---|---|
| 1 | L1₹48.9 LAccepted-AOC | L1 | Accepted-AOC Being L1 and found in order | |
| 2 | L2₹51.0 L+₹2.0 L (4.11%)Rejected-AOC C O SECTOR 02 HOUSING COLONY CHANNI HIMMAT JAMMU J K PIN 180015 | JAMMU | JAMMU AND KASHMIR | 180015 | L2 | Rejected-AOC Not being L1 | |
| 3 | L3₹55.8 L+₹6.8 L (13.9%)Rejected-AOC 99 MODEL TOWN PATHANKOT 145001 | PATHANKOT | PATHANKOT | PUNJAB | 145001 | L3 | Rejected-AOC Not being L1 | |
| 4 | L5₹59.1 L+₹10.2 L (20.8%)Rejected-AOC 29 MODEL TOWN BEHIND GURUDWARA PATHANKOT | PATHANKOT | PATHANKOT | PUNJAB | 145001 | L5 | Rejected-AOC Not being L1 | |
| 5 | L6₹77.2 L+₹28.3 L (57.8%)Rejected-AOC C O OPPOSITE MH MANDA POONCH ROAD AKHNOOR 181201 J K | JAMMU | JAMMU AND KASHMIR | 181201 | L6 | Rejected-AOC Not being L1 |
Tender Value
₹49.9 L
EMD Value
₹99,700
Closing Date
17 Aug 2021, 6:00 pmClosed
HQ 135 WE
HQ 135 WE C/O 56 APO
PROVN OF AUGMENTATION OF RESOURCES IN COVID ICU HDU AND WARDS/DEPARTMENTS FOR MILITARY HOSPITAL UNDER GE(N) AKHNOOR UNDER PARA 35 OF DWP 2020
2021_MES_463287_1
CWE/JA-20/2021-22
Open Tender
Civil Works
Item Rate
90 days
AKHNOOR
Class D Cat b i
8 documents required · 8 mandatory
₹500
GE NORTH AKHNOOR C/O 56 APO
₹99,700
3 Sept 2021
28 Jul 2021
19 Aug 2021
28 Jul 2021
17 Aug 2021
4 Aug 2021
28 Jul 2021 - 3 Aug 2021
Amount
Supply, installation, testing and commissioning of Horizontal Bed head Panel with provision of light with on off switch & electrical connection and 4 Nos of 5/15 Amps & 04 Nos Switches , head panel should be minimum 1200mm with rail, and provision for 2 Nos of Gas outlet ( 01 Nos of oxygen outlets and 01 No of vaccum outlet) safe efficient means delivering services to patients staff in both general and special care applications, It is modular design made of horizontal aluminium extructions and ability to house medical gas terminal unit and electrical socket , smooth curve surfaces , no visible screws and choice of color decor stripe, ease of installation via separate rail bracket or wall mounting plates easy removal of covers for maintenance and pipeline connection, integral twin medical rail as part of the modular extrusion design. It has segregation of services ie extra low voltage, normal voltage and medical gases, should be maintained throughout. Its aluminium extruction have a polyester coating finish on all external surfaces. It have provision of nurse call, data or monitoring sockets should be made at the point of manufacturing. It should be supplied duly pre-piped, prewired and fully tested or in carcass, complete all as per directed by Engineer-in-Charge. Make: Oxicareengineerings/ Global medlife solution/Aktiv/Life line Mediso/Tedisel / Unissi or equivalent Make approved by HQ CENC.
Supply, installation, testing and commissioning of Horizontal Bed head Panel with provision of light with on off switch & electrical connection and 4 Nos of 5/15 Amps & 04 Nos Switches , head panel should be minimum 1200mm with rail, and provision for 4 Nos of Gas outlet ( 02 Nos of oxygen outlets and 02 Nos of vaccum outlet) safe efficient means delivering services to patients staff in both general and special care applications, It is modular design made of horizontal aluminium extructions and ability to house medical gas terminal unit and electrical socket , smooth curve surfaces , no visible screws and choice of color decor stripe, ease of installation via separate rail bracket or wall mounting plates easy removal of covers for maintenance and pipeline connection, integral twin medical rail as part of the modular extrusion design. It has segregation of services ie extra low voltage, normal voltage and medical gases, should be maintained throughout. Its aluminium extruction have a polyester coating finish on all external surfaces. It have provision of nurse call, data or monitoring sockets should be made at the point of manufacturing. It should be supplied duly pre-piped, prewired and fully tested or in carcass, complete all as per directed by Engineer-in-Charge. Make: Oxicareengineerings/ Global medlife solution/Aktiv/Life line Mediso/Tedisel / Unissi or equivalent Make approved by HQ CENC.
Supply , installation , commissioning and testing of Oxygen Manifold confirming to ISO standard configured as 6+ 6 Cylinder Manifold is suitable to withstand a pressure of 145 Kg/cm2, along with high-pressure copper annealed tail pipes with end Brass adapter suitable for Oxygen Cylinders and manifold.Top frame comprising of high pressure copper pipes of size 1/2” NB x 15swg with high pressure brass fittings made of high tensile brass and connections through non return valves; high pressure copper tail pipes, made of high pressure copper pipe of size 1/4” NB x 15 swg. The design of middle and bottom frames should be provided to fit both round and flat bottom cylinders safely. The manifold should be tested (hydraulically) at 3500 psig.A High-Pressure Regulator to be mounted on the Cylinder Manifold System for reducing the cylinder pressure suitable to the line pressure.To reduce the risk of medical oxygen system from contamination due to ignition of fluorinated polymer materials, only Non Halogenated Polymer materials should be used in the Non Return Valves and high pressure side of the Pressure Regulators of the manifold system. Make: Oxicareengineerings/ Global medlife solution/Aktiv/Life line Mediso/Tedisel / Unissi or equivalent Make approved by HQ CENC.
Supply , installation , commissioning and testing of Emergency Oxygen Manifold with mox regulator confirming to ISO standard configured as 3+3 Cylinders Manifold is suitable to withstand a pressure of 145 Kg/cm2, along with high-pressure copper annealed tail pipes with end Brass adapter suitable for Oxygen Cylinders and manifold.Top frame comprising of high pressure copper pipes of size 1/2” NB x 15swg with high pressure brass fittings made of high tensile brass and connections through non return valves; high pressure copper tail pipes, made of high pressure copper pipe of size 1/4” NB x 15 swg. The design of middle and bottom frames should be provided to fit both round and flat bottom cylinders safely. The manifold should be tested (hydraulically) at 3500 psig.A High-Pressure Regulator to be mounted on the Cylinder Manifold System for reducing the cylinder pressure suitable to the line pressure.To reduce the risk of medical oxygen system from contamination due to ignition of fluorinated polymer materials, only Non Halogenated Polymer materials should be used in the Non Return Valves and high pressure side of the Pressure Regulators of the manifold system, complete all as per directed by Engineer-in-Charge. Make: Oxicareengineerings/ Global medlife solution/Aktiv/Life line Mediso/Tedisel / Unissi or equivalent Make approved by HQ CENC.
Supply, installation, commissioning and testing of Fully Automatic Oxygen Control Panel consists of two first stage regulators each capable of delivering 100-200 psig outlet pressure. The first stage regulators are connected to a common second stage regulator which delivers an outlet pressure of 60 psig. The first two regulators meant for first stage , each having 40 micron inlet filters and with switchover system incorporated fro RUNNING to RESERVE bank due to differential pressure. The control panel should provide two individual content contact pressure guages to indicate the cylinder pressure in the two wings of the manifold and common pressure guage to indicate the delivery/ line pressure. The control panel should have built in audio- visual alarm indications for automatic bank changeover. The electrical circuits shall consist of a step down transformer, 3 Nos. enunciator windows showing ' Power On’, ‘Line Pressure Normal’ and ‘Reserve in Use'. All these items shall be housed in a metal cabinet made out of MS plate and side wall & top made of MS sheet duly painted with best quality of Epoxy paint. The control panel system should be compliant to HTM0201/NFPA/DIN/EN/ISO-7396-1 and shouuld be CE certified. Note 1. No interruption of gas supply in case of power failure. 2. Status of each bank is indicated on big pressure gauges. 3. Line pressure is indicated in the small pressure gauge. Make: Greggersen/ Trumph/Tedisel Medical / Unissi /WTK
Supply, installation, commissioning and testing medical Oxygen gas terminal outlet point complete with matching probe / adopter(BS Type) with one end suitable for hose & other end suitable for terminal units complete set for Oxygen & vacuum system as per technical specification confirming to CE certified .Outlets should be manufactured with a 165 mm long Copper inlet pipe stub which is silver brazed to the outlet body. Body should be of one-piece brass construction. For positive pressure gas services, the outlet should be equipped with a primary and secondary check valve and the secondary check valve should be rated at minimum 200 psi in the event the primary check valve is removed for maintenance.The outlet assembly must have separate colour coding for each service and should accept only corresponding gas specific adapters.All outlets must be cleaned and de-greased for medical gas service, factory assembled and tested, complete all as per directed by Engineer-in-Charge.The medical gas outlets should be of quick connecting and wall mounted modular type Make: Oxicareengineerings/ Global medlife solution/Aktiv/Life line Mediso/Tedisel / Unissi or equivalent make approved by HQ CENC.
All as per srl item No 6 but for medical vaccum gas outlet point with probe/adopter, complete all as per directed by Engineer-in-Charge.
Supply, installation, testing and commissioning oxygen back pressure compensated flow meter with control range of 0-15 LPM (calibration within ±10%) with humidifier bottle 250 ML made of unbreakable polycarbonate material and auto clavable at 121 degree centigrade temperature. The flow meter shall be strict precision, durability standard and body shall made of brass chrome plated materials. The flow tube and shrouded components shall be made of clear impact resistant polycarbonate. Flow tube shall have large and expanded 0-5 lpm range for improved readability at low flows with inlet filter of stainless steel wire mesh to prevent entry of foreign particles, complete all as per directed by Engineer-in-Charge. Make: Oxicareengineer / Global med life solution/ Lifeline Medisol/ Aktiv/ Unissi/ or equivalent make approved by HQ CENC
Supply, installation , testing and commissioning of Gas Area Alarm is micro-processor based which monitor the pressures of medical gases like oxygen, vacuum levels at a specific area of piped gas system. The electronic circuitry has been such that if the pressure / vacuum in the gas pipeline drops below the present limit, the equipment shall give an audio-visual alarm. Visual alarm remains active even after pressing of “Mute” button. But it comes to normal condition when gas pressure / vacuum return to normal level. The equipment should have following features: • Display of Line Pressure for all the services (both for Pneumatic & Vacuum Services) with factory calibrated pressure sensors. • Color coded LED Display of Line pressure status (High – Caution – Normal – Caution– Low ) • Audible Alarm for High & Low pressure condition. • Test and Alarm Acknowledge (Mute) facility. • Programming facility of alarm limits from front panel (Password protected, to be done through Contractor’s engineer/ authorized person of the hospital). • Should be of light weight and with small and compact design. • Highly sensitive gas pressure sensors of reputed make should be used. • Nut & Nipples to be provided for connection with Pneumatic supply line. • Low voltage internal operation with input power supply of 220V AC. • Wall mounting facility. Make: greggersen/Unissi /WTK /Germany or Pneumatik Berlin Germany /MPS/Tri-Tech Medical-USA /Phoenix - Pipeline -UK/ Modular -US or equivalent make approved by HQ CENC
Supply, installation, commissioning and testing Medical Vacuum Pump with Receiver tank Vacuum tank-1000 ltrs, bacteria filters, vacuum filters, Vacuum switch complete set all technical specifications and as per following details: a. Vacuum Pump single stage, minimum capacity 10 HP motor having capacity of 149 cfm . Pump should be fitted with MS channel frame complete having vibration isolators with 3 Nos. Silencer at vacuum pump discharge end. The pump should be air cooled 3 sets of isolation valve on return valve, on return valve and filter should be provided with pumps the system should be expandable to received one additional pump in future. b. Drive assembly: The electrical drive for vacuum pump should be consist of TEFC induction motor operating at 1440 RPM and connected through a V belt. The motor should be suited for 415V±10%50Hz 3 phase AC supply capacity 10 HP . The reservoir should be supplied fabricated from MS sheet plates as per IS:226 and manufactured as per IS:2825. The reservoir should be supplied with safety valve, pressure gauge and drain valves etc. manufacturers test certificates to be provided d. Vacuum Switch: Vacuum Switch of 0-760 mm range should be provided, should have bot NO/NC contracts and suitable for AC supply. The medical vacuum plant shall be CE marked / ISO comprising of air cooled, oil lubricated rotary vane / reciprocating vacuum pumps to provide a flow rate. Make:Ingersoulrand/Boge/Coaire / Anest Iwara/ IR or equivalent make approved by HQ CENC.
Supply , installation testing & commissioning of Medical line valve or isolation valve 12 mm OD should be CE certified or ball valves as per BS 1057, rated at 400 psig , as well full vacuum (29.9" Hg) valves go from full off by turning the vinyl- gripped valves handle 90 degree locking type handles may be ordered separately when required. Copper tubing is factory installed to help prevent valve seat damage during soldering main and riser valves seat damage during soldering, main and riser valves are not required to be installed in a box unless specially noted or required, complete all as per directed by Engineer-in-Charge. Make: Conex/RB/RN/SIM/AUDCO/AKTIV/Matrix or equivalent Make approved by HQ CENC
All as per item No. 11 but isolation valve of size 15mm OD, complete all as per directed by Engineer-in-Charge.
All as per item No. 11 but isolation valve of size 22 mm OD, complete all as per directed by Engineer-in-Charge.
All as per item No. 11 but isolation valve of size 28 mm OD, complete all as per directed by Engineer-in-Charge.
All as per item No. 11 but isolation valve of size 42 mm OD, complete all as per directed by Engineer-in-Charge.
Supply, installation, testing and commissioning of solid drawn, seamless , deoxidized, non arsenical, half hard , tempered and greased copper pipe of following size with complete fitting / fixtures as per technical specifications and BS EN -13348:2008 all as specified and directed. (i) Copper pipe 42mmODX 1.2 mm thick
(ii) All as per item No. 16 but Copper pipe 28mmODX 0.9 mm thick, complete all as per directed by Engineer-in-Charge.
(iii) All as per item No. 16 but Copper pipe 22mmODX 0.9 mm thick, complete all as per directed by Engineer-in-Charge.
(iv) All as per item No. 16 but Copper pipe 15mm ODX 0.9 mm thick , complete all as per directed by Engineer-in-Charge.
(v) All as per item No. 16 but Copper pipe 12mm ODX 0.7 mm thick , complete all as per directed by Engineer-in-Charge.
Note : Applicable to item No. 16 to 20 (a) Installation of Piping should be carried out as per medical standard with outmost cleanliness. (b) Only pipes fittings and valves which have been degreased as per approved medical standards should be used. (c) Pipe fixing clamps for up to as per outer dia of pipes (d) Pipes should suitable nonferrous suitable for the diameter of the pipe and wherever required rigid metallic hanging or cemented supports to be used. (e) The main lines to the building to be taken overhead through heavy grade metallic poles or GI tubes with inspection points in metallic boxes where ever required. Adequate supports shall be provided while laying pipelines to ensure that pipes do not sag. Suitable sleeves shall be provided wherever pipes cross through walls / slabs. All pipe clamps shall be non-reactive to copper. (f) All Pipes joints shall be of copper to copper and should he brazed by silver brazing filler a material without flux while being brazed joints shall be continuously purged with oil free dry nitrogen to prevent the formation of copper oxide on inside surface of the joint (g) All pipes should be installed without springing or forcing (h) All pipes should be protected against mechanical injury in a manner satisfactory to authorities having jurisdiction. (j) Test: After erection, all the new pipes cleaned or purged with the help of dry nitrogen at 2 times of working pressure for 24 hours (k) Painting : All existing and proposed exposed pipes / should be painted with two coats of synthetics enamel paint & color codification as per IS:2379 of 1963 and approved medical standard. (l) All concealed pipes to have gas identification bands/labels at approaches distances. (m) Similarly all pipe in the wall should be tested /painted/labelled and properly insulated. Make: Mexflow/Janya/Metal Gem/Cubex or equivalent make approved by HQ CENC
Supply, installation, testing and commissioning of ward vacuum unit or ward vacuum unit with 600ml suction collection jar-01 No as per specifications mention below: (a) The vacuum regulator should have a vacuum gauge with analog display. The scale should be in mbar or mmhg as required by the end user. (b) 1 No of suction regulator and 1 No of 600 ml polysulfone / polycarbonate collection jar. (c) Suction regulator should be supplied with a safety bar, including and antibacterial filter and filter and anti-overflow safety device and should have wide membranes continuous suction controller. (d) Should have vacuum level 0-760 mm of Hg and should have vacuum gauge fitted with a protective bumper device. (e) Should have a on/off knob allowing for the quick restoration of a readjusted vacuum level (f) Must have central adjustment knob with a color coded for 0-760mm of Hg. Should have Polysulfone /polycarbonate 1700 cc safety jar, autoclavable at 121 degree centigrade at 5 mins, unbreakable , fitted with an anti-overflow safety device and equipped with a plastic anti-bacterial filter and should be totally transparent, to ensure perfect sucked liquid visibility Make: Oxicareengineer / Global med life solution/ Lifeline Medisol/ Aktiv/ Unissi/ or equivalent make approved by HQ CENC
Supply and fixing high pressure tube, complete all as per directed by Engineer-in charge.
Supply and fixing low pressure tube, complete all as per directed by Engineer-in charge.
Supply, installation ,commissioning and testing and electric control panel wall supported,indoor type having sheet metal 2 mm thickness painted with dry powder 3 tank process, hinged type for vacuum pump of suitable size comprising of change over 4 pole 415 volt 100 Amp-01 Nos, MCCB 415 volt 4 Pole rupturing capacity 16 kA ,100 Amp-01 Nos as incomming, MCCB 415 volt 4 Pole rupturing capacity 16 kA ,63 Amp-02 Nos as outgoing, DOL type starter for 10 HP motor with single phase preventer -02 Nos MCB SP 20 Amp -10 Nos & MCB SP 6 Amp -10 Nos incl wiring with suitable size of copper conductor for connection all as per directed by Engineer-in charge
Supply, installation , testing and commissioning of Zonal Valve box for -2 gas services Valve Box is made of Aluminium / Powder Coated M.S. Material.Valve Box Assembly consists of the following:• Lever operated quarter turn valve (i.e. 90 degree shut off ball valve, factory degreased and suitable for oxygen service) with brass body and chrome plated brass ball.• Brass fittings (Nut, Nipples and extruded brass Adapter) KE Type Seat Brass Block for pressure gauge • 2" Dial gauges (0 - 10 kg/cm2, 0 - 760mm Hg)• Nylon Bush for copper pipes holding with the valve box• Beading for box lead• Lockable cover with breakable glass so that during normal operation access has been by key. But during emergency operation, access by breaking the glass panel. Make: greggersen/ Trumph/Tedisel / Unissi /WTK or equivalent make approved by HQ CENC
M&L for earthing complete with copper plate 60cm×60cm ×3.15mm thick electrode , burried directly in ground, earth pit not less than 2.25 m deep below normal ground level with top edge of the plate not less than 1.5m below normal ground level, connected to equipment by using 20×3mm copper strip as earthing lead as earthing lead by means of bolts,nuts,checknut & washers of galvanised iron all as shown in elect plate No.5 of MES schedule and connected to earthing test point all as specified and directed including earth pit of size 30cmx30cmx45cm made of PCC 1:3:6 and PCC 1:2:4 cover, funnel, GI tubing etc. & testing on completion, all as specified. Note : The units rate includes the cost of excavation & earth work in any type of soil charcoal & salt in altemate layers, 20mm GI Pipe (medium grade) for watering & 25 mm GI protection pipe (light grade) funnel wire mesh PCC 1:2:4 type B1 chamber in cement concrete ,50 mm thick precast RCC cover duly reinforced with XPM (weight not less than 3 kg /sqm ) and CI frame etc , and including provision of test point, connecting to test point and testing on completion etc, complete all as per directed by Engineer-in-Charge.
Excavation in tranches not exceeding 1.5 m wide and not exceeding 1.5 m deep for foundation of laying cable including getting out in soft/loose soil, complete all as per directed by Engineer-in-Charge.
Returning filling in including spreading watering and well ramming in layers not exc 25 cm, complete all as per directed by Engineer-in-Charge.
Removing surplus soil to distance not exceeding 50 m and depositing where directed at a level not exceeding 1.5 m above the starting point, complete all as per directed by Engineer-in-Charge.
M & L for protection to cable with one layer of brick laid flat across the cable using all size sub class 'B' Bricks, complete all as per directed by Engineer-in-Charge.
Sand cushioning to under ground electric cables including watering and consolidation all as per clause 19.75 of SSR Part-I, complete all as per directed by Engineer-in-Charge.
Supply, laying, jointing and testing XLPE cable insulated, armoured, earthed, heavy duty electric cable with aluminium conductor, 1100 Volt grade, cross sectional area 50 sqmm 3.5 core including necessary cable lugs, glands, insulation tape etc complete all as specified and as directed. Make:- Finolex/ Gloster/ Polycab
Supply and fix UPS for 30 min back up three phase AC input 300 volt to 450 volt and output single phase AC output voltage 230 volt+- 1% of capacity 20 KVA having input power factor > 0.95, sine wave with external battery bank DC voltage 192 to 240 VDC , constant voltage constant current type charger incl I/P under and voltage cut off , phase reversal and also protection at output for overload , short circuit , under voltage , battery low and audible alarm on battery mode , battery low , overload , fault incl LCD/LED indicators for load, battery level, bypass, overload etc all as specified and directed. Make EPXI, Canon, HP
M&L for HI-L0 bed manually operated having size 2200mmx 900mm width x 590 mm H consisting of 60mmx 30mm CR tubular frame work, incl four section perforated CR sheet Top having backrest, knee rest, trendenel berg and reverse trendeelenberg operated by screw and lever mechanism with four separate attd Chromium plated folding handles.polymer moulded head and foot board, collapsible side ralling (1320mm length) , Chromium plated telephonic saline rod with four locations incl four heavy duty swivel castors of 125mm dia with 2 brakes, four side corner buffers , pretreated powder coated having four section mattresses with 04 inch thick PU Foam of at least 40 density covered with PVC Rexene -01 No all as specified and directed
MAMTA ENTERPRISES
M/S SATPAL AND BROTHERS
kailu construction co
Usha Engineering Co.
stage.html
html • 0.05 MB
tech_bid_open.pdf
tech_eval.pdf
fin_bid_open.pdf
boq_comp_chart.xlsx
xlsx
fin_eval.pdf
aoc.pdf
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.03 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .