XEN MHED Udhampur
XEN MHED Udhampur
Supply Installation Testing and Commissioning of Medical Grade Online Reverse Osmosis RO Plant of 500 Liters per Hour Capacity Double Pass along with Allied Works for the Hemodialysis Unit in Community Health Centre Katra on Turnkey Basis
2025_PWDJK_274029_1
MHDU/TS/2024-25/109
Open Tender
Machineries/ Mechanical Engg Items
Item Rate
60 days
Katra
Please refer Tender documents.
2 documents required · 2 mandatory
₹500
XEN MHED Udhampur
₹24,000
14 May 2025
10 Mar 2025
9 Apr 2025
10 Mar 2025
9 Apr 2025
10 Mar 2025
Amount
Supply, Installation, Testing and Commissioning of a Medical Grade Online Reverse Osmosis (RO) Plant of 500 Liters Per Hour Capacity, double Pass along with allied works for the Hemodialysis Unit in Community Health Centre, Katra on Turnkey basis Complete job as per site condition in all respect as detailed below; 1. RO Water Treatment Plant capable of producing at least 500 LPH of high quality Reverse Osmosis treated water as per AAMI standards (chemical and Bacteria Free) suitable for Hemodialysis. 1. Capacity to treat impure inlet Raw Water with TDS of Range 150- 1500 PPM, Hardness 500 PPM 2. Semi-automatic start, stop and auto rinse facility with LED/LCD display for the RO parameters and other functions 3. Must have online/offline Mode. 4. Must have fully automatic disinfection cycle. 5. RO Output quality: As per AAMI Standard water required for Hemodialysis and related therapies i.e. 1. Rejection rates of Bacteria(CFU) & Endotoxins(EU) should be >99% 2. Rejection rates of Total Dissolved salts (TDS) should be :> 95% 3. TDS range 0-10 PPM. 4. Water conversion factor 60-75% 5. Conductivity<10 us/cm (microSiemens/cm) 6. Hardness<10 PPM 6. RO membrane material: Spiral Wound, Modified polymide Membrane with thin film composite with PH tolerance of Membrane 2-11. 7. Conductivity measuring range of inlet water: 100-1000 us/cm (micro Siemens/cm) (±10%) 8. Conductivity measuring range of pure water: 2-200 us/cm (micro Siemens/cm) 9. Facility to Rinse, Disinfect and clean the reverse osmosis system. 10. Facility for online monitoring and display of: 1. Raw water conductivity and temperature 2. Pure water conductivity and temperature 3. Running TDS 11. Facility to pump pure water directly to dialysis unit with return line back to RO Unit(True Online Connection) 12. RO Unit should have facility to indicate the following Protection Switches a. Low Pressure Switches b. High Pressure Switches 13. Must have facility to easy addition /upgradation to membrane for higher water output capacity(without change of pump capacity) 14. Must have facility for upgrading to latest technology in the future. 15. Pre-treatment with Mesh Filter of 50 microns 16. Horizontal multistage centrifugal raw water pumps 17. Sand Filter with sand particles of different grade. 18. Activated Carbon Filter with fine carbon granules. 19. Water Softener with Regeneration, brine fill and clean cycles. Also have brine tank incorporated in the system. 20. Particle filter, cartridge type of 5 microns before the RO system 21. Dry run protection for raw water pump. 22. Pressure monitoring across all filters. 23. Microprocessor/PLC controlled double stage RO system with user friendly interface for manual/automatic. 24. RO systems with final permeate capacity of 500 LPH. 25. In build display to show the permeate water temperature, permeate conductivity, operating hour in the RO system. 26. Dry run protection for pressure pumps of RO system. 27. Automatic start stop facility with permeate tank control. 28. Online and offline operation facility as per treatment requirement. 29. Water savings technology. 30. Emergency operation in case of control failure. 31. Stainless steel conical/ dish end bottom permeate water storage tank. 32. Two numbers (one working & another stand by) permeate water transfer pumps for dialysis loop. 33. UV system in Post RO module. 34. Complete training to hospital staff for operation, routine maintenance and functioning of RO plant in all modes including online and offline modes.
Supply, Installation, Testing and Commissioning of Online 5KVA UPS system & Stabilizer for power back up of Hemodialysis Unit & RO Plant at Dialysis Center of Community Health Centre, Katra for uninterrupted Power Supply with minimum backup of 5-6 hours to the system along with 2-years Warranty as detailed below; 5 KVA True online double conversion design with input voltage range 160 - 280VAC PWM Technology & IGBT Based Pure Sine Wave Output with THD < 3% Power Factor > 0.8 Large input operating range, eliminates unnecessary battery discharge UPS startup without battery Multi-functional LCD/LED display In-built Galvanic Isolation Automatic Battery Charge in UPS Off Mode Static Bypass and Cold Start EMI/RFI and Noise Filter Generator and Solar Compatible Auto-restart Function Extended backup time suitable charger Protection: Output short Circuit, Low Voltage of Battery, Over or lack of input Voltage, Overload and so on Standard: IEC/EN62040-1, 62040-2, 62040-3, Complying to CE, BIS, RoHS, ISO:9001, ISO:14001
GLOBAL AQUATECH (BID ID -2355560)
M/s VISHWAMITRA WATER TECHNOLOGIES PRIVATE LIMITED (BID ID -2353362)
tech_bid_open.pdf
tech_eval.pdf
fin_bid_open.pdf
boq_comp_chart.xlsx
xlsx
fin_eval.pdf
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .