Providing duplex system and bacterial filters therein
2022_PWD_65150_1
ED.II.GA.Tender/22-23-5310-11 dated 15.09.2022
Open Tender
Electrical Works
Item Rate
30 days
IGMC Shimla
Please refer Tender documents.
8 documents required · 8 mandatory
₹500
EE Electrical Divn No II
₹52,500
13 Oct 2022
22 Sept 2022
3 Oct 2022
22 Sept 2022
1 Oct 2022
22 Sept 2022
Amount
Oxygen System
SITC of Oxygen Manifold and control panel
SITC of the manifold and control panel should comprise of two banks of 10-12 cylinders each In additional bank of 10 cylinders should be supplied and installed for emergency supply of oxygen.
The control panel should be meant for the supply of oxygen at 2100 litres per minute
The unit (control panel)should comply with international standards like NFPA/ HTM-2022/ BIS/CDSSO
The control panel will need to be adjusted at a pressure to ensure 4 bar pressure in the two blocks of the hospital
The control panel should be from reputed manufacturers like: Becon Maedas/Ohio Medical/ Penlon/ Medicop/ Drager/Atlas COPCO/Aktive EuCE or USFDA certified
Feed Gas:Medical grade oxygen as per Indian Pharmacopeia 99.5% at pressure 130 – 140 bar
Gas Delivery Temperature (–) 10oC to (+) 30oC and ambient relative humidity of exceeding 80%
Altitude above MSL: 2250 meters
Frost formation and pressure drop is common problem in winters and rainy season. Therefore heating arrangement for the control panel have to be provided and be documented and quoted.
Safety relief valves, pressure gauges, NRVs and Alarms etc should be provided with means to prevent tampering by Un-authorized persons
Supply Installation and Testing commissioning of 4+4 Emergency Control pannel with manifold complete in all respects as per the satisfaction of Engineer-in-charge
SITC of Wall outlets oxygen
SITC of Oxygen, air and vacuum outlets HTM or NFPA compliant should be quoted along with 2 probes for each out let. Both imported and indigenous outlets and probes should be quoted separately. The outlets and probes should be as specific and non- interchangeable. The gas outlets have heavy duty constituents and smooth connection and disconnection. The outlets must be labeled and color coded. Imported units should have UL Listing and Indian units must be IS/CE certified. Make Becon Medas, Dragger, Allied USA.
SITC of Probes for Wall Outlets
SITC of OXYGEN FLOW METER WITH HUMIDIFIER BOTTLE (CE MARKED)
The oxygen manifold should comply with HTM 02-01/NFPA 99/ DIN / EN / ISO-7396-1 standards.
Back Pressure Compensated flow meter for accurate gas flow measurement with following features:
A) Control within a range of 0-15 LPM.
B) It should meet strict precision and durability standard.
C) The flow meter body should be made of brass chrome plated materials.
D) The flow tube and shroud components should be made of clear, impact resistant polycarbonate.
E) Flow tube should have large and expanded 0-15 LPM range for improved readability at low flows.
F) Inlet filter of stainless steel wire mesh to prevent entry of foreign particles
G) The humidifier bottle is made of unbreakable & reusable polycarbonate /polysulfone/polypropylene material autoclavable at 121 degree centigrade.
SITC of Horizontal/ vertical BED HEAD PANEL (CE MARKED)
Bed head panel should conform to NEPA99/HTM-02-01/ISO-7396-I/DIN/EN.
Efficient,safe &Robust design in extruded aluminium section
Smooth curved surface and choce of base colour and fascia plates
Unit should have integrated rails system to mount accessories
The head wall system should be constructed of aluminium extrusions joined together
form a caracass to suit the particular application. Unit shall be factory assembled for
Elecrical and mechanical components
Segregation of services i.e. electrical wiring supply and medical gases shall be maintained throughout.
Front fascia plates should be removable individually to access for respective services
Bed space management system with optional equipment rail.
With all Equipment rail
Should have built in light feature
Medical gas pipe line outlets should be factory fitted by & tested by the manufacturer. Outlets configuration as per hospital requirement.
Following accessories should be provided locally with HBHP
5/15 Amp Module Electrical Socket with switches= 6 sets
IV pole=2 no.
Vaccum slide =1 no.
Sliding blocks = 2no.
Infusion pump mount =1
Monitor tray with slider= 1no.
Utility basket =1 no.
Bed Head Panel should be CE marked
SITC of Ward Vacuum Units ( CE Marked)-
It should comply with HTM 02-01/NFPA 99/ DIN / EN / ISO-7396-1 standards.
It must consist of the following:-
1. One no of Suction Regulator and 1no of 1000 ml polysulfone /polycarbonate/polypropylene collection jar.
2. Suction Regulator: Suction regulator should be supplied with a safety jar, including and antibacterial filter and an anti-overflow safety device. Should have wide membrane continuous suction controller
3. Should have vacuum levels: 0-760 mm of Hg
4. Should have vacuum gauge fitted with a protective bumper device.
5. Should have on/off knob allowing for the quick restoration of a readjusted vacuum level.
6. Must have central adjustment knob with a color coded for 0 to 760 mm of Hg. Should have Polysulfone/polycarbonate 100cc safety jar, autoclavable at 121º C, unbreakable, fitted with an anti-overflow safety device and equipped with a plastic antibacterial filter. It should be totally transparent, to ensure perfect sucked liquid visibility.
SITC of Wall outlets Vaccum (Specs same as given for Oxygen Outlets)
SITC of Oxygen, air and vacuum outlets HTM or NFPA compliant should be quoted along with 2 probes for each out let. Both imported and indigenous outlets and probes should be quoted separately. The outlets and probes should be as specific and non- interchangeable. The gas outlets have heavy duty constituents and smooth connection and disconnection. The outlets must be labeled and color coded. Imported units should have U.
ULTIMED SOLUTION
stage.html
html • 0.03 MB
tech_bid_open.pdf
tech_eval.pdf
fin_bid_open.pdf
boq_comp_chart.xlsx
xlsx
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.03 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .