� Refrigerated multipurpose bench-top�Centrifuge�with�at least 4�Litre�capacity�, with touch-screen interface,� System should be capable to use both fixed angle and swing out rotors with adapters for use with different vessel formats
� System should maintain temperature range �11 �C to 40 �C
� System should have a maximum Capacity of swing out rotor of 4 x 1000 mL, 4 x 5 MTP and fixed angle 6 x 250 mL bottle, 6 x 50 mL, 48 x 1.5/2.0 mL
� Maximum speed for fixed angle rotors should be ?14,000 rpm and RCF ? 22,000 xg force, for swing out rotors maximum speed ? 5,000 rpm and RCF 5, 000 xg force
� System should have user-friendly at least ? 6-inch touchscreen interface with an intuitive user experience providing fast and error-free operation to set speed RPM / RCF
� System should be capable of working with user defined radius correction values when working with different adapters
� System should have fast temperature function for rapid cooling of centrifuge and stand-by cooling options to hold the sample at same temperature.
� System should be capable of storing up to ? 90 programs each consisting of all run parameters (incl. ramps and radius correction settings) along with rotor and program compatibility check.
� System should have excellent temperature dynamic control technology with compressor running continuously during the run time. This results in a precise cooling performance and the set temperature is kept stable at all time.
� System should have a rotor option with Universal bucket to hold three different formats of vessels (Tubes, Plates and Bottles). To avoid too many buckets requirements for each vessel formats and occupy bench space.
� System should have Universal adapter to support tubes/plates/bottle formats without needing to change the adapters. To bring more comfort and convenience for a faster workflow and adds a cost saving, lab space saving by not having too many individual adapters for each format.
� System must be equipped with automatic rotor recognition with its maximum radius for the bucket and imbalance detection for maximum operational safety
� Timer setting � 10 s to >90h , with continuous run function
� System should have a separate short spin function key with user defined speed. To reduce the sample getting exposed to maximum speeds when using sensitive samples.
� System should have 10 acceleration and 10 deceleration steps
� Noise level at max speed should be less than 60 dB(A) for quite operation in work place
� Fixed angle rotors and rotor lids should be made of metallic and must be fully autoclavable at >110 �C
� Swing bucket rotors must be made of metallic and all the buckets, adapter and caps must be fully autoclavable at >110 �C
� System must have a smallest possible foot print and smallest lid opening height for easy sample accessing while loading and unloading samples
� Centrifuge lid with soft-touch lid closure, and low opening height for stress-free lid locking
� Centrifuge lid should possess two gas springs to support the lid securely from falling when opened and this gives additional safety for the users.
� System should have emergency lid release preferably in front side (in case of power failure)
� System should have favourite function in the home screen for faster parameter settings
� System should have user management function for enhanced security with minimum 3 user levels (Admin, normal user and restricted user)
� System should have the facility to support documentation of all the parameter settings, run records, event logs and rotor logs
� System should have the option to export all run records, event logs via USB port
� System should have digital lab support with remote monitoring solution
� System should have automatic stand-by mode to reduce energy consumption and to extend compressor life when not in use up to user defined hours.
� System should have extended display function to show both run time and pending time
� System must have an USB-port for service maintenance and software upgradation.
Rotors to be supplied:
? High speed fixed angle rotor 48 x 1.5/2.0 mL with aerosol tight lid and maximum of ?14,000 rpm and ? 22,000 xg. Both rotor and rotor lid is made of anodized aluminium and autoclavable at >110 �C for at least 20 mins.
? Fixed angle rotor 6 x 50 mL with aerosol tight lid for 50 mL conical tubes with maximum of 12,000 rpm and ?20,030 xg. Optional Adapter for > 13 mm blood collection tubes, 5.0 mL, 15 mL tubes (rotor must be 45� angle). Both rotor and rotor lid is made of anodized aluminium and autoclavable at >110 �C for at least 20 mins.
UPS & Stabilizer to be supplied
� The system should be attached with at least 5KVA stabilizer and with 2KVA maintenance free UPS.
Other Condition
� Features in the quotations should be substantiated with proper company catalogue/brochure/manual/website
� System must be European(EU) CE Certified
� System should be supplied with at least 1 year warranty with minimum of 3 years free AMC and supplier should have local service center in Bangalore and with 24x7 trained service engineers support.
� The supplier should be OEM preferably locally available.
� The supply of the equipment, installation and demo should be done by the firm at their own cost, after the installation the performance of the equipment should be with full satisfaction.
� The firm should have supplied and obtained work done satisfactory report for the same model from at least 5 Govt. organization preferably ICAR research Organization. |