DBS Implant and Accessories
2025_SMSMC_495705_1
1960_21.08.2025
Open Tender
Miscellaneous Goods
Item Wise
730 days
SMS Hospital
Please refer Tender documents.
3 documents required · 3 mandatory
₹5,000
Medical Superintendent cum Member Secretary, RMRS
₹7.4 L
Yes
SMS Hospital, Jaipur
25 Mar 2026
21 Aug 2025
20 Nov 2025
21 Aug 2025
18 Nov 2025
8 Sept 2025
29 Aug 2025
Select, Excess (+), Less (-)
Tender Inviting Authority: SMS Hospital, Jaipur
Name of Work: Technical Specification point wise DBS Implant & Accessories
Contract No: NIT : 1960 Dated: 21.08.2025
Name of the Bidder/ Bidding Firm / Company :
Sl. No. Item Description Item Code / Make Quantity Units Estimated Rate in Rs. P Addition / Deduction Addition / Deduction Values Currency Convertion against each Item Quoted Currency in INR / Other Currency Rate per unit without GST To be entered by the Bidder in Rs. P GST (If applicable in Percentage) GST per unit in Rs. P Make / Model HSN / SAC Code TOTAL AMOUNT excluding taxes in Rs. P Rate per unit with GST Rs. P TOTAL AMOUNT In Words
1 DBS Rechargeable with Directional Lead with25yr battery life & Image guided programming integrated inside clinician Programmer with MICC “multiple independent current control” technology with Zero Volt Technology,Bluetooth Connectivity with clinician and patient programmers, USFDA & CE Approved Kit contents: Rechargeable pulse generator-1 , Directional lead Kit -2 , Extension Kit -2 , Burr hole locking kit -2 , Tunneling tool -1 , Patient Remote Control Kit -1, Patient Charging Kit -1 item1 1 Nos 20 Excess(+) Full Conversion INR 1.01 Supplying, Conveying and fixing spls. Including eart item1 123.223 Nos
2 DBS Rechargeable with Non directional lead with 25yr battery life & Image guided programming integrated inside clinician Programmer MICC “multiple independent current control” technology with Zero Volt Technology,Bluetooth Connectivity with clinician and patient programmers. have 8 Contact non-directional Lead Kit with Array span of 15.5mm for maximum coverage, USFDA & CE Approved Kit contents:Rechargeable pulse generator -1, non-directional lead Kit -2, Extension Kit -2, Burr hole locking kit -2, Tunneling tool -1, Patient Remote Control Kit -1, Patient Charging Kit -1 item2 1 Nos 0 Excess(+) Full Conversion INR 1.02 Construction of chamber for 100mm sluice plates item2 213 Nos
3 DBS Non rechargeable with Directional Lead with Image guided programming integrated inside clinician Programmer MICC “multiple independent current control” technology &Bluetooth Connectivity with clinician and patient programmers,USFDA & CE Approved Kit contents: Non-Rechargeable pulse generator -1 , Directional lead Kit -2 , Extension Kit -2, Burr hole locking kit -2 , Tunneling tool -1 , Patient Remote Control Kit -1,) item3 1 Nos 0 Excess(+) Full Conversion INR 2 Construction of chamber for 100mm sluices valve item3 10 Nos
4 DBS Non rechargeable with non-directional Lead with Image guided programming integrated inside clinician ProgrammerMICC “multiple independent current control” technology withBluetooth Connectivity with clinician and patient programmers. Should have 8 Contact non-directional Lead Kit with Array span of 15.5mm for maximum coverage,USFDA & CE Approved Kit contents: Non-Rechargeable pulse generator -1, non-directional lead 1) item4 1 Nos 0 Excess(+) Full Conversion INR 1.01 Supplying, Conveying and fixing spls. Including eart item1 123.223 Nos
5 DBS implant Kit with non-rechargeable, 3T MRI compatible and sensing enabled IPG and directional leads.The implanted pulse generator (DBS neurostimulator- IPG) should have a minimum service life of 5-7 years and there should not be a need to recharge it periodically. IPG should have the capability for group programming and ability to store the programming data in IPG itself and retrieve from IPG.The system should be able to sense Local Field Potentials (LFPs) from brain targets and this data should be accessible through the physician programmer. The leads should be able to stimulate in specific directions, in addition to the ability to stimulate circumferentially. Markers on leads should facilitate directional programming even if lead rotates. The leads should also be able to sense LFPs.Burrhole kit should be there for securing lead in place. Microelectrodes for intra op stimulation and recording should be there.The extension connecting the lead and IPG should be flexible There should be a tunneller to place the extension wire and the tunneller should be metallic and firm to enable smooth tunnelling.The system should have approval for conditional full-body 1.5T & 3T MRI Compatibility.Patient programmer, enabling the patient to see and if needed, modify the stimulation under guidance, should be available. Patients should be able to mark dyskinesia, tremor etc. using the programmer. item5 1 Nos 0 Excess(+) Full Conversion INR 1.02 Construction of chamber for 100mm sluice plates item2 213 Nos
6 DBS implant Kit with rechargeable, 1.5T & 3T MRI compatible and sensing enabled IPG and directional leads.The implantable pulse generator (DBS neurostimulator- IPG) should have a service life of 15+ years if regularly recharged by the patient at specified intervals.IPG should have the capability for group programming and ability to store the programming data in IPG itself and retrieve from IPG.The system should be able to sense Local Field Potentials (LFPs) from brain targets and this data should be accessible through the physician programmer. The system should have FDA approval foradaptive DBS.The system should be able to identify the best contact for stimulation based on the LFP data.The leads should be able to stimulate in specific directions, in addition to the ability to stimulate circumferentially. Markers on leads should facilitate directional programming even if lead rotates. The leads should also be able to sense LFPs.The system should be able to predict the recharging intervals.The extension connecting the lead and IPG should be flexible. There should be a tunneller to place the extension wire and the tunneller should be metallic and firm to enable smooth tunneling. The system should have approval for conditional full-body 3T MRI Compatibility. Patient programmer, enabling the patient to see and if needed, modify the stimulation under guidance, should be available. Patients should be able to mark dyskinesia, tremor etc. using the programmer. Burrhole kit should be there for securing lead in place.Microelectrodes for intra op stimulation and recording should be there. item6 1 Nos 0 Excess(+) Full Conversion INR 2 Construction of chamber for 100mm sluices valve item3 10 Nos
7 DBS implant Kit with non-rechargeable, 1.5T & 3T MRI compatible and sensing enabled IPG and conventional ring mode (non-directional) 4- electrode-contact leads.The implanted pulse generator (DBS neurostimulator- IPG) should have a minimum service life of 3-5 years and there should not be a need to recharge it periodically.IPG should be capable of group programming and have the ability to store the programming data in IPG itself and retrieve from IPG.IPG should be able to sense Local Field PotentialsLeads should have four ring-shaped (non- directional) lead contacts; should be able to deliver stimulation circumferentially.The extension connecting the lead and IPG should be flexible. There should be a tunneller to place the extension wire and the tunneller should be metallic and firm to enable smoothtunnelling.Should have approval for conditional full body 3T Magnetic Resonance Imaging (MRI) Compatibility.Patient programmer, enabling the patient to see and if needed, modify the stimulation under guidance, should be available.Microelectrodes for intra op stimulation and recording should be there. item7 1 Nos 0 Excess(+) Full Conversion INR 1.01 Supplying, Conveying and fixing spls. Including eart item1 123.223 Nos
8 DBS implant Kit with rechargeable IPG and conventional ring mode (non-directional) 4- electrode-contact leads. The implanted pulse generator (DBS neurostimulator- IPG) should have a service life of 10-15 years if regularly recharged by the patient at specified intervals. Both constant current and constant voltage stimulation should be possible. The recharger should be able to recharge the IPG in a non-invasive manner and should be wireless. The IPG should have the capability for group programming and ability to store the programming data in IPG itself and retrieve from IPG. The leads should have four ring-shaped (non-directional) lead contacts and should be able to deliver stimulation circumferentially. The extension connecting the lead and IPG should be flexible. There should be a tunneller to place the extension and the tunneller should be metallic and firm to enable smooth tunneling. The system should have approval for conditional full-body 1.5 T MRI Compatibility. Patient programmer, enabling the patient to see and if needed, modify the stimulation under guidance, should be available. Microelectrodes for intra op stimulation and recording should be there. item8 1 Nos 0 Excess(+) Full Conversion INR 1.02 Construction of chamber for 100mm sluice plates item2 213 Nos
9 DBS implant kit with rechargeable IPG and directional leads. The implanted pulse generator (DBS neurostimulator- IPG) should have a service life of 10-15 years if regularly recharged by the patient at specified intervals. Both constant current and constant voltage stimulation should be possible. The recharger should be able to recharge the IPG in a non-invasive manner and should be wireless. The IPG should have the capability for group programming and ability to store the programming data in IPG itself and retrieve from IPG. The leads should be able to stimulate in specific directions, in addition to the ability to stimulate circumferentially. Markers on leads should facilitate directional programming even if lead rotates. The leads should also be able to sense local field potentials (LFPs), so that there is no need to replace lead if, in future, rechargeable IPGs with sensing capability become available and the patient opts for sensing. The extension wire connecting lead and IPG should be flexible. There should be a tunneller to place the extension wire and the tunneller should be metallic and firm to enable smooth tunneling. The system should have approval for conditional full body 1.5 T MRI Compatibility. Patient programmer, enabling the patient to see and if needed, modify the stimulation under guidance, should be available. Burrhole kit should be there for securing lead in place. Microelectrodes for intra op stimulation and recording should be there. item9 1 Nos 0 Excess(+) Full Conversion INR 2 Construction of chamber for 100mm sluices valve item3 10 Nos
10 DBS implant Kit with rechargeable,1.5T & 3T MRI compatible and sensing enabled IPG and conventional ring mode (non-directional) 4- electrode-contact leads. The implanted pulse generator (DBS neurostimulator- IPG) should have a minimum service life of 15+ years, if charged at regular period. IPG should be capable of group programming and have the ability to store the programming data in IPG itself and retrieve from IPG. IPG should be able to sense Local Field Potentials. Leads should have four ring-shaped (non-directional) lead contacts; should be able to deliver stimulation circumferentially. The extension connecting the lead and IPG should be flexible. There should be a tunneller to place the extension wire and the tunneller should be metallic and firm to enable smooth tunnelling. Should have approval for conditional full body1.5T & 3 T Magnetic Resonance Imaging (MRI) Compatibility. Patient programmer, enabling the patient to see and if needed, modify the stimulation under guidance, should be available. Microelectrodes for intra op stimulation and recording should be there. USFDA & CE Approved. item10 1 Nos 0 Excess(+) Full Conversion INR 1.01 Supplying, Conveying and fixing spls. Including eart item1 123.223 Nos
11 11. DBS Rechargeable Replacement kit with from Non-Rechargeable device Kit contents: Rechargeable pulse generator -1, Patient Charging Kit -1. item11 1 Nos 0 Excess(+) Full Conversion INR 1.02 Construction of chamber for 100mm sluice plates item2 213 Nos
12 12. DBS Rechargeable Replacement kit with Adapter (Medtronic to Boston Scientific Device Replacement) Kit contents: Rechargeable pulse generator -1, Patient Remote Control Kit -1, Patient Charging Kit -1, M8 adapters -02 item12 1 Nos 0 Excess(+) Full Conversion INR 1.01 Supplying, Conveying and fixing spls. Including eart item1 123.223 Nos
13 13. DBS non-rechargeable replacement kit with adapter (Medtronic to Boston Scientific Device) Kit contents: Non-Rechargeable pulse generator -1, Patient Remote Control Kit -1, M8 adapters -02 item13 1 Nos 0 Excess(+) Full Conversion INR 1.02 Construction of chamber for 100mm sluice plates item2 213 Nos
Total in Figures 4 Construction of chamber for 100mm sluice plates item5 10 Nos
Quoted Rate in Figures Select %
Quoted Rate in Words
stage.html
html • 0.02 MB
cancelled_1.pdf
pdf • 1.20 MB
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.03 MB
Tendernotice_1.pdf
PDF • 0.69 MB
1960TenderDoc.pdf
Tender Documents • 13.55 MB
BOQ_883750.xls
BOQ • 0.34 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .