CECRI 2019 22988 SUPPLY INSTALLATION TESTING AND COMMISSIONING OF ELECTROCHEMICAL WORKSTATION AND ALLIED ITEMS
2019_CSIR_38476_1
CECRI 2019 22988 ELECTROCHEMICAL WORKSTATION
Global Tenders
Equipments (Hospital / Lab)
Item Rate
90 days
KARAIKUDI
SEE CHAPTER 6 OF TENDER DOCUMENT
12 documents required · 12 mandatory
₹0
Exempted
21 Feb 2020
18 Dec 2019
6 Feb 2020
18 Dec 2019
5 Feb 2020
18 Dec 2019
Amount
SUPPLY, INSTALLATION, TESTING AND COMIISSIONING OF ELECTROCHEMICAL WORKSTATION (ANALYSER) AND OTHER ITEMS
SPECIFICATIONS FOR ELECTROCHEMICAL WORKSTATION Bipotentiostat + Galvanostat with Electrochemical Impedance Analyzer.
MAKE QUOTED TO BE SPECIFIED
MODEL QUOTED TO BE SPECIFIED
WARRANTY TO BE SPECIFIED.
Maximum Compliance Voltage: ±12 Volts or better
Output Voltage: ±10 Volts or better
Measured Voltage Resolution: 10 μV or better
Maximum output current: ± 500 mA or more
Current range: ±10 pA to ± 500 mA or bette
Measured current resolution: 100 fA or better
Potentiostat rise/fall time: 1 μs or lower
Input Impedance of electrometer :1 TΩ or better
Bandwidth of electrometer: 5 MHz or higher
Impedance analyser frequency range: 10μHz-7MHz or better
Impedance analyser AC amplitude: 200 – 1000 mV
There should be a provision for Auxiliary input/output ports for interfacing with other devices.
Apart from the available techniques the software should allow building new techniques.
Interface: USB/Ethernet/serial port
Should work with 220 V 50 Hz AC supply.
Electrochemical methods Following minimum modules must be supplied with the Bipotentiostat:
Voltametric analysis: Open circuit potential measurement (OCP) staircase cyclic and linear sweep voltammetryNormal pulse Differential pulse Differential normal pulse Square wave voltammetry
Chrono methods: Chrono amperemetry and Chronopotentiometry
iR compensation: EIS based iR compensation
EIS Methods: Potentiostatic / Galvanostatic Staircase Mott-Schottky
A comprehensive software package for the control of electrochemical workstation should be provided for data acquisition and storage and data analysis.
The software should allow plotting of various formats and should allow export of ASCII datafiles for use of the data with other software.
TRAINING NUMBER OF PERSONS TO BE SPECIFIED
TRAINING NUMBER OF DAYS TO BE SPECIFIED
CONFIRMATION ON WHETHER all the software updates ARE be free of cost DURING WARRANTY SAY YES/ NO
SUITABLE COMPUTER OF THE FOLLOWING SPECIFICATIONS
MAKE QUOTED TO BE SPECIFIED
MODEL QUOTED TO BE SPECIFIED
WARRANTY TO BE SPECIFIED.
i5 processor or better 4 GB SD RAM or higher 500 GB HDD DVD RW drive 4 USB Ports 17” TFT Colour Monitor 101 Keys Keyboard Optical mouse.
COST OF THE ABOVE COMPUTER INCLUSIVE OF GST
Biologic Science Instruments Pvt Ltd
US Dollar
European Euro
British Pound
Australian Dollar
Swiss Franc
Japanese yen
Singapore Dollar
Portion of Quoted Rate in Currency
European Euro
stage.html
html • 0.04 MB
tech_eval.pdf
fin_eval.pdf
aoc.pdf
finance_41297.pdf
boq_comp_chart.xlsx
xlsx
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.04 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .