RequiredOEM Annual Turnover Bid to RA enabled No Note: 1. OEM Turn Over Criteria: The minimum average annual financial turnover of the OEM of the offered product during the last three years
Requiredending on 31st March of the previous financial year
Requiredshould be as indicated in the bid document. Documentary evidence in the form of certified Audited Balance Sheets of relevant periods or a certificate from the Chartered Accountant/Cost Accountant indicating the turnover details for the relevant period shall be uploaded with the bid. In case the date of constitution/incorporation of the OEM is less than 3 year old
Requiredthe average turnover in respect of the completed financial years after the date of constitution shall be taken into account for this criteria. Steam Sterlizer ( 1739 pieces ) Technical Specifications 1/9 * As per GeM Category Specification Specification Specification Name Values Bid Requirement (Allowed Values) Construction Type of Steriliser VERTICAL CYLINDRICAL Steam Steriliser Pressure type
RequiredSteriliser Chamber size: 300 mm (Diameter) x 500 mm (Length)
Requiredconforming to IS: 3829 (Part 3)/1985
Requiredreaffirmed 2001
Requiredwith amendment 1 & 2
Requiredfitted with Silicon door gasket( expansion type).Electric load of the unit shall be 4 KW. VERTICAL CYLINDRICAL Steam Steriliser Pressure type
RequiredSteriliser Chamber size: 300 mm (Diameter) x 500 mm (Length)
Requiredconforming to IS: 3829 (Part 3)/1985
Requiredreaffirmed 2001
Requiredwith amendment 1 & 2
Requiredfitted with Silicon door gasket( expansion type).Electric load of the unit shall be 4 KW. Steam supply to Steriliser Built-in electrically heated Steam Generator (made of SS 316 (AISI 316) with low water protection system)
Requiredit should work with 3 phase 400/440 V 50 Hz AC supply Built-in electrically heated Steam Generator (made of SS 316 (AISI 316) with low water protection system)
Requiredit should work with 3 phase 400/440 V 50 Hz AC supply Material: Steriliser parts; Chamber
RequiredJacket
RequiredDoor
RequiredDoor Ring and shall be made of SS 304 SS 304 Type of control Automatic: MICROPROCESSOR CONTROLLED
RequiredWITH PRESELECTED STERILISATION PROGRAMS (variable programs customisable as per Buyer’s requirement to deal with sterilisation and drying of wrapped/unwrapped loads at 121 & 134 degree C ) Automatic: MICROPROCESSOR CONTROLLED
RequiredWITH PRESELECTED STERILISATION PROGRAMS (variable programs customisable as per Buyer’s requirement to deal with sterilisation and drying of wrapped/unwrapped loads at 121 & 134 degree C ) CONSTRUCTION Instruction plate on Steam Steriliser advising user to use soft water for steam generation for long service life of Steam Steriliser YES * Whether a compatible water softener/reversed Yes Yes 2/9 osmosis based system to feed water to the steam generator unit is to be supplied with the steriliser: Provision of Lock on Steriliser Chamber door: Manual Lock Manual Lock Installation The site ready for installation equipped with water supply
Requiredelectric connection cabling and drainage outlet shall be provided by Buyer. The Seller shall be responsible to carry out for installation of the offered Steam Steriliser unit on turnkey basis i.e. Any civil work required for installation of the ordered equipment i.e. flooring tilling/plastering/Paint and/or wood work shall be Seller’s responsibility. The Seller shall also provide training to personnel of buyer. * Type of Door: Hinged type Hinged type Display facility for Sterilisation Cycle parameters Chamber Temperature
RequiredAlarm indicator
RequiredLow water indicator Chamber Temperature
RequiredAlarm indicator
RequiredLow water indicator Print facility for record generation of dates
Requiredtime
Requiredload
Requiredidentification number and operating parameters (temperature and holding time etc): Yes
Requiredport given to connect an external Printer Unit Yes
Requiredport given to connect an external Printer Unit Warranty Period 3 3
Required5 Or higher No of Doors in Steriliser 1 1 Whether Type Test Certificate from a Govt/NABL/ILAC approved testing Yes Yes 3/9 laboratory is available (If Yes
Requiredto be furnished by Seller to Buyer on demand) for the offered product? Type of user friendly Interface (for microprocessor controlled Sterilizer) to specify parameters for running Sterilization cycle: NA Keyboard interface
RequiredTouch screen interface
RequiredNA Data Transfer facility through USB/RS232 port to computer/laptop on real time basis No Yes
RequiredNo Air removal from Sterilisation chamber
Requiredafter loading the steriliser
Requiredby on for more pressure/vacuums pulsing for proper steam penetration YES * Integrated Shredder Facility Integrated Shredder (for shredding the load in Sterilisation chamber) Yes
Requiredwith Microprocessor controlled Yes
Requiredwith Microprocessor controlled
RequiredYes
Requiredwith Manually controlled Shredder unit
RequiredNot available Material: Shredder parts; SS 304 SS 304
RequiredSS 316 or SS 316L Power of Motor used for Shredding NA
Requiredno integrated Shredder unit offered Must be of >= 2 HP
Required>= 1000 RPM
Required220 V
Required50 HZ AC motor
Requiredand able to deal with shredding of bio medical waste land as per Steriliser capacity
RequiredNA
Requiredno integrated Shredder unit offered Accessories (made of SS 316) for loading/unloading and handling of Load Number of Rails: 0 0
Required2 Or higher Long Handle: 1 1 Sterilisation Carriages 1 0
Required1
Required2 Or higher Pull out Trays/Tanks 0 0 Floor loading carts 0 0
Required2 Or higher Transfer carriages: 0 0
Required1
Required2 Or higher 4/9 * Specifications highlighted in bold are the Golden Parameters. * Bidders may note that In respect of non-golden Parameters
Requiredthe specifications 'Values' chosen by Buyer will generally be preferred over 'Bid requirement ( allowed Values) by the Buyer.