Rate contract for MGPS works for the year 2026 27 installed in Govt Medical College Srinagar and Associated Hospitals for the year 2026 27
2026_PWDJK_309871_1
MHDS/TS /2026-27/08/e-tendering dt 13.05.2026
Open Tender
Repair and Maintenance Works
365 days
Govt Medical College Srinagar and Associated Hospi
as detailed in nit
4 documents required · 4 mandatory
₹500
Executive Engineer MHD Srinagar
₹25,000
13 May 2026
13 May 2026
6 Jun 2026
13 May 2026
6 Jun 2026
13 May 2026
This is a percentage-rate BOQ. The schedule below shows estimated quantities and rates. Bidders quote a single percentage above or below the total estimated cost.
| Sl No | Description | Qty | Unit | Est. Rate | Est. Amount |
|---|---|---|---|---|---|
| 1 | Gas Outlets/Inlets [for Oxygen, Vacuum, NO2,CO2 & Medical Air | - | - | - | - |
| 2 | Dismantling of damaged Gas outlets | - | job | - | - |
| 3 | Providing and Fitting of Gas Outlets/Inlets/Double lock outlets using skilled labour, all tools and tackle as required to carry out the job at the site. Outlets shall be of HTM 0201 BS standard, manufactured with a 165 mm length, Copper inlet pipe stub, which is silver brazed to the outlet body. The body shall be of one-piece brass construction. For positive-pressure gas services, the outlet shall be equipped with a primary and secondary check valve, and the secondary check valve shall have a break-safe mechanism, comply with EN 737 pressure test standards, and be rated at a minimum of 200 psi in the event the primary check valve is removed for maintenance. The outlet assembly should have separate colour coding for each service and shall include the corresponding gas-specific adaptors. All outlets shall be cleaned and degreased for medical gas service and factory-assembled and tested. The medical gas outlets should be of quick-connecting and panel/wall-mounted. | - | job | - | - |
| 4 | BPC Flow meter with Humidifier | - | - | - | - |
| 5 | Providing and Fitting of a Back Pressure Compensated flow meter for accurate gas flow measurement with the following features to be fitted on existing Oxygen Gas outlets points.Control within a range of 0 – 15 lpm (calibration within +/- 10%). It should meet strict precision and durability standards. The flow meter body should be made of brass or chrome-plated materials. The flow tube and shroud components should be made of clear, impact-resistant polycarbonate. Flow Tube shall have a large and expanded 0 – 5 lpm range for improved readability at low flows. Inlet filter of stainless-steel wire mesh to prevent the entry of foreign particles. The humidifier bottle shall be made of unbreakable polycarbonate material and autoclavable at 121 degrees Centigrade temperature. | - | job | - | - |
| 6 | Ward vacuum units | - | - | - | - |
| 7 | Providing and Fitting of 600 ml Ward Vacuum Units. It shall be of lightweight and compact. The unit should consist of a regulator, 600 ml Reusable collection jar, made of unbreakable polycarbonate material and fully autoclavable at 134 degree centigrade, bracket for mounting the jar assembly on the wall, along with required tubing of convenient length & diameter. The vacuum regulator shall be adjustable and have a vacuum gauge that indicates the suction supplied by the regulator. A safety trap shall be provided inside the jar to safeguard the regulator from overflowing. | - | job | - | - |
| 8 | Providing and fitting of high pressure tubing for vacuum Ward/Theater Units. | - | mtr | - | - |
| 9 | Theatre vacuum units | - | - | - | - |
| 10 | Providing and Fitting of 2000 ml Theater Vacuum units.It shall be of lightweight and compact. The unit should consist of a regulator, 2000 ml Reusable collection jar, made of unbreakable polycarbonate material and fully autoclavable at 134 degree centigrade, bracket for mounting the jar assembly on the wall, along with required tubing of convenient length & diameter. The vacuum regulator shall be adjustable and have a vacuum gauge that indicates the suction supplied by the regulator. A safety trap shall be provided inside the jar to safeguard the regulator from overflowing. | - | job | - | - |