Loading…
Loading…
| # | Company | Status |
|---|---|---|
| 1 | Admitted-Fee/PreQual/Technical/Finance 9TH FLOOR NORTH WING GODREJ ONE PIROJSHAHNAGAR VIKHROLI EAST MUMBAI MAHARASHTRA 400079 | MUMBAI SUBURBAN | MAHARASHTRA | 400079 | Admitted-Fee/PreQual/Technical/Finance |
| 2 | Admitted-Fee/PreQual/Technical/Finance C O RED BRICK A WING 12TH FLOOR UNIT NO 1204 PLOT NO C 70 G BLOCK BANDRA KURLA COMPLEX BANDRA EAST MUMBAI 400051 | MUMBAI | MUMBAI SUBURBAN | MAHARASHTRA | 400051 | Admitted-Fee/PreQual/Technical/Finance |
Tender Value
Refer Docs
Closing Date
9 Jul 2026, 11:00 amClosed
Director TMC
TATA MEMORIAL HOSPITAL, PAREL, MUMBAI-400012
Expression of Interest for CT Simulator System
2026_TMC_914808_1
TMH/TMC/2026-27/CPU/EO/0004
Open Tender
Equipments (Hospital / Lab)
EOI
TATA MEMORIAL HOSPITAL, PAREL
Please refer Tender documents.
2 documents required · 2 mandatory
₹0
Exempted
10 Jul 2026
25 Jun 2026
10 Jul 2026
25 Jun 2026
9 Jul 2026
25 Jun 2026
Gantry
Aperture 80 cm or more
Multi-slice ( > 16 slice per rotation)
Gantry tilt angle: +/-300
Inbuilt Lasers (accuracy ≤ 1 mm)
Controls – both local & remote locations
Patient's scannable range : ≥ 150 cm
Metal-free scannable range ≥140cm
Digital readouts
Gantry tilt angle accuracy
Patients couch vertical position accuracy ≤ 1mm
Patients couch horizontal position accuracy ≤ 1mm
X-ray systems
Anode size
Maximum Heat storage of anode ( ≥ 6MHU)
Maximum anode cooling rate
Continuous anode cooling rate
Focal spot size
Beam filtration
Generator ( ≥ 70 KW)
Data Acquisition
Type of detectors : photodiode and electronics both integrated
Detector aperture
Photon conversion efficiency: specify percentage
Dynamic range up to 105 dB
Patient support system (Indexed Couch)
Flat Carbon fibre Table Top
Weight bearing capacity 160 kg or more (Table top sagging ≤ 2 mm).
Vertical limits & accuracy ( ≤ 1mm)
Total travel
Horizontal limits & accuracy ( ≤ 1mm)
Speed
Controls
TG 66 compliance
Image Characteristics
High contrast Spatial Resolution (at 0 - 2% MTF with full FOV ≥15 lp/cm)
Low contrast Resolution (5mm or less at 0.3%, for 10 mm slice thickness)
Noise(<15%)
Image reconstruction algorithms
Image reconstruction time
CT number accuracy (0 ± 4 HU for water and -1000 ± 10HU for air)
Scan Specification
Scan kV – 80-140 kV
Scan mA – 30 to 400 mA or above.
Scan field view (at least 50cm) and extended field of view (at least 70 cm)
Facility for pilot scan: scan field size & its reconstruction time
Available scanning modes: Sequence acquisition modes / Spiral acquisition modes please specify
Hardware Feature (details to be submitted)
CPU & its architecture (specify details)
Hard disk storage capacity (≥1 TB)
Image memory size
Graphic overlay
Image Bus Bandwidth
Confirm optimisation for the equipment & software quoted
Software feature
Pan / zoom facility
Multiple image display
Image annotation facility
Readout of CT numbers
Window & leveling facility
Distance measurement facility
Manual Image Sequencing facility
Graphic user Interface
Metal artifact reduction required
Oncology Special Protocol
Software:(QTY-1)
4DCT data acquisition software
Retrospective and Prospective acquisition
Phased and amplitude based acquisition
Binning of various phases
Export to existing radiotherapy network/Treatment planning system, existing Linear Accelerators and other third party network/Treatment planning system,
Interface and essential software for four-dimensional pulmonary CT scan for both amplitude and phase-bin CT data acquisitions to reconstruct the MIP (Maximum Intensity Projection), MinIP (Minimum Intensity Projection) and AIP (Average Intensity Projection) CT Image datasets should be provided. Retrospective four-dimensional CT scan acquisition options should also be available.
Hardware (QTY 1 set) - for transfer from existing system to new system
Interface for Motion managment system and compatiblity with existing network, Treatment planning system and Linear Accelerator (QTY =1 sets )
Comprehensive Motion managment system (IR Camera,mounting assembly,signal cables,marker block,interface,workstation with software and hardware) compatible with existing network, Treatment planning system and Linear Accelerator: Vendor will need to transfer existing hardware (RGSC Version 2.0) from present diagnostic 4DCT to newplanned machine with full compatibility of the system with new machine.
Image Storage
Specify storage media ≥ 1 TB
Specify storage capacity ≥ 500,000 images
CT Simulation accessories:
Leadglass window (2mm lead equivalent, dimensions ≥ 1500mmX1000mm) (QTY-1)
One Set of 3 external Moving laser System (QTY-1 set)
QA devices
Latest available CAT PHAN or equivalent (Qty No:1)
Laser QA Tool (QTY-1 )
One set of Complete diagnostic QA tool kit including Laptop with license of analysing software and associated hardware, with Beam allignment test tool,focal spot test tool,high & low resolution/contrast, integrated KVp, mAs, Timer and filtration,Dose dose rate, R, etc. Pencil ion chamber with calibration certificate (Qty no;1) , connecting cables (20 metre, Qty no:2) .(Qty No:1 set)
Colour network laser printer(Qty No:1 )
Advanced electron density phantom (Preferably more than 14 (Including Lung, Liver, Breast, Bone, Brain, Aluminum , Stainless Steel, Titanium, and True water container) (Qty No:1 )
Other Information
Export capability: Ethernet & TCP/IP consistent
DICOM-RT Ready
Should be interfaced with Hospital Information system (HIS) including the ROIS and RAS
Should be interfaced with existing PACS system for Diagnostic scan images
Patient communication system
ROIS Connectivity
CT Injector system
Water Bath / Oven
Installation Requirement
Physical Dimension & Weight:
Electrical requirement:
Air Conditioning:
Room size requirement:
Financial
Please submit the technical data sheet along with the brochure as a tender document with reference to the quoted make & model.
Please specify footprint size & weight.
Predispatch inspection at factory will be required if desired by authorities.
Demo of the quoted model , will be required if so desired by the user, after the opening of the technical bid and prior to opening of financial bid. This is for technical evaluation.
Please visit the site to check whether quoted model can be accomodated in the available space.
Local Service Support: Should have local office and service support/service engineer for attending the breakdown calls.
Service tool kit hardware & software should be provided to Biomedical Engineering department.
All the software related to the system should be handed over to USER.
Additional Requirment
Warranty : 2 Years
Penalty Clause:
Delivery Schedule
Training of staff
Regulatory Requirement (Please enclose copies)
Other Information
If 2 years warranty, 8 years CMC and Decommissioning with Buyback offer of the old Simulator conditions are not adhered to, TMC reserves right to reject the offer.
Only Yes or No should be mentioned in compliance report and additional information should be provided in detail in the remarks column.
stage.html
html • 0.02 MB
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.03 MB
Tendernotice_1.pdf
PDF • 0.61 MB
TECHCTSIM.xls
Tender Documents • 0.06 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at [email protected].