RequiredExperience with Government
RequiredAnnual Turnover
RequiredCertificate (Requested in ATC)
RequiredOEM Authorization Certificate
RequiredOEM Annual Turnover Bid to RA enabled No Note: 1. OEM Turn Over Criteria: The minimum average annual financial turnover of the OEM of the offered product during the last three years
Requiredending on 31st March of the previous financial year
Requiredshould be as indicated in the bid document. Documentary evidence in the form of certified Audited Balance Sheets of relevant periods or a certificate from the Chartered Accountant/Cost Accountant indicating the turnover details for the relevant period shall be uploaded with the bid. In case the date of constitution/incorporation of the OEM is less than 3 year old
Requiredthe average turnover in respect of the completed financial years after the date of constitution shall be taken into account for this criteria. X-Ray Machine (500 MA) ( 1 pieces ) Technical Specifications 1/9 * As per GeM Category Specification Specification Specification Name Values Bid Requirement (Allowed Values) Standards Conformity to Certification CE European * Certification number and date - * Type test approval from AREB required Yes * AREB approval certificate number of Particular Model - * Confirmity to manufacturing standards ISO 13485 * Conformity to safety standards IEC 60601-2-54 or BIS equivalents * Performance Parameters Purpose For Radiography examinations * Type of Frequency generator High frequency generator High frequency generator Frequency generator output in kW 50 * KVP range in KV 40 -125 * Current output 500 mA or more at 100 kV * Provision of Anatomical programming radiography Yes Yes Over loading protection Yes * X-ray tube positioning Floor Mount Tube stand * X ray tube should be high speed rotating anode and must have dual focus Yes * Tube rotation in degree 120 * Size of large focal spot in mm Upto 2 mm * Size of small focal spot in mm Upto 1 mm * Anode Rotation Speed (minimal) per minute 2850 * Permanent Filtration 0.9 mm Al/75 kV * Anode heat storage 300 * 2/9 capacity Heat dissipation rate 707 * Automatic exposure control device provided Yes * Maximum Exposure time in milli second upto 4 upto 4 Digital display for KVP & mAs provided Yes * Range of tube movements in vertical planes in cm 150 * Range of tube movements in longitudinal in cm 306.3 * Range of tube movements in lateral planes in cm 150 * Electromagnetic brakes with fully counter balanced mechanism for tube movement Yes * Facility to display tube angulation Yes * Centering devices to be provided Yes * Maximum source to image distance (SID) in vertical position in cm 150 * Maximum source to image distance (SID) in horizontal position in cm 150 * Collimeter rotation in degree ±180 * Multileaf collimator having halogen lamp/bright light source
Requiredwith auto shut off provision of the light
Requiredshould be available Yes * Minimum intensity of light field lamp ≥150 lux ≥150 lux Full coverage size at a 100 cm SID(collimeter size) with ± 5 percent tolerances 430 x 430 * Type of table provided Floating table top * 3/9 Table top material with low attenuation and scratch resistant surface Carbon fiber Carbon fiber Table Length in mm with ± 1 percent tolerances 1800 * Table Width in mm with ± 1 percent tolerances 600 * Table Height in mm with ± 1 percent tolerances 650 * System should have longitudinal and horizontal table top movements Yes * Table should have provision for lateral imaging
Requiredwithout patient movement Yes Yes Range of movements for the table Transverse movements of + 10cm or more & Longitudinal movements+ 40 cm with electromagnetic brakes * Facilities of height adjustments Yes Yes Maximum Patient weight supported by table in kgs 200 * Foot switches for adjusting longitudinal/side to side movements
Requiredlocking
Requiredand light adjustment No No Movable Grid with grid ratio 10:1 or more * Anti-collision and safety devices provided No No Counterbalanced
Requiredadjustable height
Requiredwall and floor mounted
Requiredvertical Bucky Yes * SID range of radiographic application including chest radiography in 1000 to 1800 * 4/9 mm Provision of doing chest radiography without grid Yes Yes Power supply Three Phase AC ( 380-440 V
Required50/60 Hz) * Number of installations of the X- ray machines in Central/State/PSU Govt Hospitals (Note: Seller should furnish a performance certificate of the device to the buyer in case demanded after placement of the order) >3 <3
Required>3 OEM and Reseller (if supplier is re-seller) shall ensure uninterrupted availability of all spares for 10 years Yes * Availability of toll free facility for technical support maintened by OEM or authorized agencies Yes Yes User/Technical/Mainten ance manuals to be supplied in English in hard and soft copy Yes * Demonstration of equipment and training to be provided after completing supplies before acceptance Yes * Warranty in years 5 * Additional Parameters Patient fixing belts and compression device (for performing excretory urography
Requiredetc) No No Voltage stabilizer servo controlled in kVA 65 * Lead lining for doors Yes * Lead Aprons - Full body 2 each (0.5pb) * 5/9 and Skirt type (Each ) Light weight zero lead radiation protection aprons provided No No Thyroid shields supplied 4 * Gonad sheet Male supplied 2 * Gonad sheet Female supplied 2 * Wall mounted stand for radiation protection aprons supplied Yes * Lead glass window Supplied Yes * Footsteps for the table supplied Yes * Lateral cassette holder provided No No Reports The Principal Manufacturer must have direct Presence/approved service center In India Yes * Copies of reports and certifications to be furnished to buyer on demand at time of supplies Yes * QA
RequiredQC and other tests as per NABH shall be carried out by OEM or authorized agency of OEM Yes * Copies of all calibration and test reports to be submitted at the time of Supply of Machine Yes * * Specifications highlighted in bold are the Golden Parameters. * Bidders may note that In respect of non-golden Parameters
Requiredthe specifications 'Values' chosen by Buyer will generally be preferred over 'Bid requirement ( allowed Values) by the Buyer. Additional Specification Documents