RequiredOEM Annual Turnover Bid to RA enabled No Electro Surgical Unit With Vessel Sealing System ( 4 pieces ) Technical Specifications * As per GeM Category Specification Specification Specification Name Values Bid Requirement (Allowed Values) Electro Surgical Unit Configuration: Advanced Electro Surgical unit with Vessel Sealing System have Mono polar Cut & Coagulation
RequiredBipolar and Vessel Sealing technology
Requiredall integrated in single system. * 1/8 Display & Control: System have user friendly 3 independent/Integrated LCD touch screen/TFT feather touch screen (of at least 14 cm diagonally) display for mono polar
Requiredbipolar and Vessel sealing technology
Requiredwith back lit LCD for good visibility in operating room
RequiredEquipment have Microcontroller based device incorporating closed loop control Coagulation for all output modes in the microcontroller firm wire. * Equipment incorporates tissue sensing/tissue feedback technology at on real time basis
RequiredIt has the capability to convert electrical energy to radio frequency energy thus providing radiofrequency seal Yes * Tissue sensing/Feedback & Radio frequency sealing: Equipment incorporate tissue sensing/tissue feedback technology at on real time basis and have the technology to convert electrical energy to radio frequency energy thus providing radio frequency seal. Equipment also identify the tissue impedance and adjust power accordingly to get the desired surgical effect on various tissue types. * Vessel Fusion system: Must be able to seal artery
Requiredveins along with tissue bundle up to & including 7 mm in diameter & sealed vessels should be able to withstand at least 3 times the systolic blood pressure. * 2/8 Low Frequency Leakage: should be between 50-60 Hz * High frequency leakage (RF): Less than 100mA in mono polar and vessel sealing mode and less than 60mA in bipolar mode. * Port Connectivity: System have RS232
RequiredUSB
RequiredEthernet port or any other computer connectivity for on field software downloads upgrades and serviceability. * Setting recall facility: System have recall facility to recall the last setting used by user before turning off the power. * Bipolar resection facility: Bipolar resection facility: System have bipolar resection (under water cutting). * System is compatible with Argon
RequiredCUSA/Waterjet (Hydrojet Technology) & Da Vinci robotic systems: No No Control Modes: Foot control mode and Hand control mode (both) available for open as well as laparoscopy in vessel sealing. * Mounting Cart: Mounting cart manufactured by the Original Equipment Manufacturer (OEM) shall be supplied with the equipment. * Monopolar and Bipolar System have separate mono polar mode that gives optimum cutting and coagulation: Yes * System have twin mono polar socket (only one socket to be activated at a time) with independent power setting: Yes * 3/8 Universal adaptor socket: System have one universal adaptor socket for foot switch operation and one bipolar socket: System have one universal adaptor socket for foot switch operation and one bipolar socket: Or higher Cutting modes: Have minimum two cutting modes for Mono polar output: [ Pure cut for clean
Requiredprecise cut in general surgery having maximum power of 300-350W & Blend mode for cutting with homeostasis having maximum power of 200 W] * Coagulation modes: Have minimum two Coagulation modes with power of 120 W or more; [ Fulgurate/Forced mode for efficient non- contact coagulation in most applications & Spray mode should have effect of varying amplitude and frequency for coagulating large tissue areas with minimum depth of necrosis.] * Bipolar Modes: Minimum 3 bipolar modes with power of 95W or more. * On screen review facility: Equipment have On Screen review of error codes and diagnostics. * Safety feature: System have patient plate contact quality monitoring system giving audiovisual alarm and deactivating output if contact between patient and patient plate is not proper to eliminate The risk of patient burns. * Auto Bipolar start and stop along with User configurable auto start facility: Yes Yes 4/8 Delay time and Auto Stop impedances range: Yes Yes Vessel Sealer System (VSS) Power output of Vessel sealing system (VSS): 150-200W (minimum) * No Auto-Cut facility: VSS have the user control mechanism so that the surgeon can decide for a repeat seal before actually cutting it or even not cutting it. (No Auto-Cut) * Sealing Time: VSS should be able to Seal vessels or tissue bundles faster than any conventional devices (preferably within 2-4 seconds) * All Laparoscopic instruments should have facility for both 5 mm/10 mm and should have a blade for instant cutting – 02 nos Yes * VSS have following vessel sealing instruments: Curved Jaw Vessel Sealer 5mm 34 to 37 cm for lap surgery- 2 units
RequiredCurved Jaw Vessel Sealer 5mm 20 to 23cm for open surgery- 2 units
RequiredDouble action Jaw Vessel Sealer 5mm 34 to 37 cm for lap surgery- 2 units
RequiredCurved Jaw scissor type open surgery instrument with separate control for seal and cut - 2 units * Audio Alerts: Equipment have Alert Tones : Patient Pad monitoring Error Alert
RequiredReactivate/Re grasp Tissue Error Alert
RequiredInstrument Checking Alert
RequiredComplete or Incomplete Seal Alert & Malfunctioning or System Error Alert * Certification and Warranty Certification: European CE European CE System meet IEC60601-1
RequiredYes * 5/8 IEC60601-1-2 & IEC 60601-2-2 requirements: Warranty Period 5 5 Or higher Installation: Installation of the equipment shall be Seller's responsibility * * Specifications highlighted in bold are the Golden Parameters. * Bidders may note that In respect of non-golden Parameters
Requiredthe specifications 'Values' chosen by Buyer will generally be preferred over 'Bid requirement ( allowed Values) by the Buyer.