| 1 | | Handheld, Non-Contact Type Digital Infrared thermometer, Minimum temperature- -30 C, Maximum temperature- +50 C, Accuracy- 1 degree Celsius, Digital display of temperature, Should have a power source of non-rechargeable and or rechargeable batteries, Should have Fahrenheit and Celsius Conversion, Should have a Single point class II laser, Should have Low Battery indication, Maximum and Minimum Reading, Battery Backup under contineous use for minimum of 8 hours | 1 | Nos | Anish Shivendra 508126,AIIMS Bibinagar (Hyderabad Metropolitan | 30 |
| 2 | | Should have a soft fluorescent, glare free bulb, Power source of 200-240V 50Hz, M.S powder coated box with frame, Viewing area should attain temperature of 37 degrees - 39 degrees C easily and maintain it, Warranty period of 1 year | 1 | Nos | Anish Shivendra 508126,AIIMS Bibinagar (Hyderabad Metropolitan | 30 |
| 3 | | High quality PT-100 Sensor for better accuracy, Temperature range adjustable from 5 degrees C above ambient to 60.0 degrees C with Micro controller based temperature controller Micro-controller controls the internal temperature within � 0.5 degrees C, Display for set value and process value, Audio Visual Alarm, Castor wheels for ease of mobility, Stainless steel trays to ensure even temperature distribution, Directional air flow to assure quick recovery after door openings, Chamber illumination with fluorescent light | 1 | Nos | Anish Shivendra 508126,AIIMS Bibinagar (Hyderabad Metropolitan | 30 |
| 4 | | Quality Standard-Manufacturing should be compliant with ISO and IEC, Should be compliant with European CE Class IIA and or US FDA, Should be compact Bench top equipment, Should gently seal Blood Bag pilot PVC tubing with no haemolysis, using radiofrequency heating, Should be capable of making wide seal of at least 2 mm width with sealing time lessthan 5sec, Electrodes should be well protected by a cover to prevent blood splutter, Sealing trigger should be automatic-on sensing tube in the slot, Should have indicator lamp for sealing process
Should have tear-seal feature to make segments that can be easily separated by hand, Charger should be compatible with Input voltage 200-240V 50 Hz Single phase Ac | 1 | Nos | Anish Shivendra 508126,AIIMS Bibinagar (Hyderabad Metropolitan | 30 |
| 5 | | Should be able to strip blood from the PVC blood tubing which helps in proper mixing of blood with anticoagulant prior to making segments, Exterior structure-Stainless steel, anti-rust, Rollers-Two, Self-Centric Teflon rod, Internal spring Keeps the stripper in open position, Screw mechanism that allows adjusting the distance between the rollers, Should include the spring for automatic return slot and a blade to cut the tubes | 1 | Nos | Anish Shivendra 508126,AIIMS Bibinagar (Hyderabad Metropolitan | 30 |
| 6 | Blood Bank Weighing Scale | Should be able to accurately weigh blood and blood components, Easy Auto-conversion of weight to volume, Auto calibration and Accuracy-1gm or ml, Micro-processor based digital display of weight and volume, Compact model with Polycarbonate or ABS molded body, Tare provision to account for the weight of the blood bag, Power supply of 200-240 V | 1 | Nos | Anish Shivendra 508126,AIIMS Bibinagar (Hyderabad Metropolitan | 30 |
| 7 | | Check compliance document enclosed in this bidfor complete specifications | 1 | Nos | Anish Shivendra 508126,AIIMS Bibinagar (Hyderabad Metropolitan | 30 |
| 8 | | Universal for all types of standard gel cards used in blood serology, blood type routine test, erythrocyte washing, micro column agglutination, immunology and other tests, Specifications- DC Brushless motor, low noise, Microprocessor control system and Digital user interface screen with soft touch control panel, LCD or LED display for visualization of RPM and Time, Acceleration and Deceleration with additional two sequential sets of parameters, Audio-Visual alarm provided for Imbalance and End of Cycle, Electric lid interlock, over speed and imbalance protection, Ergonomically designed shape for ease of operation, Light weight yet robust design, Speed control through VFD and dynamic brake system, Specialized step-ladder centrifugation technology, different step-ladder could be designed for different reagent Maximum speed 1000 rpm-4000rpm, Max volume 12 Cards or 24 cardsTimer 0-99mins, Power supply AC 110-240V 50HZ, Open program settings 10-20 | 1 | Nos | Anish Shivendra 508126,AIIMS Bibinagar (Hyderabad Metropolitan | 30 |
| 9 | | Table top Incubator for 12 or 24-gel cards in racks CE certified, Universal for all types of standard gel cards used in blood grouping, cross-matching and immunohematology procedures Maintain Temperature Range of 37 degrees C. Temperature Uniformity is within + or - less than 0.5 degrees C,, Microprocessor temperature control system, Manual adjustment of time parameters and temperature
Digital interface screen for control and visualization of remaining incubation time and real temperature
Audible alarms to indicate completion of incubation time, Time ranges from 0 to 99 mins.
Power supply of 200-220V | 1 | Nos | Anish Shivendra 508126,AIIMS Bibinagar (Hyderabad Metropolitan | 30 |
| 10 | | Island type of work bench with minimum 2 seat capacity with accessories including plumbing and electrical fittings, Length of work bench-1 to 2 meters, Width and height of work bench-0.5 to 1 meter, Granite top, Provision for under bench storage with either drawers or shutters combination or both of minimum 2 in quantity, Work bench with anti-vibration pads on bottom level, Provision for sink in work bench of 0.4 to 0.6 meters length, made of stainless steel along with flexible waste pipe, Panel mounted water faucet for the sink with appropriate drainage system, Warranty period for a minimum period of 1 year | 1 | Nos | Anish Shivendra 508126,AIIMS Bibinagar (Hyderabad Metropolitan | 30 |