RequiredExperience with Government
RequiredAnnual Turnover
RequiredCertificate (Requested in ATC)
RequiredOEM Authorization Certificate
RequiredOEM Annual Turnover Bid to RA enabled No Hematology Analyzer ( 1 ) Technical Specifications * As per GeM Category Specification Specification Specification Name Values Bid Requirement (Allowed Values) Set 1 Usage: To count
Requiredmeasure and analyse red blood cells
Requiredwhite blood cells and platelets
Requiredto measure the amount of hemoglobin in the * 1/6 blood and within each red blood cell. " Haematology Analysis is based on the principle of counting and sizing: Yes * "Analyser provides complete blood cell counting including: 5 part WBC differential without Reties and NRBC enumeration capability 5 part WBC differential without Reties and NRBC enumeration capability
Required5 part WBC differential with Reties and NRBC enumeration capability Or higher Analyses Leucocytes in their native state through: Laser based scatter analysis Laser based scatter analysis Applicable analyses available: Specify the number of parameters analysed by the machine: 25 25
Required26
Required27
Required28
Required29
Required30 Or higher Upgradable to attach with fully automated slide- maker & slide stainer in future: No Yes
RequiredNo Capability to automatically enumerate
RequiredNucleated Red Blood Cells in the CBC Difference Mode and without additional reagents: No Yes
RequiredNo Facility for PC based data management with all scatter plots
Requiredhistograms and display and in print: Yes Yes Auto loader facility with mixer capable of loading min 90 samples at any time (with Open vial
RequiredPre dilute & Closed vial modes) STAT capability with Positive Bar code identification facility for samples loading/unloading available Yes Yes
RequiredNo 2/6 Facility for User defined rules & flagging limits: Yes * "Database capability of storing sets of results and graphics
Requiredplease specify sets of results and graphics storable in database: >= 20000 >= 20000 Or higher Facility for workload recording
Requiredavailable: Yes Yes
RequiredNo Have unlimited number of user-definable control files: Yes * Auto stop function in event of unacceptable control data
Requiredavailable: No Yes
RequiredNo Ability to transmit results to host computer: Yes Yes Throughput capacity of analyser in Primary mode (samples/hour): 80 80.0 - 200.0 Or higher Linearity of Platelet to be from 0 to 3000 * 1000 cells/micro litres: Yes * WBC linearity to be from 0 to 3000 * 1000 cells/micro litres: Yes * Testing mode Selection facility available for: CBC mode CBC mode
RequiredCBC plus Differential & Reticulocyte testing mode Or higher Analyses time for cytopenic samples (RBC
RequiredPlt and WBC): Yes * Directly measures MCV: Yes * Capability to differentiate between smaller RBCs and Larger Platelets: Yes Yes Have auto purge function in the analyser’s software: Yes * Power Backup facility: Through suitable UPS (using maintenance * 3/6 free batteries) ensuring at least 60 minutes power backup. Sufficient Quality control files available (which store >= 100 runs each XB analysis): Yes Yes Printer Facility available through Inbuilt printer or external printer unit (to be supplied along with equipment): Yes Yes Offered equipment unit to be supplied with sufficient consumables (with at least 5/6th of total shelf life) required for
Requiredsufficient to carry out haematological testing of samples (please specify number of samples): 500 500
Required1000 Or higher Operating condition: Capable of operating continuously in ambient temperature of 10 to 50 deg C and relative humidity of 15 to 90% in ideal circumstances. * Storage condition: Capable of being stored continuously in ambient temperature of 0 to 50 deg C and relative humidity of 15 to 90%. * CERTIFICATION
RequiredSTANDARDS AND SAFETY Certification of product available CE European CE European Equipment shall meet internationally recognized for Electromagnetic Compatibility (EMC) for electro medical equipment: 61326-1 Yes * Certified to be compliant with IEC 61010-1
RequiredIEC 61010-2-281
Required61010-2- Yes * 4/6 101 for safety: Manufacturer have ISO certification for quality standards: Yes Yes Warranty Period 3 3
Required5 Or higher * Bold specifications are the golden parameters.