Loading…
Loading…
| # | Company | Status |
|---|---|---|
| 1 | Accepted-Technical E 29 A ROAD NO 2 I P I AREA KOTA RAJASTHAN 324005 | KOTA | RAJASTHAN | 324005 | Accepted-Technical Qualified.Offer is as per NIT. |
| 2 | Accepted-Technical 96 ELECTRONIC CITY COMPLEX BANGALORE KARNATAKA 560100 | BENGALURU URBAN | KARNATAKA | 560100 | Accepted-Technical Qualified.Offer is as per NIT. |
| 3 | Rejected-Technical NO 69 2 N R NIVAS GROUND FLOOR SHARADAMBA NAGAR RAGHAVENDRA TEMPLE ROAD JALAHALLI BANGALORE KARNATAKA 560013 | BENGALURU URBAN | KARNATAKA | 560013 | Rejected-Technical Technically disqualified. |
| 4 | Rejected-Technical 1ST FLOOR FLAT NO 1C BLOCK ORION ENCLAVE 186 MAHATMA GANDHI ROAD KOLKATA KOLKATA WEST BENGAL 700082 | KOLKATA | WEST BENGAL | 700082 | Rejected-Technical Technically disqualified. |
Tender Value
Refer Docs
EMD Value
₹71,700
Closing Date
25 Jul 2026, 3:00 pmClosed
General Manager,KTPS.
Administrative Building, Kolaghat Thermal Power Station, P.O- Mecheda, Dist-Purba Medinipur, PIN-721137.
Procurement of Complete set of Oxygen Analyser used in Air Pre Heater Inlets and Outlets of U -3,4,5,6 under C and I dept, KTPS.
2026_WBPDC_1026895_1
WBPDCL/KTPS/NIT/E3883/25-26
Open Tender
Miscellaneous Goods
Buy
90 days
Kolaghat Thermal Power Station
Please refer Tender documents.
2 documents required · 2 mandatory
₹0
₹71,700
Yes
17 Oct 2026
15 Jun 2026
28 Jul 2026
15 Jun 2026
25 Jul 2026
22 Jun 2026
| Sl No | Description | Qty | Unit | Est. Rate | Est. Amount |
|---|---|---|---|---|---|
| 1 | INSITU ZIRCONIA OXYGEN ANALYZER FOR OXYGEN
ANALYSER O2 ANALYZER,PROBE WITH CONVERTER SIGNAL
OUTPUT 4-20MA DC,DISPLAY RANGE 0 -100%,HART
Additional Data:
Complete set of zircinia base oxygen analyser:
Probe Length:2 meter,
Materials:ss316,
Flange:DIN PN10 /ANSI150-DN50A SUS 304
Probe OD:2''.
Probe protector: Material:ss316
Flange:DIN PN10/ ANSI150-
DN50A SUS 304(COUNTER FLANGE SHOULD BE PROVIDED BY VENDOR) OD:3'',ID:2.5'',Length:2050 mm.Out put signal:4-20 ma dc, Power supply:240 v ac,50 hz. Connecting wires between analyser to converter need to be provided if required, Cable should be minimum 50 metre in length. Transmitter Option : Standard ZIRCONIA BASED, System Type : Remote Probe Type : Standard , No Automatic Calibration, Probe insertion Length : 2.0 metre, Thermocouple
Type : Type k, Standard Cell ;,, Accuracy : 0.05% O2,Response Dead Time : 3 sec., Response rate :T90 is less than 10 seconds, Zero drift & Span drift : < ± 0.2% typically,
Output: 4-20mA. | 4 | SET | 3,28,685 | ₹13,14,740 |
| 2 | OXYGEN ANALYSER,ZIRCONIA BASED
Additional Data:
Complete set of zircinia base oxygen analyser: Probe Length:1 meter,Materials:ss316, Flange:DIN PN10/ANSI150- DN50A SUS 304 Probe OD: 2''. Probe protector: Material:ss316 Flange:DIN PN10/ ANSI150- DN50A SUS 304
(COUNTER FLANGE SHOULD BE PROVIDED BY VENDOR) OD:3'',ID:2.5'',Length:1050 mm.Out put signal: 4-20 ma dc, Power supply:240 v ac,50 hz. Connecting wares between analyser to converter need to provide if required, Cable should be minimum 50 metre in length.Transmitter Option : Standard ZIRCONIA BASED, System Type : Remote Probe Type : Standard , No Automatic Calibration, Probe insertion Length : 1.0 metre, Thermocouple Type : Type k, Standard Cell ;,<(>,<)> Accuracy : 0.05% O2, Response Dead Time : 3 sec., Response rate :T90 is less than 10 seconds, Zero drift & Span drift : <(><<)> ± 0.2% typically, Output: 4-20mA. ) | 8 | NO | 2,83,585 | ₹22,68,680 |
stage.html
html • 0.02 MB
tech_eval.pdf
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.03 MB
Tendernotice_1.pdf
PDF • 2.03 MB
BOQ_2522624.xls
BOQ • 0.28 MB
07DateCORRIGENDUM2.pdf
PDF • 0.69 MB
07DateCORRIGENDUM1.pdf
PDF • 0.71 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .