Single zone Tube Furnace with 1200 degC Temperature | The furnace should have Max temp of 1200 deg C, Continuous Operation-1100 degC, Ramp rate-1- 5 degC/min, Maximum acceptance of work tube with O/D 60mm,Heat up time 48min or less ,Tube Heated Length-450mm,Uniform temperature (plus/minus 5 degC) in tube length -315 mm,Thermocouple type-N Type,Maximum Power-2000W or Less,Power Supply-220 -240 V single phase,Controller should be a part of the furnace and should not be a separate Unit. It should be Eurotherm or Honeywell make, or similar quality make with 4.3inch colour touch screen with Setpoint control, Program profile control, 10 unique program profiles saved in memory, 24 segments per unique program, Ethernet communication, Data logging to a USB memory stick in a .csv file format, Real time clock, Program schedule start, Program status indication with estimated end time and date, Event indication, Control power indication and User level security, Single point temperature calibration. Inert atmosphere accessories to be provided with Extended length Quartz work tube 900 mm with 55 I/D and 60 O/D, 2 x insulation plugs for extended length work tube, Work tube end seals with 1 x gas inlet/outlet pipe, 1 x gas inlet/outlet pipe with thermocouple gland, Probe thermocouple access with Up to 1200 degC, Diameter 3 mm gland and blanking plug to suit optional type N probe thermocouple, Work Tube Support , Flow Meter with the advanced inert gas package including Advanced inert gas package, with solenoid valve switched by program segment output Nitrogen N2- 1-8 l/min, Gas inlet equal to 6 mm outside diameter push in fitting, Manual on/off ball valve, Pressure relief valve, Pressure gauge, Electrically operated valve. The spare parts to be provided by vendor include 2 pair of gloves,5 KVA Servo Stabilizer, Tong to Remove sample, Quartz crucible 25 ml and 15 ml 2 No Each,Borosilicate glass Desiccator with porcelain plate-200 mm, N2 Cylinder with two stage Regulator. |