RequiredExperience with Government
RequiredAnnual Turnover
RequiredCertificate (Requested in ATC)
RequiredOEM Authorization Certificate
RequiredOEM Annual Turnover Bid to RA enabled No Automated Hematology Analyser ( 5 Part ) ( 1 pieces ) Technical Specifications * As per GeM Category Specification Specification Specification Name Values Bid Requirement (Allowed Values) Standards Certifications for the product CE European (with verified number) CE European (with verified number) Certification number and date - * Confirmity to quality ISO 9001 & ISO 13485 ISO 9001 & ISO 13485 1/9 management standards Confirmity to electrical safety standards IEC 61010-1
RequiredIEC 61010-2-081
RequiredIEC 61010-2-101 or BIS equivalent * Performance Parameters UMDNS Name and Code "Analyzers
RequiredLaboratory
RequiredHematology
RequiredCell Counting
RequiredAutomated [17-741] Analyzers
RequiredLaboratory
RequiredHematology
RequiredCell Counting
RequiredSemiautomated [17-742] Analyzers
RequiredPoint-of-Care
RequiredWhole Blood
RequiredHematology [18-513] Computer Aided Detection Systems
RequiredPathology Slide
RequiredBlood Cell Classification [24-248]" * Clinical Purpose Hematology Analyzers are used to count and characterize blood cells for disease detection and monitoring. Basic analyzers return a complete blood count(CBC) with a 5-part differential white blood cell(WBC) count.Sophisticated analyzers measure cell morphology and can detect small cell populations to diagnose rare blood conditions. * Type of Configuration benchtop * Type of system offered Open system Open system Type of automation fully automatic * Equipment should have automatic start up
Requiredshut down and sample analysis Yes * Analysis principle LASER based scatter analysis LASER based scatter analysis Type of cell counting "5 part WBC differential without Retic and NRBC enumeration capability" "5 part WBC differential without Retic and NRBC enumeration capability" 2/9 Should have multi channel analyssi for better resolution and reproductability Yes * Analysis available WBC
RequiredLympho#
RequiredLympho %
RequiredNeutrophil#
RequiredNeutrop hil %
RequiredBasophil#
RequiredBasoph il%
RequiredEasinophil #
RequiredEasino phil%
RequiredMonocyte#
RequiredMono cyte%
RequiredRBC
RequiredHGB
RequiredMCV
RequiredP CV
RequiredMCH
RequiredMCHC
RequiredRDW- SD
RequiredRDW CV
RequiredPLT
RequiredMPV
RequiredP DW
RequiredPLT
RequiredPLCR
RequiredPLCC
RequiredHCT
RequiredPCT
RequiredRET#
RequiredRET%
RequiredNRBC #
Requiredsynovial
Requiredser ous) parameters like WBC-BF
RequiredRBC-BF
RequiredMN #
RequiredMN%
RequiredPMN%
RequiredPMN#
RequiredTC-BF# No No IPF analysis No No PLT-0 Analysis No No PLT-F Analysis No No IG % analysis No No IG# analysis No No RET-HE analysis No No HFR analysis No No LFR analysis No No MFR analysis No No IRFanalysis No No Discrete analysis modes available CBC
RequiredCBC+DIFF CBC
RequiredCBC+DIFF Analysis method for WBC Tri angle Laser flow cytometry Tri angle Laser flow cytometry method for platelet measurement electrical impedance electrical impedance RBC Measurement method electrical impedance electrical impedance Hb measurement cyanide free colorimtery cyanide free colorimtery Retic measurement not provided not provided no of sample racks to cater to different tube sizes 6 6 3/9 Maximum sample Aspiration volume needed in all modes 20µl or less 20µl or less Minimum sample volume required in all modes 20µl or less 20µl or less Throughput capacity in CBC/Differential 71-80 71-80 Throughput capacity in Retic mode NA NA Type of modes of sample running Closed vial Closed vial Linearity of Platelet " 0 to 3000 * 10 3 cells micro litres" * RBC Linearity 0 to 8 x 10^6 per micro litre or more * Hemoglobin linearity 0 to 25 gm per litre 0 to 25 gm per litre Rectic linearity NA NA WBC linearity " 0 to 300 * 1000 cells micro litres" * Directly measures MCV Yes * Time taken by the analyser to produce the test results(Analysis time) in seconds 40-60 * Availability of Auto dilution No No should have clot detection facility No No Automatic probe wipe Yes * Extended Analyses for cytopenic samples Yes * Number of reagents 5 * Quality assurance system with calibration and controls Yes * Atleast 3 quality control programs Yes * Type of Calibration Both manual and automatic Both manual and automatic Direct aspiration for capillary blood from finger prick Yes Yes Floating discriminator for platelets and RBC counting for reliable Yes * 4/9 RBC and PLT data Separate diluting nozzles for RBC and WBC Yes Yes Double bathing mechanism Yes Yes Upgradable to integrate with attachment No No Data Management and Display Type of data management PC based PC based Display LCD LCD PC hard disk >500 GB >500 GB Inbuilt monitor size in inches NA (for PC based) NA (for PC based) Processor of PC Provided with system I 5 or more I 5 or more RAM 4 GB 4 GB HDD 500 GB 500 GB PC Monitor size in inches if PC provided externally 17 INCHES OR MORE 17 INCHES OR MORE The processor and RAM of the board system should be latest version Yes * HIS/LIS Interface HL7 HL7 Type of external storage USB USB No of USB Port provided 4 4 Printer facility Yes Yes Type of printer unit External External Printer type LASER printer LASER printer Display and print provided scatter plot and histograms * Auto loader facility Yes * Facility for user defined flagging Yes * "Database capability of storing sets of results and graphics" >= 50000 >= 50000 L J Plot facility Yes Yes 5/9 delta check for cumulative review Yes Yes On line QC option No No patient moving average Yes Yes QC File management Yes Yes Facility for workload recording Yes * "Flagging in event of unacceptable control data" Yes * Ability to transmit results to host computer Yes * Make of PC provided (Reputed make PC to be provided ) - * Type of user Interface or data entry all three all three Have auto cleaning function in the analyser’s software Yes * Power Requirements Type of power supply in VAC
RequiredHz 100-240 VAC
Required50-60 Hz * Power Backup facility UPS * Type of UPS Offline * Rating of UPS in KVA 1 KVA 1 KVA Back up time in minutes 30 min 30 min AVAILABILITY OF Sufficient Number of Quality control fileswhich store 100 or more XB Analysis Yes * ACCESSORIES
RequiredSPARE PARTS AND CONSUMABLES "Offered equipment unit to be supplied with sufficient consumables (with at least 2/3rd of total shelf life) required for
Requiredsufficient to carry out haematological testing of samples" 500 500 Should have on board regent facility and automatic reagent Yes * 6/9 inventory management Net work integration with lab information system feature Yes Yes Operating temperature and humidity "Capable of operating continuously in ambient temperature of 10 to 35 deg C and relative humidity of 15 to 85% in ideal circumstances" * Miscellaneous Parameters Warranty in years 3 3 Number of installations in Central/State/PSU Govt Hospitals ( Hint: Seller should supply a performace certificate of the device to the buyer incase demaneded after placement of order) Less than 3 Less than 3 OEM/Reseller ( if supplied by reseller shall ensure uninterputed availabilty of all spares for 10 years) Yes * Availability of toll free facility for technical support maintened by OEM or authorised agencies Yes Yes During the warranty period it shall be ensured that all the break down calls are attended with in 48 hrs and the complete details of service agents/contact details should be furnished to buyer and consignee at time of supplies Yes * User/Technical/Mainten ance manuals to be supplied in English in hard and soft copy Yes * Details of equipments and procedures required for local Yes * 7/9 calibration and routine maintenance to be supplied and advanced maintenance task documentation also to be furnished List of important spares and accessories
Requiredwith their part numbers to be supplied to the buyer at the time of supplying the equipment Yes * Installation and Demonstration of equipment and training to be provided after completing supplies before acceptance Yes * The Principal Manufacturer must have direct Presence/approved service center In India Yes * Copies of reports and certifications to be furnished to buyer on demand at time of supplies Yes * Calibration certificates as per NABH requirement Yes * * Bold specifications are the golden parameters. * Bidders may note that In respect of non-golden Parameters
Requiredthe specifications 'Values' chosen by Buyer will generally be preferred over 'Bid requirement ( allowed Values) by the Buyer.