RequiredOEM Authorization Certificate Bid to RA enabled Yes Steam Sterlizer ( 2 pieces ) Technical Specifications * As per GeM Category Specification Specification Specification Name Values Bid Requirement (Allowed Values) Construction Type of Steriliser HORIZONTAL RECTANG ULAR/SQUARE- Steam Steriliser Pressure type for Hospital and Pharmaceutical use
Requiredconforming to IS: 3829(part 1)/1999
Requiredfitted with Silicon door gasket( expansion type). Steriliser Chamber size (L x W x H): 1200 mm x 600 HORIZONTAL CYLINDRICAL Steam Steriliser Pressure type for Hospital and Pharmaceutical use
RequiredSteriliser Chamber size: 400 mm (Diameter) x 1100 mm (Length)
Requiredconforming to IS: 3829(part 1)/1999
Requiredfitted with Silicon door gasket( expansion 1/6 mm x 600 mm
RequiredElectric load of the unit shall be 18 KW. type). Electric load load of the unit shall be 9 KW
RequiredHORIZONTAL RECT ANGULAR/SQUARE- Steam Steriliser Pressure type for Hospital and Pharmaceutical use
Requiredconforming to IS: 3829(part 1)/1999
Requiredfitted with Silicon door gasket( expansion type). Steriliser Chamber size (L x W x H): 1200 mm x 600 mm x 600 mm
RequiredElectric load of the unit shall be 18 KW. Steam supply to Steriliser From Built-in Steam Generator electrically heated using immersion heaters wired for operation on 3 phase 4 wire 400/440 V 50 Hz AC supply [with immersion heaters having mineral filled sheathed heating elements conforming to IS: 4159/2002 duly BIS marked and complying with safety requirements as per IS : 302-2-201(1992)] From Built-in Steam Generator electrically heated using immersion heaters wired for operation on 3 phase 4 wire 400/440 V 50 Hz AC supply [with immersion heaters having mineral filled sheathed heating elements conforming to IS: 4159/2002 duly BIS marked and complying with safety requirements as per IS : 302-2-201(1992)] Material: Steriliser parts; Chamber
RequiredJacket
RequiredDoor
RequiredDoor Ring and shall be made of SS 316 or SS 316L SS 304
RequiredSS 316 or SS 316L Type of control Automatic: MICROPROCESSOR CONTROLLED
RequiredWITH PRESELECTED STERILISATION PROGRAMS (variable programs customisable as per Buyer’s requirement to deal with sterilisation and drying of wrapped/unwrapped loads at 121 & 134 degree C ) Automatic: MICROPROCESSOR CONTROLLED
RequiredWITH PRESELECTED STERILISATION PROGRAMS (variable programs customisable as per Buyer’s requirement to deal with sterilisation and drying of wrapped/unwrapped loads at 121 & 134 degree C ) CONSTRUCTION Instruction plate on Steam Steriliser YES * 2/6 advising user to use soft water for steam generation for long service life of Steam Steriliser Whether a compatible water softener/reversed osmosis based system to feed water to the steam generator unit is to be supplied with the steriliser: Yes Yes Provision of Lock on Steriliser Chamber door: Manual and Automatic Locks
Requiredboth Manual and Automatic Locks
Requiredboth Installation The site ready for installation equipped with water supply
Requiredelectric connection cabling and drainage outlet shall be provided by Buyer. The Seller shall be responsible to carry out for installation of the offered Steam Steriliser unit on turnkey basis i.e. Any civil work required for installation of the ordered equipment i.e. flooring tilling/plastering/Paint and/or wood work shall be Seller’s responsibility. The Seller shall also provide training to personnel of buyer. * Type of Door: Radial Type Radial Type Display facility for Sterilisation Cycle parameters Chamber Temperature
RequiredBatch Number
RequiredTime and date
RequiredAlarm indicator
RequiredLow water indicator Chamber Temperature
RequiredBatch Number
RequiredTime and date
RequiredAlarm indicator
RequiredLow water indicator Print facility for record generation of dates
Requiredtime
Requiredload
Requiredidentification number and operating parameters Yes
Requiredport given to connect an external Printer Unit Yes
Requiredport given to connect an external Printer Unit 3/6 (temperature and holding time etc): Warranty Period 5 5 Or higher No of Doors in Steriliser 1 1 Whether Type Test Certificate from a Govt/NABL/ILAC approved testing laboratory is available (If Yes
Requiredto be furnished by Seller to Buyer on demand) for the offered product? Yes Yes Type of user friendly Interface (for microprocessor controlled Sterilizer) to specify parameters for running Sterilization cycle: Keyboard interface Keyboard interface Data Transfer facility through USB/RS232 port to computer/laptop on real time basis Yes Yes Air removal from Sterilisation chamber
Requiredafter loading the steriliser
Requiredby on for more pressure/vacuums pulsing for proper steam penetration YES * Integrated Shredder Facility Integrated Shredder (for shredding the load in Sterilisation chamber) Not available Not available Material: Shredder parts; SS 304 SS 304 Power of Motor used for Shredding NA
Requiredno integrated Shredder unit offered NA
Requiredno integrated Shredder unit offered Accessories (made of SS 316) for loading/unloading and handling of Load Number of Rails: 2 2 Or higher Long Handle: 1 1 Sterilisation Carriages 1 1
Required2 Or higher Pull out Trays/Tanks 2 2 Floor loading carts 2 2 Or higher 4/6 Transfer carriages: 1 1
Required2 Or higher * Specifications highlighted in bold are the Golden Parameters. * Bidders may note that In respect of non-golden Parameters
Requiredthe specifications 'Values' chosen by Buyer will generally be preferred over 'Bid requirement ( allowed Values) by the Buyer.