Executive Engineer MHD, Srinagar
Office of Executive Engineer, MH Division Srinaga
Improvement of Medical Air by way of providing and Fitting of pressure regulator at SICU(Ward-17) and PICU (Ward-10), Automatic Medical Gas Alarm system in ENT OT and other minor works of SMHS Hospital , Srinagar
2024_PWDJK_264463_2
MHDS/TS /2024-25/119/e-tendering, Dated13-11-2024
Open Tender
Machineries/ Mechanical Engg Items
Item Wise
5 days
SMHS HOSPITAL
as detailed in the nit
3 documents required · 3 mandatory
₹100
Executive Engineer MHD SGR
₹1,640
2 Jan 2025
13 Nov 2024
4 Dec 2024
13 Nov 2024
4 Dec 2024
14 Nov 2024
Amount
Improvement and augmentation of Medical Air by way of providing and fitting of high air pressure regulator for Medical air pipe line in SICU (ward-17) and PACU (ward-10). The regulator shall be installed in medical air pipe line by means of brazing with silver electrode. The job further includes providing and fitting of necessary copper fittings complete in all respects. The regulator shall be confirm to ISO 6953-1, non- raising "press to lock " adjusting knob for locking at any set pressure, diaphragm operated, relieving type and testing thereof.
Providing and fitting of pressure switch having sensing element: copper alloy, differential: 0.009-0.4bar, range: (-)1...0 bar for vacuum pump installed at Burn centre
Providing and fitting of 63 Amp. 04 pole MCCB with breaking capacity of 16 KA at 240 volts and 10KA at 415 volts for vacuum pump.
M/s THE ENTERPRISES (BID ID -2270506)
tech_eval.pdf
boq_comp_chart.xlsx
xlsx
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .