Loading…
Loading…
| # | Company | Amount | Rank | Status |
|---|---|---|---|---|
| 1 | 1₹2.3 LAccepted-AOC 3724 OPP KALI BARI MANDIR OPP KALI BARI MANDIR AMBALA CANTT HARYANA 133001 | AMBALA | HARYANA | 133001 | 1 | Accepted-AOC alloted L-1 | |
| 2 | 2₹9.3 L+₹7.0 L (301.4%)Rejected-AOC | 2 | Rejected-AOC L-2 | |
| 3 | 3₹12.5 L+₹10.2 L (441.9%)Rejected-AOC | 3 | Rejected-AOC L-3 |
Tender Value
₹9.6 L
EMD Value
₹19,135
Closing Date
6 Jul 2020, 3:00 pmClosed
Executive Engineer
PHED Fatehabad
Supplying of Jar test apparatus Laminar Air Flow Chamber HOT AIR OVEN Bench top pH Tester Benchtop TDS meter BenchTop Turbidity Meter Dissolved Oxygen Meter Colony Counter for District Lab at Public Health Engineering Division
2020_HRY_132954_1
9.57 Laboratory Equipment Fatehabad
Open Tender
Civil Works
Works
90 days
Fatehabad
Please refer Tender documents.
2 documents required · 2 mandatory
₹1,000
Yes
Engineer in Chief
₹19,135
Yes
1 Oct 2020
22 Jun 2020
6 Jul 2020
22 Jun 2020
6 Jul 2020
22 Jun 2020
Amount
N.S. Item
Jar test Apparatus - Make-Electronics India/ Iso certified Company Specifications • Accepts six samples (40,000) • LED Digital Display • Digital indication of speed and timer • Digital timer from 0 to 99 minutes • Two pre-set programs • Built in illumination for easy viewing of sample • Speed range 25 to 200 rpm • Power-230V,50Hz,100 W
HOT AIR OVEN -Make Iso certified Company (Specifications) • Inner Chamber SIZE:18”X18”X18”inch(LxWxD) • Double walled inner chamber available in stainless steel with grade304. • Outer chamber made of Stainless Steeel • Glass wool/ PUF insulation to prevent temperature loss. • Outer double wall door is provided with gas kit. • Stainless steel removable & adjustable shelves . • No. of Shelves-3nos • Digital display with mains ON/OFF, switch indicating lamps & Fuse • Digital microprocessor based PID Temp controller with LED Display • accuracy of temperature +/- 0.5 degree • Temperature range adjustable from Room Temp. to2600C (resolution 0.10C) • Air flow systems providing Blower for even distribution of temperature.tt • Safety thermostat to prevent over heat. • Electric Supply:220/230V AC,50/60 Hz,2500watt
BennchTop Turbidity Meter -Make Eutech /Hach Specifications • Turbidity Range 0 to 1000 NTU • Resolution & Accuracy 0.01 NTU (0 to 9.99 NTU); 0.1 NTU (10 to 99.9 NTU); 1 NTU (100 to 1000 NTU) & ± 2% of measured range + 0.01 NTU • No. of Calibration Points Up to 3 points • Calibration Options 4 points (0.02, 10, 100, 1000) • Light Source White light (tungsten) • Response Time < 6 seconds • Power 12 VDC, using AC adapters 110 VAC or 220 VAC, 50/60 Hz
Laminar Air Flow Cabinet Horizontal Specification • Working Chambers SIZE: 3`x2`x2` feet • Work Bench made of stainless steel with grade 304. • Outer chamber made of Wooden Duro Board. • Hepa having accurate performance rating as high as 99.97% with cold dop and -99.97% with hot dop. • Air quality should meet ISO Class 5 standards. • Front Door made of Acrylic sheet. • Fluorescent light : Minimum200lux intensity • This blower is fitted with ¼ h.p motor (ISI MARKED) and operates at minimum noise. • UV Lamp: Phillips Holland. • Heavy duty Blower Dynamically Balanced • Pressure : Manometer 0-50mm range • Air velocity:0.5m/s • ISO/ISI Certified Company
Colony Counter -Make Electronics India/Others Specifications • 6 Digit Display • 7-Segment LED Display • Dish Size-110 mm • Magnification- x1.7 surface magnification • Power-230 V ± 10% AC, 50 Hz • Dimension-225 x 250 x 280 mm (L x B x H) (Approx. • Auto Marker Pen • Audible Confirmation of Each Count • Uniform Glare Free Illumination • Wolffhugel Glass Grid with Focusing Facility • LED Based Illumination
TDS Meter -Benchtop Make-Eutech/Hach/Others Specifications •TDS Range 0.000 to 99'999 ppm •Resolution ±0.01% Full Scale •Accuracy ±0.5% Full Scale + 1 LSD •Temperature Range -5.0 to 105.0°C (23.0 to 221.0°F) (Meter) •Resolution 0.1°C (0.1°F) •Accuracy ±0.3°C (±0.5°F) •Calibration Options Automatic (single or multi-point up to 4 points) &manual (up to 5 points - 1 per range) •TDS Factor 0.40 to 1.00 •Cell Constant 0.1, 1.0, 10.0 •Display Large Custom LCD - 105 X 75 mm (screen size) with contrasted background •Memory 100 Data Sets •Special Functions Auto-Calibration; Stability Indicator, Replatinising Circuit & Self-diagnostic •Power Requirement 110VAC/ 220VAC, centre negative
BenchTop-pH Meter-Make Eutech/Hach/Others Specifications •pH Range -2.00 to 16.00 pH •Resolution 0.01 pH •Accuracy ±0.01 pH •Cal. Points Up to 5 Buffer Sets USA (1.68, 4.01, 7.00, 10.01, 12.45), NIST (1.68, 4.01, 6.86, 9.18, 12.45) ORP •Range ±2000 mV •Rel. mV Range ±2000 mV •Resolution 0.1 mV (±199.9 mV) / 1 mV (beyond) •Accuracy ±0.2 mV (±199.9 mV) / ±2 mV (beyond) Temperature •Range (Meter) 0.0 to 100.0 ºC / 32.0 to 212.0 ºF •Resolution 0.1 ºC / 0.1 ºF •Accuracy ±0.3 ºC / ±0.5 ºF (0 to 70 ºC) •Compensation ATC / MTC (0 to 100 ºC) (pH only) •Calibration Offset in 0.1 ºincrements; •offset range: ±5.0 ºC / ±9.0 ºF •Power Requirements 9 V DC adapter, 1.3 A (100/240 VAC, SMPS)
Dissolved Oxygen Meter with Optical Probe Make Hach/Others Specifications Data Memory : 500 results Data Storage : GLP/ISO compliant reading data stored with calibration details.Calibration details and check standard readings documented in data log. Automatically store in "press to read" mode and interval measurement mode. Manually store in "continous read" mode. Display Lock Function : Continuous measurement or "press to read" mode available with averaging function for LDO measurement. Display Type : 240 x 160 pixel Display readings from one or two probes LDO : dissolved oxygen, pressure, temperature, ORP/Redox: mV, temperature, Sodium: Sodium, mV, temperature DO Measurement Range : 0.1 - 20 mg/L Calibration Curves Display : Calibration summary data logged and displayed Calibration Intervals/Alerts/Reminder: 2 hours to 7 days Probe-Specifications Measurement Range : 0 to 20.00 mg/L (or ppm) Measurement Range : 0 to 200 % air saturation Operating temperature range : 0 to 50 °C (32 - 122 °F) Pressure Range : max. 3.5 bar (50 psi) Range : 0 to 20.00 ppm Repeatability : ± 0.1 ppm (mg/L) Resolution : 0.01 ppm DO Response time : at 20 ºC: To 60 s Response Time - 20 °C : To 40 s
Bacteriological Incubator Specification • capacity approximately 95 liters • Size-24”X24”X24”inch • Double walled inner chamber available in stainless steel with grade304 • Outer chamber made of CRCA Epoxy powder coat finish • Glass wool/ PUF insulation to prevent temperature loss • Full length inner plexi-glass door for viewing sample • Outer double wall door is provided with magnetic gas kit and lock & key arrangement. • Stainless steel removable & adjustable shelves Digital display with mains ON/OFF, switch indicating lamps & Fuse • Digital microprocessor based PID Temp controller with LED Display • accuracy of temperature +/- 0.5 degree • Temperature range adjustable from Room Temp to 900C • Air flow systems providing fans for even distribution of temperature. • Safety thermostat to prevent over heat. • Electric Supply:220/230V AC,50/60 Hz,300watt
Hot Plate Specifications:- • 8 Inch round • Heat Resistant • Include Cord & Plug • ISO/ISI Certified Company
R-INITIATIVE ENTERPRISES
RAJ KUMAR CONT
stage.html
html • 0.05 MB
tech_bid_open.pdf
tech_eval.pdf
fin_bid_open.pdf
boq_comp_chart.xlsx
xlsx
fin_eval.pdf
aoc.pdf
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.03 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .