RequiredOEM Authorization Certificate Bid to RA enabled No SPECTROPHOTOMETER ( 1 pieces ) Technical Specifications * As per GeM Category Specification Specification Specification Name Values Bid Requirement (Allowed Values) TYPE Type of Spectrophotometer UV - VISIBLE UV - VISIBLE Type of Optical System Double Beam Double Beam Monochromator Type NA Reflexive Grating
RequiredHolographic Grating
RequiredCzerny-turner
RequiredNA
RequiredScanning grating Detector Type Photodiode Photo Multiplier Tube 1/6 (PMT)
RequiredPhotodiode Spectral Bandwidth Type Variable Bandwidth Variable Bandwidth LIGHT SOURCE Light Source Xenon Flash Halogen Lamp
RequiredDeuterium Lamp
RequiredLED
RequiredTungsten Halogen Lamp
RequiredXenon Flash Diffraction Grating (lines/mm) 800 600
Required800
Required1000
Required1200
Required1400
Required1600
Required1800 Switching of Light Source Automatic Automatic Wavelength Setting And Scanning Automatic Automatic OPTICAL PERFORMANCE Source Wavelength - Minimum 190 190.0 Source Wavelength - Maximum 900 900
Required960
Required1000
Required1100
Required1500
Required2000
Required2500
Required2700
Required3000
Required3200
Required3300
Required1650 Or higher Resolution 0.5 0.01
Required0.5 Wavelength Accuracy (+/-) 0.3 0.3 Bandwidth (nm) 0.5
Required1
Required2.0
Required4.0 0.5
Required1
Required1.8
Required2.0
Required0.1 to 5.0 If variable band width
RequiredSpecify the incremental step value(mm) 1.0 * Scanning Speed (nm/min) 3800 * Focal length (mm) 50 to 99 0 to 49
Required50 to 99
Required100 to 149
Required150 to 199
Required200 to 249
Required250 to 299
Required300 and above Photometric measurement modes Absorbance
RequiredTransmitta nce
RequiredConcentration (Wavelength
Requiredtime)
RequiredReflectance Transmittance
RequiredAbsorbance
RequiredConcentration (Wavelength
Requiredtime)
RequiredReflectance
RequiredEnergy
RequiredPolarisation Photometric Absorbance (Max) (Abs) (+/-) 4 0.2 - 8 Photometric Absorbance Accuracy (Max) (+/-) 0.018 0.001 - 0.5 Photometric Transmittance (Max) (+/-) 100 0 - 100 2/6 Photometric Transmittance Accuracy (Max) (+/-) 0.5 0 - 0.5 Photometric Reflectance (Max) (+/-) 100 0 - 100 Photometric Reflectance Accuracy (Max) (+/-) 0.5 * Stray Light Correction Automatic * Baseline Correction Automatic * GENERIC Display Through PC Through PC Display type NA LCD with Backlight
RequiredLED
RequiredLCD touchscreen
RequiredTFT
RequiredNA Display Size 21 21.0 - 30.0 Or higher Test result printing Through external printer Through inbuilt Printer
RequiredThrough external printer If inbuilt printer
RequiredType of printer Inkjet b/w Thermal printer
RequiredDot matrix
RequiredLaser b/w
RequiredInkjet b/w
RequiredColor laser
RequiredColor inkjet
RequiredN/A Suitable software for connecting PC & Printer Yes * Connectivity RS 232 * Internal storage(GB) 32 0 - 80 Sample switching/selection Automatic * Capacity of Sample holder/Sample Carousel 2 * Mximum Path Length of sample 100 0 - 120 Number of Cuvettes supplied-inclusive in the scope of supply 10 * Power Supply 230 V ± 10
Required50 Hz * Any other features/accesories covered in scope of supply - * Spares and Consumables inclusive in the scope of supply Yes Yes
RequiredNo 3/6 List of Items and Quantity of each item included in the offer - * OPERATING CONDITIONS Minimum Operating Temperature -5 -5
Required-10
Required0
Required5
Required10 Maximum Operating Temperature 40 35
Required40
Required45
Required50
Required55
Required60 Operating Humidity (RH) (%) at 40 degree C 90 80
Required85
Required90
Required95 WARRANTY Onsite Warranty 3 1
Required2
Required3
Required4
Required5 Availability of Spares (at extra cost) 10 2
Required3
Required4
Required5
Required6
Required7
Required8
Required10 Certification Availability of UL/CE Certification as per EN 61010-1:2010 Safety requirements for electrical equipment for measurement
Requiredcontrol
Requiredand laboratory use General requirements Yes Yes
RequiredNo Conformance to EMC/EMI as per EN 61326-1:2013 Electrical equipment for measurement
Requiredcontrol and laboratory use Yes Yes
RequiredNo Availability of Test Reports from Central Govt/NABL approved/ILAC acredited lab to prove conformity to the specification Yes Yes
RequiredNo If yes test report to furnished
Requiredto the buyer on demand Yes * Test Report number and date - * Name and address of the Lab - * * Bold specifications are the golden parameters. * Bidders may note that In respect of non-golden Parameters
Requiredthe specifications 'Values' chosen by Buyer will generally be preferred over 'Bid requirement ( allowed Values) by the Buyer.