Executive Engineer MHD SKIMS Soura
SKIMS SOURA
Replacement of old wornout Oxygen and Carbon dioxide Control Panels with automatic Control Panel along with allied works in Manifold at SKIMS Soura.
2026_PWDJK_321501_2
MHD/SKIMS/JVC/36 of 2026-27 Dated. 19.08.2026
Open Tender
Machineries/ Mechanical Engg Items
Item Rate
20 days
SKIMS SOURA
Please refer Tender documents.
8 documents required · 8 mandatory
₹0
Exempted
19 Aug 2026
19 Aug 2026
29 Aug 2026
19 Aug 2026
29 Aug 2026
19 Aug 2026
Select, Excess (+), Less (-) Less (-)
Tender Inviting Authority: Executive Engineer Mechanical & Hospital Division, SKIMS/JVC, Soura, Srinagar.
Name of Work:Replacement of old obsolete Oxygen and Carbon dioxide Control Panels with automatic Control Panel along with allied works in Manifold at SKIMS Soura. (W=02)
Contract No: e NIT NO : MHD/SKIMS/JVC 35 OF 2026-27 dated: 19 /08/2026
Sl. No. Item Code / Make Quantity Units Estimated Rate Addition / Deduction Addition / Deduction Values Currency Convertion against each Item Quoted Currency in INR / Other Currency BASIC RATE with taxes In Figures To be entered by the Bidder Rs. P Excise Duty VAT Freight Charges ( Unloading & Stacking) Any Other Taxes/Duties/Levies Other Taxes 2 IIIrd Party i.e DGS&D / RITES etc Inspection Charges @0.34%+Service Tax Less for Cenvat Credit,if any respect of Supplies Under full Excise Duty Category
1 Replacement of old obsolete nitrous oxide battery bank by new oxygen bank. The job includes dismantling the old system and then Providing ,installation, testing and commissioning of 6x6 new oxygen battery bank complete with lead (High pressure pipes), interconnecting copper pipes, 12 No. copper tail pipes with NRV and fittings for cylinders, mounting angle iron 50x50x5 mm to be fixed on the wall with clamps/dash fasteners to hold the 6x6 cylinder system in place, isolation control valve on the supply line. The job includes providing/fitting new fully automatic analog oxygen panel. The control panel should be capable of regulating minimum 1500 LPM flow for twin bank (06+06) with auto regulation and shall be designed with standing the oxygen cylinder pressure in the range of 0-200 kgf/cm2 with safety features related to hazard and continuity of flow. The panel shall operate at a constant pressure upto 6 bar through two banks of bottled gas cylinders with following status and indication system. Rest of the specifications as per Annexure B of the e-NIT. 1 Job 0.0000 Excess(+) Full Conversion INR
2 Replacement of old obsolete carbon dioxide panel by new carbon dioxide panel. The job includes first dismantling the old panel and then Providing, installation, testing and commissioning of new fully automatic analog Carbon dioxide panel. The control panel should be capable of regulating minimum 1500 LPM flow for twin bank (06+06) with auto regulation and shall be designed with standing the cylinder pressure in the range of 0-200 kgf/cm2 with safety features related to hazard and continuity of flow. The panel shall operate at a constant pressure upto 7 bar through two banks of bottled gas cylinders with following status and indication system. Rest of the specifications as per Annexure B of the e-NIT. 1 Job 0.0000 Excess(+) Full Conversion INR
3 Replacement of old obsolete oxygen panel by new oxygen panel. The job includes first dismantling the old panel and then Providing,installation, testing and commissioning of new fully automatic analog oxygen panel. The control panel should be capable of regulating minimum 1500 LPM oxygen flow for twin bank with auto regulation and shall be designed with standing the oxygen cylinder pressure in the range of 0-200 kgf/cm2 with safety features related to hazard and continuity of flow of medical grade oxygen. The panel shall operate at a constant pressure upto 6 bar through two banks of bottled gas cylinders with following status and indication system.Rest of the specifications as per Annexure B of the e-NIT. 1 Job 0.0000 Excess(+) Full Conversion INR
4 Providing and fixing of 80 mm forged ball valve with following specifications: Rest of the specifications as per Annexure B of the e-NIT. 4 No 0.0000 Excess(+) Full Conversion INR
5 Providing and fitting of 80 mm NB C class MS piping confirming to IS code 1239 connecting from refrigerant air dryer to air compressors, including all standard procedures for laying jointing cutting, welding, polishing and providing of all necessary fittings , bends, tees etc. The job includes color coating of approved shade. 30 RMT 0.0000 Excess(+) Full Conversion INR
Total in Figures
Quoted Rate in Figures 150mm dia. pipe (with 32 mm insulation), effective dia. 214 mm Select
Please Enable Macros to View BoQ information
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.03 MB
Tendernotice_1.pdf
PDF • 1.05 MB
BOQ_1074244.xls
BOQ • 0.24 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at [email protected].