Loading…
Loading…
| # | Company | Amount | Rank | Status |
|---|---|---|---|---|
| 1 | L1₹15.75Admitted-Finance PLOT NO 63 UDYOG VIHAR PH IV GURUGRAM 122015 GURGAON HARYANA 122001 | GURUGRAM | HARYANA | 122015 | ₹15.75 | L1 | Admitted-Finance |
| 2 | L2₹20.71+₹4.96 (31.5%)Admitted-Finance | ₹20.71+₹4.96 (31.5%) | L2 | Admitted-Finance |
| 3 | L3₹24.26+₹8.51 (54.0%)Admitted-Finance MUMBAI | ₹24.26+₹8.51 (54.0%) | L3 | Admitted-Finance |
| 4 | L4₹38.85+₹23.10 (146.7%)Admitted-Finance | ₹38.85+₹23.10 (146.7%) | L4 | Admitted-Finance |
| Sl No | Description | Qty | Unit | Meril Diagnostics Pvt Ltd L3 | SD Biosensor Healthcare Pvt. Ltd L1 | Alere Medical Pvt. Ltd. L4 | Biolab Diagnostics (I) Pvt. Ltd. L2 |
|---|---|---|---|---|---|---|---|
| 1.00 | HBsAg (Rapid Test)
1. Should be coated with monoclonal antibodies covering all subtypes and variants of HBsAg
2. The assay should be able to detect surface antigen to Hepatitis B virus.
3. Should be compatible with plasma and serum both.
4. Adequate documents detailing the principle, components, biosafety, methodologies, validity criteria, interpretation of results, performance characteristics, storage conditions, limitation of assays, manufacturing and expiry dates should be provided with each kit.
5. The kit should have approval of the statutory authority from the country of origin
6. In case of imported kits it should be registered and licensed by the DCG(I)
7. In case of indigenous manufactures should be licensed by the competent authourity/licensing authority defined under Drugs and Cosmetics Act (1940) and medical device rule 2017
8. The kit should have minimum shelf life of 60% or 12 months (whichever is more) at the port/place of discharge of consignees.
9. The total procedure time shall not be more thean 30 minutes.
10. The assay components should include positive and negative controls sufficient for conducting 20% of the tests (10% negative and 10% positive controls) which may be provided along with the kits if not a part of the kit.
11. The assay should have sensitivity of 100% and specificity of more thean or equal to 98% as per the office order of MoHFW vide F. No 29/Misc./4/2016-DC(65) dated 13/6/2017
12. The control dot/band should be able to detect the presence of human immunoglobulin and should not be just a “procedural control”or meant merely for checking the flow of reagents or integrity of antigens except except lateral flow technology
General specifications
1. The manufacturer/ authorized agent should ensure maintenance of cold chain during storage and transport of kits at 2-8 C. The cumulative time temperature indicator technology should be used on each kit and be per-qualified by WHO.
1. The pack size should not be more than 50 tests wherein each test is individually packed .
2. 8 Kits should be supplied along with the procurement lot of which four kits will be used for validation, subject to which the kits of each batch and lot no. will be supplied to program and four kits will be retained for evaluation close to the expiry of the kit or in case of legal dispute with regards to performance of the kit as per specified parameters
3. The kit will bw evaluated on the above parameters by the centers approved by the program
The committee approved the specification of HBsAg(rapid test)
1.The control dot/band should be able to detect the presence of human immunoglobulin and should not be just a “procedural control” or meant merely for checking the flow of reagents or integrity of antigens
General specifications
1. The manufacturer/authorized agent should ensure maintenance of could chain during storage and transport of kits at 2-8° C. The cumulative time temperature indicator technology should be used on each kit and be pre-qualified by WHO.
2. The pack size should not be more than 50 tests wherein each test is individually packed.
3. 8 kits should be supplied along with the procurement lot of which four kits will be used for validation, subject to which the kits of each batch and lot no. will be supplied to program and four kits will be retained for evaluation close to the expiry of the kit or in case of legal dispute with regards to performance of the kit as per specified parameters
4. The kit will be evaluated on above perameters by the centers approved by the program
The committee approved the specifications for Hepatitis B surface Antigen (Rapid Test) | 17,49,440 | Units | 8.3 ₹8.71 Lowest | - | 22 ₹23.1 | 8.78 ₹9.22 |
| 2.00 | Anti-HCV Antibody (Rapid Test)
1. Should utilize recombinant and /or synthetic peptide antigens for core, NS3, NS4 and NS5.
2. The assay should detect total anti HCV antibodies
3. Should be compatible with plasma and serum both.
4. Adequate documents detaling the principle, components, biosafety, methodologies, validity criteria, interpretation of results, performance characteristics, storage conditions, limitation of assays, manufacturing and expiry dates should be provided with each kit.
5. The kit should have approval of the statutory authority from the country of origin
6. In case of imported kits it should be registered and licensed by the DCG (I)
7. In case of indigenous manufactures should be licensed by the competent authority/Licensing authority defined under Drugs and Cosmetics Act (1940) and medical device rule 2017
8. The kit should have minimum shelf life of 60% or 12 months (whichever is more) at the port/place of discharge of consignees.
9. The total procedure time shall not be more than 30 minutes.
10. The assay component should include positive and negative controls sufficient for conducting 20% of the tests (10% negative and 10% positive controls) which may be provided along with the kits if not a part of the kit.
11. The assay should have sensitivity more than or equal to 99% and specificity of more than or equal to 98% as per the office order of MoHFW vide F. No 29/Misc./4/2016-DC (65) Dated 12/7/2017
12. The control dot/band should be able to detect the presence of human immunoglobulin and should not be just a “ procedural control” or meant merely for checking the flow of reagents or integrity of antigens except in lateral flow technology.
General Specifications
1. The manufacturer/ authorized agent should ensure maintenance of cold chain during storage and transport of kits at 2-8 C. The cumulative time temperature indicator technology should be used on each kit and be per-qualified by WHO.
2. The pack size should not be more than 50 tests wherein each test is individually packed.
3. 8 kits should be supplied along with the procurement lot of which four kits will be used for validation, subject to which the kits of each batch and lot no. will be supplied to program and four kits will be retained for evaluation close to the expiry of the kit or in case of legal dispute with regards to performance of the kit as per specified parameters
4. The kit will be evaluated on the above parameters by the centers approved by the program
The committee approved the specification of Anti HCV antibody (rapid test)
11. The assay should have sensitivity more than or equal to 99% and specificity of more than or equal to 98% as claimed by the manufacturer in the kit literature
12. The control dot/band should be able to detect the presence of human immunoglobulin and should not be just a “procedural contol” or meant merely for checking the flow of reagents or integrity of antigens
General Specifications
1. The manufacturer/ authorized agent should ensure maintenance of cold chain during storage and transport of kits at 2-8° C. The cumulative time temperature indicator technology should be used on each kit and be pre-qualified by WHO.
2. The pack size should not be more than 50 tests wherein each test is individually packed.
3. 8 kits should be supplied along with the procurement lot of which four kits will be used for validation, subject to which the kits of each batch and lot no. will be supplied to program and four kits will be retained for evaluation close to the expiry of the kit or in case of legal dispute with regards to performance of the kit as per specified parameters
4. The kit will be evaluated on the above parameters by the centers approved by the program
The committee approved the specifications for Anti-HCV Antibody Kits (Rapid test) | 5,83,147 | Units | 14.8 ₹15.54 | 15 ₹15.75 | 15 ₹15.75 | 10.94 ₹11.49 Lowest |
Tender Value
₹5 Cr
EMD Value
₹5 L
Closing Date
16 Dec 2021, 6:00 pmClosed
MD RMSCL
RMSC, Swasthya Bhawan, Tilak Marg C-Scheme, Jaipur
SURGICALS SUTURES (DRUG ITEMS)
2021_RMSCL_247082_1
F.02(110)RMSCL Proc. SS(MD) NIB-22 2021 660 Date
Open Tender
Supply of Materials/Hiring of Goods
Item Wise
60 days
All DDW
Please refer Tender documents.
4 documents required · 4 mandatory
₹2,000
MD, RMSCL
₹5 L
Yes
Conference hall, RMSC
16 Feb 2022
12 Nov 2021
17 Dec 2021
12 Nov 2021
16 Dec 2021
20 Nov 2021
16 Nov 2021
HBsAg (Rapid Test) 1. Should be coated with monoclonal antibodies covering all subtypes and variants of HBsAg 2. The assay should be able to detect surface antigen to Hepatitis B virus. 3. Should be compatible with plasma and serum both. 4. Adequate documents detailing the principle, components, biosafety, methodologies, validity criteria, interpretation of results, performance characteristics, storage conditions, limitation of assays, manufacturing and expiry dates should be provided with each kit. 5. The kit should have approval of the statutory authority from the country of origin 6. In case of imported kits it should be registered and licensed by the DCG(I) 7. In case of indigenous manufactures should be licensed by the competent authourity/licensing authority defined under Drugs and Cosmetics Act (1940) and medical device rule 2017 8. The kit should have minimum shelf life of 60% or 12 months (whichever is more) at the port/place of discharge of consignees. 9. The total procedure time shall not be more thean 30 minutes. 10. The assay components should include positive and negative controls sufficient for conducting 20% of the tests (10% negative and 10% positive controls) which may be provided along with the kits if not a part of the kit. 11. The assay should have sensitivity of 100% and specificity of more thean or equal to 98% as per the office order of MoHFW vide F. No 29/Misc./4/2016-DC(65) dated 13/6/2017 12. The control dot/band should be able to detect the presence of human immunoglobulin and should not be just a “procedural control”or meant merely for checking the flow of reagents or integrity of antigens except except lateral flow technology General specifications 1. The manufacturer/ authorized agent should ensure maintenance of cold chain during storage and transport of kits at 2-8 C. The cumulative time temperature indicator technology should be used on each kit and be per-qualified by WHO. 1. The pack size should not be more than 50 tests wherein each test is individually packed . 2. 8 Kits should be supplied along with the procurement lot of which four kits will be used for validation, subject to which the kits of each batch and lot no. will be supplied to program and four kits will be retained for evaluation close to the expiry of the kit or in case of legal dispute with regards to performance of the kit as per specified parameters 3. The kit will bw evaluated on the above parameters by the centers approved by the program The committee approved the specification of HBsAg(rapid test) 1.The control dot/band should be able to detect the presence of human immunoglobulin and should not be just a “procedural control” or meant merely for checking the flow of reagents or integrity of antigens General specifications 1. The manufacturer/authorized agent should ensure maintenance of could chain during storage and transport of kits at 2-8° C. The cumulative time temperature indicator technology should be used on each kit and be pre-qualified by WHO. 2. The pack size should not be more than 50 tests wherein each test is individually packed. 3. 8 kits should be supplied along with the procurement lot of which four kits will be used for validation, subject to which the kits of each batch and lot no. will be supplied to program and four kits will be retained for evaluation close to the expiry of the kit or in case of legal dispute with regards to performance of the kit as per specified parameters 4. The kit will be evaluated on above perameters by the centers approved by the program The committee approved the specifications for Hepatitis B surface Antigen (Rapid Test)
Anti-HCV Antibody (Rapid Test) 1. Should utilize recombinant and /or synthetic peptide antigens for core, NS3, NS4 and NS5. 2. The assay should detect total anti HCV antibodies 3. Should be compatible with plasma and serum both. 4. Adequate documents detaling the principle, components, biosafety, methodologies, validity criteria, interpretation of results, performance characteristics, storage conditions, limitation of assays, manufacturing and expiry dates should be provided with each kit. 5. The kit should have approval of the statutory authority from the country of origin 6. In case of imported kits it should be registered and licensed by the DCG (I) 7. In case of indigenous manufactures should be licensed by the competent authority/Licensing authority defined under Drugs and Cosmetics Act (1940) and medical device rule 2017 8. The kit should have minimum shelf life of 60% or 12 months (whichever is more) at the port/place of discharge of consignees. 9. The total procedure time shall not be more than 30 minutes. 10. The assay component should include positive and negative controls sufficient for conducting 20% of the tests (10% negative and 10% positive controls) which may be provided along with the kits if not a part of the kit. 11. The assay should have sensitivity more than or equal to 99% and specificity of more than or equal to 98% as per the office order of MoHFW vide F. No 29/Misc./4/2016-DC (65) Dated 12/7/2017 12. The control dot/band should be able to detect the presence of human immunoglobulin and should not be just a “ procedural control” or meant merely for checking the flow of reagents or integrity of antigens except in lateral flow technology. General Specifications 1. The manufacturer/ authorized agent should ensure maintenance of cold chain during storage and transport of kits at 2-8 C. The cumulative time temperature indicator technology should be used on each kit and be per-qualified by WHO. 2. The pack size should not be more than 50 tests wherein each test is individually packed. 3. 8 kits should be supplied along with the procurement lot of which four kits will be used for validation, subject to which the kits of each batch and lot no. will be supplied to program and four kits will be retained for evaluation close to the expiry of the kit or in case of legal dispute with regards to performance of the kit as per specified parameters 4. The kit will be evaluated on the above parameters by the centers approved by the program The committee approved the specification of Anti HCV antibody (rapid test) 11. The assay should have sensitivity more than or equal to 99% and specificity of more than or equal to 98% as claimed by the manufacturer in the kit literature 12. The control dot/band should be able to detect the presence of human immunoglobulin and should not be just a “procedural contol” or meant merely for checking the flow of reagents or integrity of antigens General Specifications 1. The manufacturer/ authorized agent should ensure maintenance of cold chain during storage and transport of kits at 2-8° C. The cumulative time temperature indicator technology should be used on each kit and be pre-qualified by WHO. 2. The pack size should not be more than 50 tests wherein each test is individually packed. 3. 8 kits should be supplied along with the procurement lot of which four kits will be used for validation, subject to which the kits of each batch and lot no. will be supplied to program and four kits will be retained for evaluation close to the expiry of the kit or in case of legal dispute with regards to performance of the kit as per specified parameters 4. The kit will be evaluated on the above parameters by the centers approved by the program The committee approved the specifications for Anti-HCV Antibody Kits (Rapid test)
Biolab Diagnostics (I) Pvt. Ltd.
Meril Diagnostics Pvt Ltd
Alere Medical Pvt. Ltd.
stage.html
html • 0.03 MB
tech_bid_open.pdf
tech_eval.pdf
boq_comp_chart.xlsx
xlsx
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.03 MB
Tendernotice_1.pdf
PDF • 0.06 MB
DurgDoc.pdf
Tender Documents • 2.54 MB
BOQ_441068.xls
BOQ • 0.29 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .