Executive Engineer M and H, Dn. I Jaipur
Executive Engineer M and H, Dn. I Jaipur
SITC of Oxygen Generator Plant at Rajeev Gandhi Government Hospital, Alwar.
2020_MEDIC_198692_2
NIT 37/2020-21 EE M and H Dn. I Jaipur/
Open Tender
Electrical Works
Percentage
60 days
Alwar
Please refer Tender documents.
3 documents required · 3 mandatory
₹1,000
RTGS/NEFT/Any other money Transfer System
₹67,000
23 Sept 2020
9 Sept 2020
18 Sept 2020
9 Sept 2020
17 Sept 2020
9 Sept 2020
Amount
PART-A (NON-BSR ITEM)
SITC of Pressure Swing Absorption Oxygen Generator Plant capacity should be minimum 30M 3 per hour and system should able to produce gas equivalent to 90 (Approx. )Jumbo Cylinders per day, having single compressor and dryer.
Purity of Oxygen 93%± 3% @4.1 bar pressure
Rotary Screw type air compressor of suaitable capacity ( 1 Nos) as per afore mention capacity 400 V /3 Phase /50 HZ ,Outlet pressure 4.1 BAR,Maximum Output Pressure 7.5 BAR, Power Supply 110-240 V, 1 phase , 50 Hz , UPS Module for control only with medical upgrade kit , Coal Tower , carbon filter , Bacteria/ Sterile .
The unit should comprise four stage air filtration , consists of moisture seperator , 0.1 micron pre filter before the Refrigerated air dryer , another two filters 0.1 and 1 micron after the dryer then a carbon tower which makes sure 0 oil and vapour in the down stream air. The carbon should last for more than two years . and it should be provided 1 micron filter after carbon tower. Beneath this filter a moisture alarm should be incorporated to switch off the PSA system in case of dryer auto drain failure .
This unit should also comprise of 1nos air receiver 1000 Litres , oxygen /Buffer tanks 1000 Ltrs as per standards, an in-line flow meter , certifying the capacity shall be installed at the outlet of the oxygen receiver
In case PSA Oxygen Generator failure , an auto switch over system to regular supply for oxygen is incorporated so that automatically back up from cylinders shall be taken over. To show this effect an alarm system should be installed with this system.
All the system piping and fittings , all the interconnections , and piping up to manifold room shall be of stainless steel .
This system should be comprising of OFF -Spec Valve to prevent the supply of impure oxygen to go into the system if the purity goes to below the 90% .
The Unit can be monitor on line through by service team /authorised person 24/7 on no extra cost
The Connected electricity load should not be more than 35 KW
The Flow Meter certifying the capacity should be also there
The Voltage Stabilizer of suitable capacity and standerd make should be provided. Make of servomate/servokone or equivalent make.
Above Oxygen Generator should be ISO:13485/CE/US FDA certified. In support of this certification , Notified third party inspection certificate with test results is mandatory.
UNISSI INDIA PVT LTD
Tridentpneumatics.com
Agastya Aeroworks Pvt Ltd
Tridentpneumatics.com
stage.html
html • 0.03 MB
tech_bid_open.pdf
tech_eval.pdf
fin_bid_open.pdf
boq_comp_chart.xlsx
xlsx
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.04 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .