| Sl No | Description | Qty | Unit | Est. Rate | Est. Amount |
|---|---|---|---|---|---|
| 1SPECTROFLUOROMETER | |||||
| 1.01 | • Optical System and Monochromator: Optical Design: Single/Double Monochromator using Holographic concave grating - Excitation & Emission Range: Zero Order, 200 nm to 750 nm (or) better - Wavelength Accuracy: Should be at least +/-1.5 nm (or) Better - Wavelength Repeatability: +/- 1.0 nm (or) Better - Spectral Bandwidth (Slits): Variable and minimum should be 1, 2.5, 5,10,20, L5, L10nm on Both excitation and emission monochromators - High-Order Diffraction Removal: Automatic software-controlled higher-order diffraction cut filters must be provided as standard on both channels.
• Light Source and Detection Light Source: High-stability, long-life 150 W Ozone-Free Xenon Arc Lamp with shielded lamp housing (rated lifetime ≥ 3,000 hours).- Validation Standard: Built-in / integrated Mercury (Hg) lamp for automatic wavelength calibration and instrument validation.- Photofading Protection: Automatic software-controlled shutter system to protect sensitive biological or chemical samples from photobleaching during idle periods. – Detectors; Excitation Channel: Silicon Photodiode for real-time reference monitoring. Emission Channel: High-performance, high-sensitivity Photomultiplier Tube (PMT). Sensitivity (Signal-to-Noise): Water Raman Peak S/N ratio ≥ 8,000 : 1(RMS)
• Performance & Measurement Capabilities -Wavelength Scan Speed: 10 to 20,000 nm/min (or better). - Slew Speed: 30,000 nm/min or better. - Photometric Range: -10,000 to +10,000 or wide dynamic range (≥6 orders of magnitude). Software Package: Windows 10/11-compatible control & processing software supporting: 2D & 3D Excitation-Emission Matrix (EEM) scanning, Fixed wavelength & multi-wavelength kinetics / time-course decay, Synchronous scan mode (constant wavelength and constant energy), Relative Quantum Yield calculation, Absorbance measurement & automatic spectral correction, Instrument self-diagnostics and auto-accessory recognition
• Sample Compartment & Accessories - Liquid Sample Handling: Standard 10 mm rectangular cell holder assembly. Must include 1 pair (2 Nos.) of Quartz Fluorescence Cuvettes (10 mm path length, 3.5 mL volume, 4-sided polished windows) with PTFE stoppers/lids. Solid/Powder Sample Handling: Dedicated Solid/Powder Sample Holder for thin films, powders, and opaque substrates. Temperature Control: Provision/capability to integrate a Peltier-based thermostatted cell holder system. Expansion Options: System must support future upgradeability to Micro-volume cell holders (up to 5µL).
• Warranty & Delivery Terms: Warranty: Minimum 3-Year comprehensive warranty on the complete system including light source and accessories from the date of successful installation. Installation & Training: Free on-site installation, demonstration, and user training by factory-trained service engineers. | 1 | No | 100 | ₹100 |
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.03 MB
Tendernotice_1.pdf
PDF • 10.79 MB
BOQ_836200.xls
BOQ • 0.32 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at [email protected].