| 1.01 | Item: SITC of EIP Chlorine Analyser
Measurement Range : 0-20 mg/L (or ppm)
Measurement Technique : Membrane-covered Three electrode amperometric Technology
Accuracy Technology : ± 0.1 mg/L
Repeatability : ± 0.05m/L(25°C)
Re-Productivity : ± 4%
Resolution : 0.01
Response Time : 90 Second to 90%(T-90), 25°C
Operating Temperature : 0 to 50°C
Operating Pressure : Maximum 15 PSI
Sample pH : 4-12 greatly reduced pH interference
Flow Requirement Minimum: 30L/hr (0.6cm/s)
Flow Requirement Maximum: 60L./hr (1.3cm/s)
Material: Body Housing: PVC, ABS, Membrane: Hydrophilic PTFE, Electrod: Silver-Silver Halide/Gold/316SS.
Temperature Compensation :- In-built, Automatic | 1 | Set | 10,75,000 ₹12,68,500 | 10,79,000 ₹12,73,220 |
Tender Value
Refer Docs
EMD Value
₹25,500
Closing Date
6 Mar 2026, 3:00 pmClosed
Executive Engineer (EM) M-1
Chitra Vihar BP Station Laxmi Nagar Delhi-110092
SITC of CMS system at 23 Block Trilokpuri BPS.
2026_DJB_287951_5
NIT No-19 EE (EM) M-1/ (25-26)
Open Tender
Electrical Works
30 days
Chitra Vihar BP Station Laxmi Nagar Delhi-110092
Please refer Tender documents.
3 documents required · 3 mandatory
₹500
A/c.50448339804
₹25,500
31 Mar 2026
28 Feb 2026
6 Mar 2026
28 Feb 2026
6 Mar 2026
28 Feb 2026
Item: SITC of EIP Chlorine Analyser Measurement Range : 0-20 mg/L (or ppm) Measurement Technique : Membrane-covered Three electrode amperometric Technology Accuracy Technology : ± 0.1 mg/L Repeatability : ± 0.05m/L(25°C) Re-Productivity : ± 4% Resolution : 0.01 Response Time : 90 Second to 90%(T-90), 25°C Operating Temperature : 0 to 50°C Operating Pressure : Maximum 15 PSI Sample pH : 4-12 greatly reduced pH interference Flow Requirement Minimum: 30L/hr (0.6cm/s) Flow Requirement Maximum: 60L./hr (1.3cm/s) Material: Body Housing: PVC, ABS, Membrane: Hydrophilic PTFE, Electrod: Silver-Silver Halide/Gold/316SS. Temperature Compensation :- In-built, Automatic