blood pressure recording units Blood Pressure Recording Units ( 35 ) Technical Specifications * As per GeM Category Specification Specification Specification Name Values Allowed Values CONFIGURATION DETAILS Type Of SPHYGMOM ANOMETERS or Blood Pressure recording unit: Digital Mercurial
Aneroid
Digital Conformity to Indian Standrad for SPHYG MOMANOMETERS NA for digital IS:3390 latest for Mercurial
IS 7652 latest for Aneroid
NA for digital Measuring device Automated Mechnical
Automated Type Portable Portable
Mountable/Fixed
Inclined
Integrated ISI MARKED NA for digital Yes
NA for digital CM/L No and validity date NA * Measurement Method Oscillometric Oscillometric
Auscultatory method
others Display for Digital BP instrument: LCD(Liquid Crystal Digital) Display LCD(Liquid Crystal Digital) Display
NA for 1/4 others Range: Pressure measurement (mmHg) 0 to 300 * Range:Pulse measurement for digital with accuracy ± 5%. (per minute) 40 to 200 * Accuracy:Pressure measurement (mmHg) +/- 3 * Pressure Detection Capacitive Pressure sensor * Inflation Automatic Pressure Application by pump Automatic Pressure Application by pump
Manual Deflation and rapid Air Release Automatic Pressure Release Valve Automatic Pressure Release Valve
Manual Cuff Range Medium (22-32)cm Small (18-22)cm
Medium (22-32)cm
Large (32-45)cm Dimensions of the Equipment (mmXmm) 300X300 * Dimensions of the Display screen (mmXmm) 20X30 * Operating Temperature range 0 to 55 degree Celsius * Stroage Temperature range (-10) to 55 degree Celsius * Automatic power off if system is idle for 3 minutes. Yes * Weight (in gram) 450 * Power Power Source: AA size chargeable battery set AA size chargeable battery set
Either Charageble Battery Set Or {220 Volt
50 Hz AC Supply ] Type of AA Size chargeable Battery (AA Size) Alkaline Alkaline
Lithium
Ni- cd WARRANTY
CERTIFICATIONS
TEST REPORTS Availability of Type Test Report covering complete test parameters as per IS3390 latest for Mercurial and IS7652 latest for Aneriod type
from any ILAC/NABL Yes * 2/4 accredited/Central Gov LABFor digital type including Environmental test from any ILAC/NABL accredited/Central Gov LAB Type test report No and date BP-(275)01 28/07/16 * Name and address of the Lab where test Conducted. NA * CE Marking Certificate for B.P. Apparatus Model. Yes * Conformity to OILM- R-16-1 and OILM- R-16-2 for both Non- invasive mechanical sphygmomanometers and Non-invasive automated sphygmomanometers are for guidance and All the Gazette of india related to legal metrology issued time to time
are applicable to the specification as well as to the product. Yes * Others Certificate ie ROHS
WHO
GMP etc. NA * For Digital Sphygmomanometer
Availability of Type test reports consisting of verification of all the features & functional parameters & environmental tests sequences as under: a) Dry heat test (For 16 hrs at a temp of 55degree C in accordance with IS: 90 00/part-3/section-3/197 7)latest Yes * For Digital Sphygmomanometer Cold test(For 4 hrs at a temp of (-10) degree C in accordance with IS: 9 000/part-2/section-3/19 77latest c) Damp Heat Yes * 3/4 Cyclic Test ( For 2 cycles of 24 hrs at a temp 55 degree C & 95% RH in accordance with IS: 9000/part-5/sec tion-1/1981)latest Note:- The BP Instrument shall be checked for all the parameters before conditioning After completion of the above environmental tests sequence
with a recovery period of 1 to 2 hrs
BP Instrument shall be functional " * Bold specifications are the golden parameters. Consignees and Quantity S.No. Consignee Address Quantity Delivery Days 1 Saurav Jyoti Patgiri 786001
ICMR-RMRC
NE Region
Dibrugarh
Assam 35 15