RequiredExperience with Government
RequiredAnnual Turnover
RequiredCertificate (Requested in ATC)
RequiredOEM Authorization Certificate
RequiredOEM Annual Turnover Bid to RA enabled No Note: 1. OEM Turn Over Criteria: The minimum average annual financial turnover of the OEM of the offered product during the last three years
Requiredending on 31st March of the previous financial year
Requiredshould be as indicated in the bid document. Documentary evidence in the form of certified Audited Balance Sheets of relevant periods or a certificate from the Chartered Accountant/Cost Accountant indicating the turnover details for the relevant period shall be uploaded with the bid. In case the date of constitution/incorporation of the OEM is less than 3 year old
Requiredthe average turnover in respect of the completed financial years after the date of constitution shall be taken into account for this criteria. Cardiac Monitor With Defibrillator ( 17 pieces ) Technical Specifications 1/8 * As per GeM Category Specification Specification Specification Name Values Bid Requirement (Allowed Values) Standards Conformity to Certification Both US FDA and European CE European CE
RequiredBoth US FDA and European CE Certification number and date - * Confirmity to Manufactures Certification ISO 9001 & ISO 13485 ISO 9001 & ISO 13485 Conformity to Safety Standards IEC 60601-2-4 or Equivalent BIS * Performance Parameters Capability parameter of defibrillator ECG monitoring
Requiredexternal defibrillation
Requiredexternal pacing (transcutaneous)
Requiredinternal defibrillation and recording ECG monitoring
Requiredexternal defibrillation
Requiredexternal pacing (transcutaneous)
Requiredinternal defibrillation and recording Technology of defibrillator Biphasic technology * Modes in defibrillator Automated external defibrillation and manual Automated external defibrillation and manual Maximum energy selection in joules 200 * Capability of ECG Monitoring ECG leads
Requiredmulti function electrodes
Requiredpaddles * Number of wave- forms 3 3
Required4 Patient compatibility to defibrillate Adult
Requiredpediatric patients Adult
Requiredpediatric patients Type of display TFT TFT
RequiredLCD Size of display screen in inch 6.5 * Provision to display ECG waveform on bright high resolution display Yes * Facility to have synchronized cardio version Yes Yes Ability to display CPR index on screen real time Yes Yes Ability to filter out Yes Yes 2/8 CPR artefacts to see organised rhythms without interrupting chest compression Provision of within built rechargeable battery Yes * Battery backup to deliver number of shocks at maximum energy 100 * Weight of defibrillator with battery and paddles in Kg 6.9 * Provision of in built recorder Yes Yes Provision of printing ECG trace & stored information Yes Yes Facility of External non-invasive pacing Yes Yes
RequiredNo Type of external transcutaneous pacing modes Both Demand mode and fixed mode Demand mode
RequiredFixed mode (asynchronous)
RequiredBoth Demand mode and fixed mode Pulse width of External non- invasive pacing in milli seconds 40 40 Ability to measure chest compression rate and depth in real time with visual feedback on screen with rate and depth indicator Integrated with defibrillator Integrated with defibrillator Facility to monitor EtCO2 No No Facility to monitor NIBP No No Facility to monitor SPO2 No Yes
RequiredNo Upgradability to monitor EtCO2 Yes Yes
RequiredNo Upgradability to monitor NIBP Yes Yes
RequiredNo Upgradability to monitor SPO2 Yes Yes ECG monitoring Using 5/6 lead Using 5/6 lead 3/8 Provision of user selectable alarm settings Yes * The machine should work on mains as well as on rechargeable battery Yes * Ability to select energy from paddles Yes * Ability to charge and discharge through paddles as well as from main unit Yes * Facility for internal defibrillation No No Charging time for maximum energy in seconds 7 * Mechanism of self test of unit Automatic and manual * Unit should do self test with facility to give printout of defibrillator testing report and also have code ready indicator on unit Yes * Provision of stainless steel trolley with lockable castors Yes Yes Suitability of defibrillator for transport on ground (ambulance) No Yes
RequiredNo Defibrillator suitable to use at high altitudes and in air craft No No Defibrillator should display selected energy Yes Yes Defibrillator should display delivered energy Yes Yes Suitability of defibrillator for transportation in air (aircraft) No No Availaibility of suitable protection Yes Yes
RequiredNo 4/8 for dust and water Accessories Li-ion Battery 1 1 ECG cable 1 1 NIBP pediatric cuff with hose Not provided 1
Required2
RequiredNot provided NIBP adult cuff with hose Not provided 1
Required2
RequiredNot provided Reusable airway adaptor to be used with ETCO2 main stream/side stream sensor and cable Not provided 1
Required2
RequiredDisposable airway adpater - 2 each for intubated and non intubated patient with EtCO2 mainstream sensor and cable
RequiredNot provided Disposable airway adpater with EtCO2 mainstream sensor and cable Not provided One each for incubated and non intubated patient
Required2 each for incubated and non incubated patient
RequiredNot provided External defibrillator paddles (pediatric in built in adult) 1 1 Multi Function Defibrillator & Monitoring pads/gel sheets 12 12 Recorder paper roll 10 10 Reusable CPR feedback sensor/or similar product reused 1 1 Miscellaneous Parameters Warranty in years 5 5 Original User Technical manuals to be supplied in English in hard and soft copy Yes * Original copy of operating and service manuals to be supplied in English in hard and soft copy Yes * Calibration and preventive maintenance certificate has to be provided as per requirement Yes * Installation
RequiredDemonstration and Yes * 5/8 training to be provided at consignee end at any part of country The Principal Manufacturer Must Have Direct Presence/approved service center In India Yes * Advanced maintenance task required shall be documented and procedures required for local calibration and routine maintenence and list of imp spare parts with accessories along with part numbers to be furnished Yes * Availabilty of toll free facility for technical support maintened by OEM or authorized agencies Yes * Provide Certificate from the principal that spare parts will be available from next 10 years at the time of supplies to the buyer Yes * Copies of all certifications to be provided to buyer on demand at time of supplies Yes * * Specifications highlighted in bold are the Golden Parameters. * Bidders may note that In respect of non-golden Parameters
Requiredthe specifications 'Values' chosen by Buyer will generally be preferred over 'Bid requirement ( allowed Values) by the Buyer.