Requiredshould be as indicated in the bid document. Documentary evidence in the form of certified Audited Balance Sheets of relevant periods or a certificate from the Chartered Accountant/Cost Accountant indicating the turnover details for the relevant period shall be uploaded with the bid. In case the date of constitution/incorporation of the OEM is less than 3 year old
Requiredthe average turnover in respect of the completed financial years after the date of constitution shall be taken into account for this criteria. Semi Automatic Bio Chemistry Analyser ( 8 pieces ) Technical Specifications 1/5 * As per GeM Category Specification Specification Specification Name Values Bid Requirement (Allowed Values) Configuration & Physical Characteristics Usage: The equipment; Semi automatic bio chemistry analyser is used for measurements of bio-chemistry tests
Requiredusing wet chemistry reagent(s) and have ability to use external cuvettes and integrated flow cell. Semi automated Bio Chemistry Analyser is with built in software for the calculation and curve plotting . * User Interface: Mechanical key board Mechanical key board Flow cell type: Permanent flow cell Permanent flow cell Light Source: Halogen Lamp * Wavelength Selection & Measurement range (wavelengths): Automatic Wavelength selection facility with 6 position static photometer covers standard wavelength in the range of 340 to 630 nm Automatic Wavelength selection facility with 6 position static photometer covers standard wavelength in the range of 340 to 630 nm
RequiredAutomatic Wavelength selection facility with 8 position static photometer covers standard wavelength in the range of 340 to 700 nm 6 fixed and 2 free positions Or higher Absorbance range: 0 to 3 unit 0 to 3 unit Or higher Analyzer resolution: 0.0001 absorbance unit or better * Measurement range (temperature): 25
Required30
Required37 degree Celsius with 1 degree increment
RequiredPeltier * Memory for storing test results
Requiredsufficient for: >= 200 test results and 5000 patient results * Heat Dissipation ability: Analyser shall maintain nominal temperature
Requiredwith suitable heat dissipation/cooling mechanism. * 2/5 Installation Type: Stationary installation in laboratory * Number of programmable Channels for chemistries: >=100 to 200 >=100 to 200
Required>= 200 Or higher Assay/Calculation types available in analyser: End point with factor or standard
RequiredFixed time with factor or standard
RequiredKinetic; Enzyme kinetics with factor or standard
RequiredAbso rbance
RequiredSample blanking End point with factor or standard
RequiredFixed time with factor or standard
RequiredKinetic; Enzyme kinetics with factor or standard
RequiredAbsorbance
RequiredSample blanking
RequiredDifferential mode with factor or standard Calibration Types available: Linear factor
RequiredMulti point
RequiredPoint to point
RequiredLog- Logit
RequiredSpline
Required1- point calibration
Required2-point calibration Linear factor
RequiredMulti point
RequiredPoint to point
RequiredLog-Logit
RequiredSpline
Required1- point calibration
Required2-point calibration
RequiredPolygonal multi standard (Calibration Curve ) 3 levels control facility with day to day Levy-Jennings curve stored and displayed
Requiredavailable: Yes Yes Sample incubator facility available: Yes Yes Flow Cell Flow cell material Quartz + Metallic Quartz + Metallic Flow Cell Volume(minimum): - * Flow Cell Optical Path: 10 mm * Aspiration Volume - * ENERGY SOURCE Power Source: Compatible with electric power from by 220V-240V 50Hz (+/-10%) AC source * Power Back up with provision of UPS (with maintenance fee batteries) for each equipment unit: At least 30 minutes power back up * CONSUMABLES Equipment is suitable for use of consumables under: Open System Open System Or higher ACCESSORIES Accessories to be supplied with each unit of Analyser equipment Micro pipettes (5 No.) - 2 variable (5-50 )
Required(100-1000)
RequiredSmall * 3/5 : Tips- 500 nos. & Large Tips- 500 nos. OPERATING CONDITIONS Operating & Storage Conditions: Equipment is suitable for storage and capable of operating continuously in ambient room temperature of 8 degree C to 37 degree C and relative humidity of 15 to 90% in ideal circumstances. * Parts of the Device that are designed to come into contact with the patient or the operator should either be capable of easy disinfection or be protected by a single use/disposable cover: Yes * PRINTER AND COMPUT ER/LABORATORY INTERFACE Printer facility: Yes
RequiredIn-built Thermal Printer * Computer interface: By RS232 port/USB/HDMI port facility * Lab Interface system (LIS) compatible viz analyser equipment preloaded with Software needed for lab interface: Yes Yes Quality Certifications & Warranty Certification available for the product: CE * Equipment shall meet internationally recognised for Electromagnetic Compatibility (EMC) for electro medical equipment: 61326-1 and certified to be compliant with IEC 61010-1
RequiredIEC 61010-2-281 Yes Yes Equipment Manufacturer have ISO 13485 certification for quality standards: - Yes Warranty Period: 3 year 3 year
Required5 year Or higher 4/5 * Specifications highlighted in bold are the Golden Parameters. * Bidders may note that In respect of non-golden Parameters
Requiredthe specifications 'Values' chosen by Buyer will generally be preferred over 'Bid requirement ( allowed Values) by the Buyer.