Tender Value
₹4.8 L
EMD Value
₹9,987
Closing Date
9 Sept 2026, 11:00 amClosed
EX. ENGINEER (ELECT)/Old Delhi Zone (CSPZ)
OFFICE OF THE EX. ENGINEER (ELECT)/Old Delhi Zone (CSPZ) UNDER GROUND CAR PARKING ASAF ALI ROAD, DELHI-110002
Improvement of Electrical installation and fans in existing X Ray room, pharmacy and gallery Mrs GLM Hospital
2026_MCD_289566_1
30/MWEEE-I/2026-27/TC
Open Tender
Electrical Works
Percentage
60 days
CSPZ
refer special conditions
2 documents required · 2 mandatory
₹590
₹9,987
10 Sept 2026
2 Sept 2026
9 Sept 2026
2 Sept 2026
9 Sept 2026
2 Sept 2026
This is a percentage-rate BOQ. The schedule below shows estimated quantities and rates. Bidders quote a single percentage above or below the total estimated cost.
| Sl No | Description | Qty | Unit | Est. Rate | Est. Amount |
|---|---|---|---|---|---|
| 1 | Wiring for light point/ fan point/ exhaust fan point/ call bell point with 1.5 sq.mm FRLS/HFFR PVC insulated copper conductor single core cable in surface/ recessed medium class PVC conduit, with modular switch, modular plate, suitable GI box and earthing the point with 1.5 sq.mm FRLS/HFFR PVC insulated copper conductor single core cable etc. as required. (a) Group C | 45 | Each | 1,704 | ₹76,680 |
| 2 | Rewiring for light point/ fan point/ exhaust fan point/ call bell point with 1.5 sq.mm FRLS/HFFR PVC insulated copper conductor single core cable and 1.5 sq.mm FRLS/HFFR PVC insulated copper conductor single core cable as earth wire in existing surface/ recessed steel/PVC conduit including dismantling as required. Group C | 10 | Each | 1,064 | ₹10,640 |
| 3.1 | Wiring for circuit/ submain wiring alongwith earth wire with the following sizes of FRLS PVC insulated copper conductor, single core cable in surface/ recessed medium class PVC conduit as required. (a) 2 X 2.5 sq. mm + 1 X 2.5 sq. mm earth wire | 160 | Meter | 311 | ₹49,760 |
| 3.2 | b) 2 X 4 sq. mm + 1 X 4 sq. mm earth wire | 90 | Meter | 373 | ₹33,570 |
| 3.3 | c) 2 X 6 sq. mm + 1 X 6 sq. mm earth wire | 90 | Meter | 482 | ₹43,380 |
| 3.4 | d) 4 X 4 sq.mm + 2 X 4 sq.mm earth wire | 40 | Meter | 587 | ₹23,480 |
| 4 | Supplying and drawing following sizes of FRLS/HFFR PVC insulated copper conductor, single core cable in the existing surface/ recessed steel/ PVC conduit as required. 5 x 6 sq. mm | 80 | meter | 524 | ₹41,920 |
| 5 | Supply of XLPE insulated PVC sheathed Al. Conductor Armoured Cable 1.1KV grade conforming to IS-7098 of following size:-4C*25 sqmm | 90 | Meter | 237 | ₹21,330 |
| 6 | Laying and fixing of one number PVC insulated and PVC sheathed / XLPE power cable of 1.1 KV grade of following size on wall surface as required. | 90 | Meter | 64 | ₹5,760 |
| 7 | Supplying and fixing following way, horizontal type three pole and neutral, sheet steel, MCB distribution board, 415 V, on surface/ recess, complete with tinned copper bus bar, neutral bus bar, earth bar, din bar, interconnections, powder painted including earthing etc. as required. (But without MCB/RCCB/Isolator) (a) 8 way (4 + 24), Double door | 1 | Each | 6,337 | ₹6,337 |
stage.html
html • 0.03 MB
tech_eval.pdf
boq_comp_chart.xlsx
xlsx
fin_eval.pdf
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.03 MB
Tendernotice_1.pdf
PDF • 2.84 MB
Tendernotice_2.pdf
PDF • 17.56 MB
Tendernotice_3.pdf
PDF • 0.68 MB
DNIT01_0001.pdf
Tender Documents • 8.07 MB
BOQ_337796.xls
BOQ • 0.33 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .