Current Voltage Source Meter | Dual channel Sourcing and Measuring Voltage and Current simultaneously Resolution better or equal to 6 and half digits At least 10 A pulsing current on both channels simultaneously Compatible with 1 and half A or better DC current on both channel with TRIAX or banana cable Voltage ranges 200 mV to 200 V with resolution of 100 nv Current ranges 100 nA to 1 and half A DC or better with resolution of 1pA Should provide memory sweep and programming functions for voltage and current with internal memory capacity Memory buffer 20000 rdgs per buffer or better Should have 4 wire or remote sense measure capability Communication port GPIB USB Ethernet and LXI Compatible software to collect the data Warranty 3 Years OEM should have NABL accredited calibration lab and service center in India Training and Installation free of cost Vendor should be authorized from OEM | Sunita Srivastava 123031,Central Universityof Haryana |