RequiredOEM Annual Turnover Bid to RA enabled No Dental X Ray Machine ( 1 pieces ) Technical Specifications * As per GeM Category Specification Specification Specification Name Values Bid Requirement (Allowed Values) Standards Conformity to Standards CE (with 4 digit notified number) CE (with 4 digit notified number)
RequiredUSFDA
RequiredBIS (As per medical directive) Certification number and date - * 1/4 Type test approval from AERB required Yes * AERB type test certificate date and number - * Electrical safety standards IEC- 60601/IS 13450 * Performance Parameters Purpose Dental X ray machine is used for taking Intra Oral Periapical and Occlusal X ray * Tube voltage in KV 70 KV 70 KV Tube current in mA 4 - 8 mA 4 - 8 mA Focal spot in mm 0.4 * Should have Total filtration and Inherent Filtration Yes * Total filtration in mm Al 2 * Inherent filtration in mm Al 1.5 * Minimum range of exposure time in sec 2 * Power supply 220-240 V
Required50 Hz single phase * Based on DC Current Yes * RVG/X ray Film compatible Yes * Machine should have a stable base with sturdy locking wheel mechanism Yes * Machine shall be wall mountable Yes * Should have digital control equipped with an easy ready display indicating with precision Yes * Stool for the patient provided Yes Yes
RequiredNo Soft positioning arms for accurate tube positions Yes * Light and flexible movements Yes * Internally lead coated Yes * 2/4 head tube and cone to avoid scattered radiations Exclusive angular indicating system for head positioning in various radiography techniques No Yes
RequiredNo High efficiency and greater sharpness of the radiography Yes * Warranty in Years 3 3
Required5 Additional Parameters Should have high voltage generator with high efficiency in the emission of the X rays Yes * Should have free swivel head
Requiredwhich allows easy positioning of head Yes * Number of light weight lead equivalent aprons 2 (with lead thickness 0.5 mm) 2 (with lead thickness 0.5 mm)
Required3 (with lead thickness 0.5 mm) To be supplied with Thyroid shield Yes * User manuals to be supplied Yes * Reports Manufacturer should have ISO certification Yes * ISO certification date and number - * Copies of all certifications and reports to be provided to buyer on demand at time of supplies Yes * * Bold specifications are the golden parameters. * Bidders may note that In respect of non-golden Parameters