Loading…
Loading…
| # | Company | Amount | Rank | Status |
|---|---|---|---|---|
| 1 | L1₹31.9 LAccepted-AOC AT UNIT NOS TF B 07 A TO 15 PHOENIX MARKET CITY L B S ROAD KURLA WEST MUMBAI MAHARASHTRA 400070 | MUMBAI SUBURBAN | MAHARASHTRA | 400070 | L1 | Accepted-AOC LOWEST QUOTED | |
| 2 | L2₹35.0 L+₹3.1 L (9.79%)Rejected-Finance | L2 | Rejected-Finance SECOND LOWEST | |
| 3 | Rejected-Technical 66 4743 M G ROAD NORTH END ERNAKULAM ERNAKUALM KERALA 682035 | ERNAKULAM | KERALA | 682035 | - | Rejected-Technical Technically Not Qualified |
Tender Value
Refer Docs
EMD Value
₹35,000
Closing Date
15 Jul 2022, 4:00 pmClosed
THE REGISTRAR
M G UNIVERSITY KOTTAYAM
Supply and installation of the scientific equipment Flow Cytometer Analyzer
2022_MGU_497498_1
MGU/DST-PURSE/2017-18/ETEN/2
Open Tender
Equipments
Turn-key
90 days
M G UNIVERSITY
Please refer Tender documents.
6 documents required · 6 mandatory
₹5,200
Yes
₹35,000
Yes
9 May 2023
1 Jul 2022
18 Jul 2022
1 Jul 2022
15 Jul 2022
1 Jul 2022
Amount
Flow cytometer Analyzer
(A) Specification for Flow cytometer Analyzer (i) The flow Cytometer analyzer should be compact, Transportable, automated & bench top with a blue laser (488 nm) & red laser (640 nm). All Lasers & their excitation & collection optics should be fixed & pre-aligned. (ii) The System should have the capability of four fluorescence detectors and two light scatter detectors (forward & side scatter) with six parameters measurement. (iii) The flow cytometer system should be based on hydrodynamic focusing on the laminar flow within the Flow Cell (Quartz). (iv) The System should have pre-optimized detector settings. PMT detectors should be provided standard along with the system. (v) The Flow cytometer should have the capability of user-changeable optical filters. (vi) The system should be able to detect a minimum particle size of 0.5 micron and capable of aspirating samples with volumes of a minimum of 50ul. (vii) The system should have a data acquisition speed 10,000 events per sec. (viii) The system should accommodate a wide range of sample input tubes in manual loading with various sizes of microcentrifuge tubes, 12x75 mm or smaller sample tubes made of polypropylene or polystyrene, 48 / 96well plates (flat, round and v-bottom) and 96 well deep-well plates. (ix) System should be able to perform absolute counting without adding beads to the samples. (x) The system should have Fluorescence Sensitivity of at least MESF FITC <75; PE <50 (xi) The system should have 18-bit or higher signal processing, and digital data should be collected with a 5-7-decade dynamic range (B) Software & Data Management The system should come with suitable software for acquisition and data analysis for all applications. Should quote for both online & offline compensation software for data analysis. Design should be compatible with bead array applications. The company should provide all the software updates for the next three years for the system. Data management system: All in One PC workstation with the latest workstation & suitable monitor. (C) A startup kit should be provided along with the equipment for safe installations & demonstrations of the instrument (D) 5KV UPS with a minimum of two hours backup for working the instrument in emergencies.
BD India (P) Ltd
stage.html
html • 0.05 MB
tech_bid_open.pdf
tech_eval.pdf
fin_bid_open.pdf
boq_comp_chart.xlsx
xlsx
fin_eval.pdf
aoc.pdf
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.04 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .