RequiredExperience with Government
RequiredAnnual Turnover
RequiredOEM Authorization Certificate
RequiredOEM Annual Turnover Bid to RA enabled No Note: 1. OEM Turn Over Criteria: The minimum average annual financial turnover of the OEM of the offered product during the last three years
Requiredending on 31st March of the previous financial year
Requiredshould be as indicated in the bid document. Documentary evidence in the form of certified Audited Balance Sheets of relevant periods or a certificate from the Chartered Accountant/Cost Accountant indicating the turnover details for the relevant period shall be uploaded with the bid. In case the date of constitution/incorporation of the OEM is less than 3 year old
Requiredthe average turnover in respect of the completed financial years after the date of constitution shall be taken into account for this criteria. Ventilator For Adult & Paediatric Ventilation ( 2 pieces ) Technical Specifications * As per GeM Category Specification 1/8 Specification Specification Name Values Bid Requirement (Allowed Values) Set 1 Usage: Medical ventilator is suitable for adult & Paediatric ventilation. It is a ventilator/a machine designed to move breathable air into and out of the lungs
Requiredto provide breathing for a patient who is physically unable to breathe
Requiredor breathing insufficiently. * Ventilation monitoring facility: The Ventillator is a Microprocessor Controlled ventilator with integrated facility for Ventilation monitoring. * Ventilator Features Have Standard hinged arm holder for holding the circuit: Yes * Have Coloured Touch screen: Yes * Colour Touch screen Size (diagonal): 12 12
Required13
Required14
Required15
Required17 Or higher Has facility to measure and display 3 waves- Pressure and Time
RequiredVolume and Time and Flow and Time
Required2 loops (P-V
RequiredF-V) with facility of saving of 2 Loops for reference: Yes * Has Status indicator for Ventilator mode
RequiredBattery life
Requiredpatient data
Requiredalarm settings
Requiredclock etc: Yes * Has Graphic display to have automatic scaling facility for waves: Yes Yes
RequiredNo Has Trending facility of minimum duration of: 48 24
Required48 Or higher Facility for automatic compliance & Leakage compensation for circuit available: Yes * 2/8 Ventilator Settings/Range Parameters Tidal Volume: 20 ml – 2000 ml 20 ml – 2000 ml Pressure (insp) : 2 – 90 cm H2O * Pressure Ramp/Flow pattern: Yes * Respiratory Rate : 1 to 150 bpm * I : E Ratio : 5:1 to 1:10 * CPAP/PEEP : 0 – 50 cm H2O * Pressure support : 2 – 75 cmH2O * FIO2 : 21 – 100% 21 – 100% Pause Time : 0- 2 sec * Pressure trigger : 0.5 – 15cmH2O * Flow Trigger : 0.2 – 12 LPM * Available parameter monitoring facility of Ventilator Airway Pressure (Peak & plateau) Yes Yes Tidal volume (Inspired & Expired) Yes Yes Minute volume (Inspired and Expired) Yes Yes Spontaneous Minute Volume Yes Yes Total Frequency Yes Yes FIO2 dynamic Yes Yes Use selector Alarms for all measured & monitored parameters Yes Yes Intrinsic PEEP and PEEPi Volume (or trapped Volume) Yes Yes Resistance ( Rinsp & Rexp)& Compliance ( Cdyn & Cstat) Yes Yes Respiratory mechanics: Yes Yes Ventilation Modes & Other Features Modes of Ventilation available: Volume controlled
RequiredPressure Controlled
RequiredPressure Support
RequiredSIMV (Pressure Control and volume control) with pressure support
RequiredCPAP/PEEP
RequiredInverse Ratio Ventilation
RequiredNon Invasive ventilation
RequiredYes * 3/8 Apnea/backup ventilation
RequiredMMV+PSV ASV (Adaptive Support Ventilation) or equivalent
RequiredAPRV bivent/Bi-level/BiPAP or equivalent & PRVC Expiratory block is autoclavable and no routine calibration required: Yes * Nebuliser with capability to deliver particle size of 3 to 5 micron & to be used in both Off and On line Yes * Apnea/backup ventilation with Automatic switch over from spontaneous to back and vica-versa Yes * Automatic Patient connection and disconnection Detection facility Yes * Non-forced Slow Vital Capacity
Requiredphysiologic Dead space
RequiredRSBI
RequiredImposed work of Breathing ( WOBi)
RequiredExpiratory Time constant ( Tcexp) Yes * Medical Air Compressor Ventilator has Stand- alone Medical Air compressor snap fit with the Ventilator module to provide an oil free Medical air
Requiredwith Peak output flow should be minimum 160 LPM and Air quality complying with ISO compressed air purity class: Yes
RequiredMedical Ventilator has Stand-alone Medical Air compressor. Yes
RequiredMedical Ventilator has Stand-alone Medical Air compressor. Or higher Medical Air Compressor should automatically activate in the event of wall air supply loss and replacement of internal filters should be performed without removing the compressor
Requiredhave Yes * 4/8 washable air filter: Reusable face mask & nasal Mask Reusable face & nasal mask shall be with textured dual flap silicone cushion flap for easy fit/Removable forehead support and pad are to match the angle of patient’s forehead Stability/Provision of selector shall be for easy fit and angle headgear attachments/Shall be autoclavable Yes * Accessories Accessories to be supplied with each unit of the equipment: Adult & Paediatric autoclavable silicone breathing circuits: 02 Each
RequiredReusable Masks (Medium
RequiredLarge) with each machine: 02 sets each
RequiredReusable and autoclavable Flow sensor and exhalation valve - 2 each
RequiredAccessories for non invasive ventilation: 2 sets
RequiredHumidifier: Servo controlled with digital monitoring of inspired gas temperature complete with heating wire - 01 no & (Filter paper for humidifier for 100 uses : 01 lot) Adult & Paediatric autoclavable silicone breathing circuits: Nil 0 unit
RequiredReusable Masks (Medium
RequiredLarge) with each machine: Nil/0 unit
RequiredReusable and autoclavable Flow sensor and exhalation valve - Nil/0 unit
RequiredAccessories for non invasive ventilation: Nil/0 unit
RequiredHumidifier: Servo controlled with digital monitoring of inspired gas temperature complete with heating wire - 01 no & (Filter paper for humidifier for 100 uses: 01 lot)
RequiredAdult & Paediatric autoclavable silicone breathing circuits: 01 Each
RequiredReusable Masks (Medium
RequiredLarge) with each machine: 01 sets each
RequiredReusable and autoclavable Flow sensor and exhalation valve - 1 each
RequiredAccessories for non invasive ventilation: 1 set
RequiredHumidifier: Servo controlled with digital monitoring of inspired gas temperature complete with heating wire - 01 no & (Filter paper for humidifier for 5/8 100 uses : 01 lot)
RequiredAdult & Paediatric autoclavable silicone breathing circuits: 02 Each
RequiredReusable Masks (Medium
RequiredLarge) with each machine: 02 sets each
RequiredReusable and autoclavable Flow sensor and exhalation valve - 2 each
RequiredAccessories for non invasive ventilation: 2 sets
RequiredHumidifier: Servo controlled with digital monitoring of inspired gas temperature complete with heating wire - 01 no & (Filter paper for humidifier for 100 uses : 01 lot) Or higher Certification & Warranty Whether Ventilator unit along with medical air compressor unit are US FDA certified: No Yes
RequiredNo Whether Ventilator unit along with medical air compressor unit are European CE certified: Yes Yes Manufacturer is ISO certified for quality standards: Yes Yes Equipment meets IEC-60601-1-2: General Requirements of Safety for Electromagnetic Compatibility or should comply with 89/366/EEC; EMC- directive Yes * Warranty Period: 5 3
Required5 Or higher Power Supply Offered Equipment works on 220-240V AC as well as batteries
RequiredMains adaptor to be supplied: Yes * Whether offered Ventilator unit to be supplied with suitable Servo Yes Yes 6/8 controlled Stabilizer CVT (along with minimum 30 minutes power back up for ventilation) : Resettable over current breaker shall be fitted for protection of equipment: Yes * * Specifications highlighted in bold are the Golden Parameters. * Bidders may note that In respect of non-golden Parameters
Requiredthe specifications 'Values' chosen by Buyer will generally be preferred over 'Bid requirement ( allowed Values) by the Buyer.