Gmers Medical College Attached Hospital,Porbandar
Pathology Laboratory Equipment
352551
2/2026-27 OF PORBANDAR SR. No.05
Open
Medical Equipments/Waste
Works
Porbandar
15 documents required · 15 mandatory
₹5,000
D.M.F MEDICAL SUPERINTEDENT GMERS MEDICAL COLLEGE ATTACHED HOSPITAL PORBANDAR
₹1.5 L
7 Oct 2026
7 Oct 2026
7 Oct 2026
22 Oct 2026
7 Oct 2026
LABORATORY EQUIPMENTS
Sr. No. Name of Equipment Specifications Qty Per Unit Estimate Price Total Unit Estimate Price
1 FUJI Dry Chemistry NX-600 Specifications State-of-the art proprietary dry chemistry technology from Japan. Operations should be completely waterless (dry). Result quality is not impacted by the quality of water. Analyzer should be able to deliver 22 results in 15 minutes. Measurement time for electrolytes should be less than 1 minute for electrolyte tests. All electrolyte parameters (Na,K,Cl) should be available in a single slide The analyzer should support 28 colorimetric tests and 3 electrolyte tests. Sample volume required should be 10 μL. The analyzer should have automated calibration using QC Card. No trained manpower required. • Measurement principle for biochemistry should be colorimetric method. • Measurement principle for electrolytes should be potentiometric method using ion-selective electrode slides. • The analyzer should support serum and plasma samples. The analyzer should have a throughput of up to 128 tests per hour. • The analyzer should have built in incubator. • The analyzer should support automatic identification of test item and reagent lot using slide information. • The anayzer should have automatic dilution function. • The analyzer should have a 7-inch color touch panel display. The analyzer should have a built-in thermal printer. • The analyzer should support data transmission to PC via RS-232C, USB and LAN interfaces. • The analyzer should operate on single-phase AC power supply, 100–240 V ±10%, 50–60 Hz. • The analyzer should operate within a temperature range of 15–32°C. • The analyzer should support optional barcode reader. 1 800000 800000
2 Cell Counter 5 Parts Specifications: Principle: • Flow cytometry +tri-angle laser scatter+ chemical dye method for WBC 5-diff differential analysis and W8C counting • Impedance method for RBC and PLT counting • Cyanide free reagent for HGB measurement Mode: CBC, CBC+DIFF Throughput: 60 samples per hour Parameters: • 25 reportable parameters: WBC, Lym%, Mono%, Neu%, Bas%, Eos%, Lym#, Mono#, Neu#, Bas#, Eos#, RBC, • HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PCT, PDW, PLCC, PLCR, • 4 research parameters: ALY%, ALY#, LIC%, UC# Graph: 3-D DIFF scatter gram 3 2-D DIFF scatter gram 2-D BASO scatter gram 3 histograms for WBC, RBC and PLT Sample Mode: Venous whole blood, Capillary whole blood, pre-diluted Sample Volume: Venous/Capillary whole blood: 18 μl Reagent: Z5 LD Lyse, Z5 LB Lyse, Z5 DN Diluent, Probe cleanser Calibration: Manual calibration Auto calibration: Fresh blood, calibrator QC Rule: L-J, X-B Screen: 8.4-inch touch screen Interface: • 4 USB port for external printer, external barcode scanner, mouse, keyboard, etc. • 1 network port, HL7 protocol, support LIS. Storage Capacity: >=50,000 results including numeric and graphical information Power Supply: (100-240) V~, 50/ 60Hz, 200 VA Operation environment: • Temperature: (10-30) °C • Humidity: (20-85) % • Atmospheric Pressure: (70-106) kPa Weight: 28.5 Kg Dimension (mm): 230(W)* 455(0) * 435 (H) 1 1500000 1500000
3 Tosho - Immunology — AIA 360 Technical Characteristics: • Benchtop system • Stackable (1-6 modules) • 2 Independent sections x 6 positions • Plug and play • Integrated PC (gateway) to manage up 2 modules • All-inclusive kit • Unique Single Dose Concept • Broad Parameter Menu of more than 80 parameters • Strip & SPR compatibility verification • Optical detection of sample • Automat ed detection of leakage and clogs • Bi-directional connectivity • Brand new software interface • Compliance with cybersecurity (Secure remote access & prevent against cyberattack) • Supported by an acknowledged and high- quality service team. • Remote maintenance through 1 2100000 2100000
4 Automated ESR Analyser Technical Characteristics: Application: Automated Erythrocyte Sedimentation Rate (ESR) measurement using modified Westergren method — direct from EDTA tubes used for CBC (no need for citrate tubes) — reducing sample prep and waste. Method: Modified Westergren ESR measurement. No citrate tubes required — use EDTA tubes directly. 10-channel random access ESR testing saves time & increases throughput. Fast results (~20 min). User-friendly touch screen interface. Minimal bio-waste and environmentally friendly operation. Connectivity to LIS and peripherals (barcode, printer) Certifications & Standards: ISO 13485 certified, US FDA Registered, CDSCO Registered (India), CE Certified, Electrical/EMC: IEC 61010-1, IEC 61010-2-101 & IEC 61326 series standards 1 350000 350000
5 Urine Analyser (Strip Analyser / Automated Urine Analyser) Technical Characteristics: Operation Mode: Semi-automatic urine analyzer Measurement Method: Reflectance photometer with CCD color image sensor and LED light source Test Throughput: Up to 720 tests/hour (~5 sec/test) Measurement Cycle: ~5 seconds per strip Analytes: Works with URiSCAN 4 – 11 parameter urine test strips (e.g., blood, protein, glucose, pH, SG, nitrite, leucocytes, bilirubin, urobilinogen, ketones, etc.) Memory Capacity: ~3,000 patient test records, ~1,000 control records, ~30 calibration records Display: 7″ TFT LCD touchscreen (800 × 480) Software Languages: Multilingual support (English, German, Spanish, Italian, Portuguese, Chinese, Russian, Korean, etc.) Printer: Built-in high-speed thermal printer (203 dpi) Host Communication: RS-232C and USB interfaces for PC/lab network connection. Peripherals Supported: Barcode reader, external printer. Safety Standard: EN 61010-1 / EN 61010-2-101 EMC Standard: EN 61326-1 / EN 61326-2-6 1 50555 50555
6 Microscope (Olympus Imported Model No. CX23) Specifications: • Objectives 4x, 10x, 40x (spring loaded), 100x (spring loaded, oil), antifungal • Nose Pieces Quadruple Nosepiece (Ball bearing type) with rubber grip, click stop for exact position • Stages Rectangular stage size 135 x 124 mm, X/Y travel range 76mm x 50mm. Low drive movement controls, single specimen holder • Voltage Supply 100 - 240 V • Optional Accessories Phase Contrast kit with 10x and 40x LP Series Phase Objective, Polarizing 1 174500 174500
7 Centrifuge The centrifuge machines are mainly used for solid-liquid separation in chemical, pharmaceutical and other industries. We are the renowned manufacturer, exporter and supplier of centrifuge machines made from export body quality. With the innovative and latest technology available, we bring forth revolutionize new design machines. Its easily operational due to it speed control switch and provide an optional timer manual and both digital. With our expert panel, these machines are well researched according to the international parameters of scientific production. Make: • Type: Swing Out Medico Head • Cap: 6 Tubes (15 ml each) • Rpm: 3500-4000 rpm • Speed Control Switch • Make: Deluxe • Type: Rotor Angle Head (Doctor) Size: 6 Tube 1 49998 49998
TOTAL:- 5025053
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
RAW_HTML
Laboratpray Equipment final.pdf
SLB - Laboratpray Equipment final.pdf
SLB - Laboratpray Equipment final.docx.xlsx
XLSX
JAHER NIVIDA NO.2 2026-27.pdf
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .