RequiredExperience with Government
RequiredAnnual Turnover
RequiredCertificate (Requested in ATC)
RequiredOEM Authorization Certificate
RequiredOEM Annual Turnover Bid to RA enabled Yes Note: 1. OEM Turn Over Criteria: The minimum average annual financial turnover of the OEM of the offered product during the last three years
Requiredending on 31st March of the previous financial year
Requiredshould be as indicated in the bid document. Documentary evidence in the form of certified Audited Balance Sheets of relevant periods or a certificate from the Chartered Accountant/Cost Accountant indicating the turnover details for the relevant period shall be uploaded with the bid. In case the date of constitution/incorporation of the OEM is less than 3 year old
Requiredthe average turnover in respect of the completed financial years after the date of constitution shall be taken into account for this criteria. SPECTROPHOTOMETER ( 1 pieces ) Technical Specifications * As per GeM Category Specification 1/5 Specification Specification Name Values Bid Requirement (Allowed Values) TYPE Type of Spectrophotometer VISIBLE VISIBLE Type of Optical System Single Beam Single Beam Monochromator Type NA NA Detector Type Photodiode Photodiode Spectral Bandwidth Type Fixed Bandwidth Fixed Bandwidth LIGHT SOURCE Light Source Tungsten Halogen Lamp Tungsten Halogen Lamp Diffraction Grating (lines/mm) 1200 1200 Switching of Light Source Automatic Automatic Wavelength Setting And Scanning Automatic Automatic OPTICAL PERFORMANCE Source Wavelength - Minimum 320 320.0 Source Wavelength - Maximum 1000 1000
Required1100
Required1500
Required2000
Required2500
Required2700
Required3000
Required3200
Required3300
Required1650 Or higher Resolution 0.3 0.3 Wavelength Accuracy (+/-) 1 1.0 Bandwidth (nm) 4.0 4.0 If variable band width
RequiredSpecify the incremental step value(mm) NA * Scanning Speed (nm/min) 0 * Focal length (mm) 0 to 49 0 to 49 Photometric measurement modes Absorbance Absorbance Photometric Absorbance (Max) (Abs) (+/-) 0.4 0.4 Photometric Absorbance Accuracy (Max) (+/-) 0.001 0.001 - 0.001 Or lower Photometric Transmittance (Max) (+/-) 0 0.0 2/5 Photometric Transmittance Accuracy (Max) (+/-) 0 0.0 - 0.0 Or lower Photometric Reflectance (Max) (+/-) 0 0.0 Photometric Reflectance Accuracy (Max) (+/-) 0 * Stray Light Correction N/A * Baseline Correction N/A * GENERIC Display In-built display In-built display Display type LCD touchscreen LCD touchscreen Display Size 2.5 2.5 - 30.0 Or higher Test result printing Through external printer Through external printer If inbuilt printer
RequiredType of printer N/A N/A Suitable software for connecting PC & Printer Yes * Connectivity USB * Internal storage(GB) 0 0.0 - 80.0 Or higher Sample switching/selection Automatic * Capacity of Sample holder/Sample Carousel 1 * Mximum Path Length of sample 0 0.0 Number of Cuvettes supplied-inclusive in the scope of supply 10 10 Power Supply 230 V ± 10
Required50 Hz * Any other features/accesories covered in scope of supply - * Spares and Consumables inclusive in the scope of supply No Yes
RequiredNo List of Items and Quantity of each item included in the offer - * OPERATING CONDITIONS Minimum Operating Temperature 0 -5
Required-10
Required0
Required5
Required10 3/5 Maximum Operating Temperature 35 35.0 Operating Humidity (RH) (%) at 40 degree C 85 85.0 WARRANTY Onsite Warranty 1 1.0 Availability of Spares (at extra cost) 2 2.0 Certification Availability of UL/CE Certification as per EN 61010-1:2010 Safety requirements for electrical equipment for measurement
Requiredcontrol
Requiredand laboratory use General requirements Yes Yes Conformance to EMC/EMI as per EN 61326-1:2013 Electrical equipment for measurement
Requiredcontrol and laboratory use Yes Yes Availability of Test Reports from Central Govt/NABL approved/ILAC acredited lab to prove conformity to the specification Yes Yes If yes test report to furnished
Requiredto the buyer on demand Yes * Test Report number and date - * Name and address of the Lab - * * Specifications highlighted in bold are the Golden Parameters. * Bidders may note that In respect of non-golden Parameters
Requiredthe specifications 'Values' chosen by Buyer will generally be preferred over 'Bid requirement ( allowed Values) by the Buyer.