| Sl No | Description | Qty | Unit | Laboratory Equipment Stores L1 | KERALA STATE INDUSTRIAL ENTERPRISES LTD L2 |
|---|---|---|---|---|---|
| 1.00JJM 2021-22 - Setting up / Upgradation / Operation of Laboratories for WQM&S - Procurement of 70 nos. Membrane Filtration System for various Quality Control Laboratories of Kerala Water Authority | |||||
| 1.01 | Membrane filtration units with Manifold 3- place (minimum) and vacuum pump
I. Vacuum Pump
• The Vacuum pump should transfer filtered liquids directly to waste• Does not require a vacuum flask to be connected between the pump and filtration assembly as direct transfer ensures no handling of liquid waste• The Vacuum Pump should be capable to be used both on work bench and inside Laminar flow hood as well• Complies to the ISO 8199 Standard for Water Quality (general guidance on the enumeration of microorganisms by culture) which describes the conditions for membrane filtration. • Vacuum Pump should be maintenance free and comply with Electrical regulations & norms. • Tubing should be autoclavable silicone.• Vacuum :Maximum700mbar • .Flow rate (measured with water at 20°C).Minimum 3.5l/min• Protection type IP64, Protection Class III .• Temperature range of the handled sample+5°Cto+80°C.• Maximum viscosity of the liquid 150cSt.
II. Manifold
• Manifold to have different filtration heads which shall fit to both reusable and disposable filtration devices. • Each component shall be able to remove quickly and easily without the use of tools. • Flexible to choose the flow direction, the quick connector and the quick plug fit either end of the manifold pipe. • Universal Filter heads for manifold. Material of Construction – 316 L Stainless Steel or Anodized HE30 aluminum. • All components shall be autoclavable.
III. Funnel
• Should supply 100ml graduated, transparent autoclavable funnel having protective rim and vented lid • Funnel material shall be clear with all around level indicators.
IV. Hard copy of instrument manual with specification details
V. IQ/OQ/PQ documentation shall be provided
VI. Warranty 5 years for the Membrane Filtration system including Vacuum pump | 70 | No | 2,62,000 ₹1,83,40,000 Lowest | 3,20,000 ₹2,24,00,000 |
Tender Value
₹2.5 Cr
EMD Value
₹1 L
Closing Date
5 Nov 2022, 3:00 pmClosed
Director, SRI, Nettoor
SRI, Nettoor
JJM 2021-22 - Procurement of 70 no Membrane Filtration System for various Quality control laboratories of Kerala Water Authority
2022_KWA_517616_1
02/2022-23/SRI/NTR
Open Tender
Laboratory and scientific equipment
Supply
120 days
As per Annexure A and B
Please refer tender documents
11 documents required · 11 mandatory
₹7,500
₹1 L
Yes
SRI Nettoor (online)
9 Jan 2023
18 Oct 2022
8 Nov 2022
18 Oct 2022
5 Nov 2022
18 Oct 2022
18 Oct 2022 - 25 Oct 2022
26 Oct 2022
Amount
JJM 2021-22 - Setting up / Upgradation / Operation of Laboratories for WQM&S - Procurement of 70 nos. Membrane Filtration System for various Quality Control Laboratories of Kerala Water Authority
Membrane filtration units with Manifold 3- place (minimum) and vacuum pump I. Vacuum Pump • The Vacuum pump should transfer filtered liquids directly to waste• Does not require a vacuum flask to be connected between the pump and filtration assembly as direct transfer ensures no handling of liquid waste• The Vacuum Pump should be capable to be used both on work bench and inside Laminar flow hood as well• Complies to the ISO 8199 Standard for Water Quality (general guidance on the enumeration of microorganisms by culture) which describes the conditions for membrane filtration. • Vacuum Pump should be maintenance free and comply with Electrical regulations & norms. • Tubing should be autoclavable silicone.• Vacuum :Maximum700mbar • .Flow rate (measured with water at 20°C).Minimum 3.5l/min• Protection type IP64, Protection Class III .• Temperature range of the handled sample+5°Cto+80°C.• Maximum viscosity of the liquid 150cSt. II. Manifold • Manifold to have different filtration heads which shall fit to both reusable and disposable filtration devices. • Each component shall be able to remove quickly and easily without the use of tools. • Flexible to choose the flow direction, the quick connector and the quick plug fit either end of the manifold pipe. • Universal Filter heads for manifold. Material of Construction – 316 L Stainless Steel or Anodized HE30 aluminum. • All components shall be autoclavable. III. Funnel • Should supply 100ml graduated, transparent autoclavable funnel having protective rim and vented lid • Funnel material shall be clear with all around level indicators. IV. Hard copy of instrument manual with specification details V. IQ/OQ/PQ documentation shall be provided VI. Warranty 5 years for the Membrane Filtration system including Vacuum pump
KERALA STATE INDUSTRIAL ENTERPRISES LTD
stage.html
html • 0.03 MB
tech_bid_open.pdf
tech_eval.pdf
fin_bid_open.pdf
boq_comp_chart.xlsx
xlsx
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.04 MB
KWAGeneralConditionsforeTenderingV5.pdf
PDF • 0.69 MB
Tendernotice_1.pdf
PDF • 0.22 MB
NITFiltration.pdf
Tender Documents • 1.08 MB
TCFiltration.pdf
Tender Documents • 0.19 MB
BOQ_742671.xls
BOQ • 0.28 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .