| 1.00Elemental Analyser for automatic determination of C-H-N-S (CHNS Analyser) | |||||
| 1.01 | Fully Automated PC controlled Elemental Analyser with following specifications for analysing solid and liquid samples
1 Instrument
Sample weight Range 0.1 to 1000 mg or better.
Measuring range 0.01% to 100% for all elements. Also mention the absolute measuring range of all elements with proof
Standard deviation ≤0.1% of absolute
Measurement Time ≤10 minutes in CHNS mode
Controller Must be equipped with electronic mass flow controller and digital flow sensors for accurate readout of flow
Pressure and flow sensors All pressure and flow sensors must be electronic for better accuracy. There should be no analogue pressure gauges. Flow meters on the analyser must be computer controlled.
Kit for the measurement of fluorinated samples.
2. Furnace System
Furnace Should have two zone furnace system, separate for combustion and reduction with independent temperature control for each furnace.
Temperature Controlled furnace Temperature should be 11000C or more. Documentary proof needs to be attached
Additional features Should have possibility to use ceramic/quartz ash finger to handle high inorganic/halogen or fluorine contents sample.
Warranty
10-years free replacement warranty on the furnace
3. Separation System
Separation Advanced Chromatographic separation of gases.
Full separation of all analytes and there should be no peak tailing or peak overlap.
Sulphur being a crucial element, should be analysed with higher column temperature (separation column) than Nitrogen, Carbon, and Hydrogen. All necessary accessories/ options in this regard should be quoted as mandatory items.
4. Autosampler System
Sampler system Electrical auto sampler system with 60 positions or more. Pneumatic auto-sampler is not acceptable.
Requirement
Zero blank sample injection system
5. Detector System
Detector Oxygen intrusion free thermistor technology-based Temperature stabilised TCD detector or IR detector for measurement of C-H-N-S.
Warranty
10 years warranty on TCD or IR detector needs to be offered.
6. Consumables
Consumables Consumables for 5,000 sample analyses in CHNS mode.
500 Tin capsules need to be offered separately.
Sample preparation devices Sample preparation devices for both solid and liquid samples
Gases
UHP grade Helium gas cylinder with double stage SS regulator - 1 No
UHP grade Oxygen gas cylinder with double stage SS regulator – 1 No
7. Software
Windows based, compatible with the recent version i.e., windows 10 PRO or higher and should have display of set and actual pressures, flow rates, temperatures, number of samples analysed with provision for setting maintenance interval with warning when maintenance needed. System/Software updates to be given free of cost.
Should have segmented leak check through software to enable identification of exact position of leaks.
Windows PC with i5 or higher with 8GB Ram or higher and 1TB HDD +Windows 10 OS or higher with 21” monitor needs to be offered.
8. Warranty
The entire instrument along with all accessories should cover a warranty of 3 years and further annual maintenance contract for 2 years | 1 | Nos | 73,75,000 ₹73,75,000 | 37,78,439 ₹37,78,439 Lowest |
Tender Value
Refer Docs
EMD Value
₹43,700
Closing Date
17 Nov 2025, 3:00 pmClosed
Registrar KFRI
Kerala Forest Research Institute , Peechi P.O., Thrissur District, Kerala, Pin-680653
supply, installation, demonstration and commissioning of Elemental Analyser for automatic determination of C-H-N-S
2025_KCSTE_817092_1
KFRI/1363/2025-P2-02
Open Tender
Equipments
Item Rate
60 days
KFRI PEECHI
Please refer Tender documents.
12 documents required · 12 mandatory
₹6,500
Yes
₹43,700
Yes
5 Jan 2026
31 Oct 2025
18 Nov 2025
31 Oct 2025
17 Nov 2025
31 Oct 2025
Amount
Elemental Analyser for automatic determination of C-H-N-S (CHNS Analyser)
Fully Automated PC controlled Elemental Analyser with following specifications for analysing solid and liquid samples 1 Instrument Sample weight Range 0.1 to 1000 mg or better. Measuring range 0.01% to 100% for all elements. Also mention the absolute measuring range of all elements with proof Standard deviation ≤0.1% of absolute Measurement Time ≤10 minutes in CHNS mode Controller Must be equipped with electronic mass flow controller and digital flow sensors for accurate readout of flow Pressure and flow sensors All pressure and flow sensors must be electronic for better accuracy. There should be no analogue pressure gauges. Flow meters on the analyser must be computer controlled. Kit for the measurement of fluorinated samples. 2. Furnace System Furnace Should have two zone furnace system, separate for combustion and reduction with independent temperature control for each furnace. Temperature Controlled furnace Temperature should be 11000C or more. Documentary proof needs to be attached Additional features Should have possibility to use ceramic/quartz ash finger to handle high inorganic/halogen or fluorine contents sample. Warranty 10-years free replacement warranty on the furnace 3. Separation System Separation Advanced Chromatographic separation of gases. Full separation of all analytes and there should be no peak tailing or peak overlap. Sulphur being a crucial element, should be analysed with higher column temperature (separation column) than Nitrogen, Carbon, and Hydrogen. All necessary accessories/ options in this regard should be quoted as mandatory items. 4. Autosampler System Sampler system Electrical auto sampler system with 60 positions or more. Pneumatic auto-sampler is not acceptable. Requirement Zero blank sample injection system 5. Detector System Detector Oxygen intrusion free thermistor technology-based Temperature stabilised TCD detector or IR detector for measurement of C-H-N-S. Warranty 10 years warranty on TCD or IR detector needs to be offered. 6. Consumables Consumables Consumables for 5,000 sample analyses in CHNS mode. 500 Tin capsules need to be offered separately. Sample preparation devices Sample preparation devices for both solid and liquid samples Gases UHP grade Helium gas cylinder with double stage SS regulator - 1 No UHP grade Oxygen gas cylinder with double stage SS regulator – 1 No 7. Software Windows based, compatible with the recent version i.e., windows 10 PRO or higher and should have display of set and actual pressures, flow rates, temperatures, number of samples analysed with provision for setting maintenance interval with warning when maintenance needed. System/Software updates to be given free of cost. Should have segmented leak check through software to enable identification of exact position of leaks. Windows PC with i5 or higher with 8GB Ram or higher and 1TB HDD +Windows 10 OS or higher with 21” monitor needs to be offered. 8. Warranty The entire instrument along with all accessories should cover a warranty of 3 years and further annual maintenance contract for 2 years
3V Solution (BID ID -2440240)
stage.html
html • 0.04 MB
tech_bid_open.pdf
tech_eval.pdf
boq_comp_chart.xlsx
xlsx
fin_eval.pdf
aoc.pdf
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.03 MB
Tendernotice_1.pdf
PDF • 0.17 MB
AGREEMENT.pdf
Other Document • 0.04 MB
tenderdoc.pdf
Tender Documents • 0.46 MB
BOQ_1369178.xls
BOQ • 0.24 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .